FREQUENCY ANALYSIS OF CATHETER SYSTEMS USED FOR INVASIVE BLOOD PRESSURE MONITORING by Daniel Michael Chernoff // SUBMITTED IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREES OF BACHELOR OF SCIENCE and MASTER OF SCIENCE at the MASSACHUSETTS INSTITUTE OF TECHNOLOGY June, 1982 O Daniel Michael Chernoff 1982 The author hereby grants to MIT permission to reproduce and to distribute copies of this thesis document in whole or in part. ..... ........ ofi................... Department of Electrical Engineering and Computer Science, June, 1982. Signature of Author. C .... M ... ., ......T ..... Supervisor-Ac Certified by.......~ge Roger G. Mark, M.D., Ph.D. Thesis Supervisor-Academic Certified by......avid David ..* ......... .................. ... EI11js, Thesis Supervisor (Cooperating Company) -/ Archives OF TECHNOLOGY Acceptedur .. " m - a............. - -I ...... ' " "....,"." C*T" *'1"92 Arthur C. Smith, Chairman Departmental Committee on Graduate Stu ents S1DSIRARlIFB 1U ACKNOWLEDGEMENT I am grateful for the guidance, support and surplus of enthusiasm which Mr. David Ellis supplied during the research and writing of this thesis. truly invaluable. His help and criticisms were Dr. Roger Mark provided perspective on the problem and aided in directing the focus of the research. I wish to thank the many people at Hewlett-Packard's Waltham Division who supplied ideas and moral support throughout this project. TABLE OF CONTENTS Page ABSTRACT.. . . . . .•·. · · iii LIST OF TABLES. . . . .. . .. . . . . . . LIST OF FIGURES . . . .. .. . . . . . . vii . vi CHAPTER a . . a . a a a . . . . . INTRODUCTION 1.1 1.2 1.3 1.3.1 1.3.2 1.4 1.4.1 1.4.2 1.5 1.5.1 1.5.2 Brief History of Invasive Monitoring .0. Requirements for Accurate Waveform Reproduction. . . . . . . . The Fluid-Filled Catheter.. . . . . Models . . . . . . . . . . . . . . Distortion . . . . . . . . . . . . Frequency Response Measurement Techniques . . . . . . . . . . . . . Direct techniques.. . . . . . . . Indirect techniques.. . . . . . . Compensation . . . . . . . . . . . . Mechanical compensation.. . . . . Electrical compensation.. . . . . MODELING 2.1 OF THE CATHETER ---- ~ ~ SYSTEM. . . . 1 2 . 17 . . 17 General Model - Mechanical/Electrical Analogies.. . . . . . . . . . Theoretical Calculation of Line Constants . . . . . . . . 2.2.1 Longitudinal impedance . . . 2.2.2 Transverse impedance . . . . 2.3 Transmission Line Formulation. 2.3.1 Telegraph equations and propagation constant . . . . . . . 2.2 2.3.2 Characteristic impedance . . . . . . . . . . . . . . . . . .. 2.3.3 Boundary conditions and reflection coefficient . . . . . . 2.3.2 Natural frequencies.. . . . . . . 2.4 Lumped Model Approximation . . . . . 2.5 Effect of Trapped Air Bubbles. . . . 2.5.1 Compliance of air bubbles.. . .... 2.5.2 Relationship between bubble location and resonant frequency ... ...... . . 38 CHAPTER Page . . . . . . . . . . 3 MATERIALS . 3.1 3.2 3.3 3.5 3.6 3.7 Extension Tubing . ....... · Transducers. . .......... · Flush Bag and Fluid. . ......· Slow/Fast Flush Unit . ..... · Bench Equipment. . ........ · Flow Source. ... ........ · Tap Generator. ......... . · 4 METHODS..AND RESULTS.. 3.4 . . . . . · · · · · · · . .. . 41 · · · · · · · 42 42 43 44 44 45 46 . . 4.1 Determination of Line Parameters . 4.1.1 Resistance . . . . . . . . . . . 4.1.2 Inertance. . . . . . . . . . . . 4.1.3 Compliance . . . . . . . . . . . 4.2 Determination of Resonant Frequency for the Bubble-free System . . . . 4.2.1 Experimental . . . . . . . . . . 4.2.2 Theoretical . .. ......... 4.3 Air Bubble Experiments . . . . . . 4.3.1 Resonant frequency as a function of bubble location . . . . . . . 4.4 Tap and Flush Experiment . . . . . 4.4.1 Experiment . . . . . . . . . . . · · · · . · · · · · · · · . . . . . . . 53 · · · · 59 59 62 65 · · · · · · 65 70 71 . . . . . Tap and Flush Responses.. 5.1 Initial conditions . . . . . . . . 5.1.1 Transient solution . . . . . . . . 5.1.2 5.2 Extraction of Resonant Frequency and Damping from Fast-Flush. . . . . . . 5.3 Anticipated Usage and Clinical Acceptability.. . . . . . . . . . . . 49 50 51 · · · · DISCUSSION . . . . . . . . . . . . . . . BIBLIOGRAPHY . 49 . 80 81 83 84 89 90 94 APPENDICES Electrical/Hydraulic Analogies . . . Calculation of fn and D from Step Response . . . . . . . . . . . . . . 98 99 FREQUENCY ANALYSIS OF CATHETER SYSTEMS USED FOR INVASIVE BLOOD PRESSURE MONITORING by Daniel Michael Chernoff Submitted to the Computer Science the requirements and Master of Department of Electrical Engineering and on May 1i, 1982 in partial fulfillment of for the Degrees of Bachelor of Science Science in Electrical Engineering. ABSTRACT The resonant behavior of fluid-filled catheter manometers can produce severe distortion in the monitored blood pressure waveform. Although gradual improvements in components has resulted in catheter systems having adequate frequency response for human blood pressure measurement, these systems are frequently compromised by the presence of occult air bubbles in the fluid column. A general model was developed to predict the frequency response of catheter systems in terms of a limited number of lumped second-order sections. This model was experimentally verified by direct frequency response measurement and by independent measurement of component characteristics. This model also successfully predicts the effect of bubble size and location on the frequency response. A previously-proposed technique of measuring the frequency response of a fluid-filled catheter system in vivo was described and theoretically justified in terms of the lumped element system model. This technique is believed to have significant clinical application in dynamic in situ testing of catheter systems. Utilization of this technique may result in higher user confidence in the catheter system. Thesis Supervisor: Title: Dr. Roger G. Mark, M.D., Ph.D. Matsushita Associate Professor of Electrical Engineering in Medicine LIST OF TABLES Table Page 4-1 Tubing resistance measurements.. . . . . . 52 4-2 Tubing compliance measurements.. . . . . . 57 LIST OF FIGURES Figure 1-1 1-2 Page Magnitude and phase response of a typical catheter system . . . . . . . . 8 Waveform distortion due to nonideal frequency response . ...... . . . . . . . . 9 . . 2-1 Modeling the catheter system ....... 2-2 R' and L' as a function of frequency. 24 2-3 Relative error of predicted natural frequency as a function of compliance ratio . .. . . . . . . . . . . . . 34 . . . . . 3-1 Schematic of tap generator.. 4-1 Setup for direct compliance measurement 4-2 Tubing compliance vs. frequency . 4-3 Diagram of laboratory setup for testing of catheter system.. . . . . . . . . . 4-4 4-5 4-7 . .. 58 . . 60 61 Comparison of second-order model with . n functkio transfr Model and experimental transfer functio with a discrete bubble at various locations in the fluid column . . . . . 64 67 Diagram of setup for tap and flush experiments 4-8 48 56 Transfer function for the bubble-free system. . . . . . . . . . . . . . . . . experimentaml 4-6 S. 20 . . . . . . . . . . . . Transfer function with and without a bubble midway in the fluid line . . . 73 74 Figure 4-9 4-10 Page Square wave, tap, and flush time responses for bubble-free system. . ..... Square wave, tap, and flush responses with bubble midway in the fluid line. 5-1 5-2 75 Simulated response to pressure step at input. . ....... . . ....... .. . . . 78 . 87 Simulated response to fast-flush. . ..... 88 CHAPTER 1 INTRODUCTION Present-day invasive accomplished using measurement a site measurement monitoring fluid-filled to an system catheter externally has characteristics, often existence of one or pressure represented more as resonant frequency peaks and well frequency frequency components of the pressure low-pass filtering to yield satisfactory with the phase highest permitting waveform the resonance beyohd. the signal, This nonlinear will usually a the response second-order, equipment, at from transducer. shift. With properly assembled modern occur typically leading located nonideal is simple reproduction while suppressing hig h-frequency artifact. However, a number of factors, most often trapped air in the fluid line, contribute to a low resonant fre quency and subsequent waveform that cannot be corrected by distortion low-pass of filtering. distortion may have serious consequences, since various of the waveform are used in clinical diagnosis. satisfactory frequenc y response before lead to the erroneous and dangerous even catheter This features Measurement insertion assumption frequency response remains satisfactory while the the that system of can the is in use. There is strong evidence to suggest that dynamic changes in the system - clott ing at coalescence of microb ubbles, frequency response. the etc. catheter - can tip, movement seriously alter and the 2 The primary objective of this study is to examine techniques of determining the approximate frequency response of system in vivo (while attached to the patient). We catheter a will examine two direct time-domain techniques for doing-this: (1) The fast-flush technique proposed by Gardner (1970); (2) A flow impulse produced by tapping the with method catheter or extension tubing. We will compare these two excitations a (pressure step at the catheter tip) that cannot be performed in vivo but is an established technique for measuring frequency response. Because the theoretical model we choose to represent the system has a considerable influence on our interpretation of the above stimuli, the other major objective of this study will be to examine transmission-line and lumped-element catheter system. Using independent evidence, we models of will the determine which model is the more accurate representation and how the models may be reconciled. This analysis will aid considerably our understanding of how bubble size and frequency response of the system and the position time affect response two in the to the Invasive blood pressure monitoring of the critical-care and proposed excitations. 1.1 Brief History of Invasive Monitoring post-surgical patient has become medical practice. The visual almost display commonplace of the blood in modern pressure waveform often yields to the clinician valuable information the dynamic state of the cardiovascular system. The monitor sites pressure at a number of important vasculature, such as in the great vessels heart, is often an invaluable and ability tool. to in the of the chambers diagnostic on Long-term monitoring has led to the incorporation of high- and low-pressure alarms into monitoring systems, resulting in faster response of hospital personnel to potentially life-threatening conditions. There are at present two methods in common use for blood pressure monitoring. The catheter-tip relatively new device, prompted by advances and semiconductor technology, which invasive manometer is a in microelectronics consists of strain-gauge transducer located on the tip of a a very, small catheter. These manometers have excellent frequency response characteristics, but suffer the disadvantages of high-cost, extreme high temperature sensitivity. The older and fragility, more consists of an external strain-gauge transducer common coupled and method to the recording site via a hollow fluid-filled catheter, first reported in its modern form by Lambert and Wood (1947). generally constructed of nylon, and is filled system, while having with a polyethylene, a saline significant catheter-tip manometer, suffers from response, which at times makes it PVC, The catheter is woven dacron, or solution. cost advantage relatively inadequate measurement of the blood pressure waveform. This for poor recording over the frequency high-fidelity 1.2 Requirements for Accurate Waveform Reproduction A number of studies have been published which recommend a certain minimum bandwidth for faithful reproduction of the blood pressure waveform. Geddes (1970) these reports, in which it is provides apparent response" depends bot• on the nu:ure degree of accuracy standardized. Bruner required, (1981) and he that of neither summary of "adequate frequency tie waveform of which and the has been has correctly observed that there still little consensus on the :aini:hIu: these systems, a offers an frequency excellent requirements analysis of is of the practical difficulties which have prevented such a consensus from being reached. The pressure waveform has also been subjected fourier analysis to determine the number of harmonics neccessary to achieve a. certain fidelity in a reconstructed waveform. analyses all show the magnitude dropping off rapidly with showed that the the amplitude waveform had fallen to 11.8% of the harmonic of the by the fourier number: components sixth These components Hansen of to (1949) an arterial harmonic; McDonald (1960) found the amplitude of the fifth harmonic from a number of pressure recordings to be less than 20% of the fundamental. These findings tend to support the view that the higher harmonics not contribute significantly to the arterial pressure it should be stressed that limited number of harmonics reconstruction is not of equivalent through a distorting measurement system, since a to the wave, wave from passing do but a it underdamped nature of the system causes nonlinear gain and phase shift within the passband. 5 Hone of these studies directly analyzed the effect of an underdamped second-order system on the pressure waveform. Gardner (1931) has done this, specifying an approximate range, in the form of a chart, of resonant frequencies arid damping coefficients which yield acceptable reproduction of waveforms. This represents a practical if "demanding" incomplete effort defining an acceptable frequency response in terms of the important features of a waveform (particularly pressure at preserving systolic diastolic values). This type of analysis is clearly needed, and along with a clear definition of the waveform features to be preserved, if an objective evaluation of the adequacy of a given *transducing system with known frequency response is to be made. 1.3 The Fluid-Filled Catheter 1.3.1 Models The mechanical properties of fluid-filled which lead to inadequate frequency response great deal of study. Hansen and Warburg as a harmonic oscillator (system with have systems undergone one the degree of freedom), The coefficients of frequency response this model could be shown to be related to the compliance of elements of the system, the physical dimensions of the mechanical lumped-element system consisting of system a of the catheter, and the mass of fluid filling the system. This is analogous to and dashpot in series, or an electrical a (1949) modeled the system generally extending the work of Frank (1903). the second-order equation governing catheter mass, composed a spring, of an inductance, capacitance, and resistance. Tnis early in .odeling work of a number (1963), Shapiro of researchers. and Krovetz Krovetz et al (1974), (1980), was extended by tha experimental .1or Fox et These (1970), al Falsetti (1978), and Yanof et their analysis. Another Latimer (1968), group of the and Li et al basic equations of fluid flow in tubes, study in pulse wave transmission the al but model for Vierhout started derived et detail including (1978), al (1974), Shinozaki fundamental workers, et al who evaluated the frequency response in more retained the second-order system as (1966), include from the originally arterial to system, and developed a transmission line model for the catheter system. This model has been shown to second-order model in be more determining accurate the than location the of simple the first resonant frequency from the physical constants of the system in predicting the presence of higher order since the primary goal of many workers resonances. has been to However, fit observed frequency response in the lower frequency range model, and not to determine the response a and the with priori, a the second-order system has been the more commonly cited model. 1.3.2 Distortion An ideal pressure measurement system is one which frequency response to well beyond interest, and either zero or linear shift corresponds to a pure time the highest phase shift delay). This has flat harmonic (linear of phase guarantees an output which is at worst a time-delayed but otherwise undistorted version of the input. The physical las governing pressure transmission in a long flexible fluid column make this extremely difficult to achieve. The magnitude and phase response of a typical catheter system is shown in Figure 1-1. There is present a large resonant peak to which will amplify harmonics that lie close the resonance, and a subsequent falling off of the frequency response following the resonance which will attenuate higher harmonics. Moreover, the phase shift is highly nonlinear, with a sudden degree phase shift near the frequency response is resonant typical of frequency. This underdamped 180 type of second-order systems. Figure 1-2 illustrates the effect this has on a simulated arterial pressure type waveform. shows the input to the catheter, while Figure output from tne system. The distortion resulting from the recording system, output signal nonideal particularly the of distortion Figure 1-2(b) 1-2(a) shows exhibits serious characteristics appearance of the of the spurious oscillations and large error in systolic pressure. While it is theoretically possible signal given the output waveform and the to recover frequency the response the system, the latter is generally unknown. Therefore, deal of work has gone into measuring the frequency input a of great response of catheter systems, either to perform this reconstruction or simply to evaluate the adequacy of the system. 38 r IH(f) 3 4a feO) -1' ~NI %ill a- Figure 1-1. pFmSg . Magnitude and phase response of a typical catheter system BED I V'~+Le& 277j1 *-T-' 81 0 x. -- ---t -. -- ti-i H- r - I .' EBED 00022, r T --- - t -!r- -. 7. r 17 : K- : I-r ti V-. :F:i-- - -- iIt input blood pressure waveform (a) -•- otu of . cathetr-t.ansd ! ---.. --- s yste ·-- i iT..: - -i;i'' .. : .: ; - i-i::i~li~l-ii-_::1;i·-_i~-: _ j,~::,::-;i- .7 r V .: :::...~. :l-~_i-~~-~~--i-, :-i27? LIi· output of catheter-transducer system (b) Figure 1-2. Waveform distortion due to nonideal frequency response 10 1.4 Frequency Response Measurement Techniques A number of techniques hive been used i~duce to the frequency response of catheter systems. These techniques can separated into two classes: involve (1) direct techniques, which be analyzing the system response to a known external input; and (2) indirect techniques, which assume the a particular model for catheter (typically second-order) and make additional assumptions about the frequency content of class, the techniques can the be input signal. further Within divided each into time, frequency, and correlation-domain approaches. We will examine how each of these techniques has been used. 1.4.1 Direct Techniques Frequency-domain techniques all require input having a known spectrum. If the system of is a signal linear and time-invariant, the energy at each frequency in the output signal is uniquely associated with the energy in the input signal that frequency, so an input which is flat in frequency and at phase will produce an output which is a scaled version of the transfer function. The impulse function and white noise both are flat frequency and phase, but practical considerations noise the better choice for frequency-domain is no evidence of white-noise excitation direct frequency-response measurement make white measurement. There having in possibly because other methods exist which been catheter do data to be analyzed directly in the frequency requiring less sophisticated instrumentation. in not used systems, require domain, for the therefore 11 An example of one such alternate method is gain and phase shift at a number of technique was used quite (1968). discrete successfully Swept frequency measurement by is (1980), and Gardner (1981). measurements described thus source substituted for the frequencies. and standard used by an require appropriately artery requirement is difficult to meet at in the the This engineering all that the Latimer Rothe Unfortunately, far removed from the patient and measure Latimer a technique which has been recently been to the and Kim of the system driven catheter clinical be pressure tip. This environment. Therefore it is not surprising that these methods have primarily been used in experimental work. Direct time-domain techniques for response all consist of exciting measuring the system the frequency with well-described time signal, usually a step function a of simple pressure ("pop" excitation). The pressure step is achieved by pressurizing a closed system and then suddenly relieving the pressure catheter tip, typically response observed at the by bursting output a of rubber the at the membrane. The system can analyzed. In order to make the analysis of the response straightforward, priori. While the most system researchers system, some (Melbin and Spohr, 1974) have responses demonstrated to the order generally have assumed 1969; Gabe, systems pressure is which step, then waveform described a although a second-order 1972; Krovetz et exhibit be al, higher-order only specifically suggests that reflections from impedance Krovetz mismatches (i.e. transmission-line phenomena) may be responsible. One source (Attinger, 1969) reports that 30% of a large number of systems tested exhibited higher than second-order behavior. However, second-order approximation often - is satisfactory, assumption makes the waveform analysis particularly the and this simple (see chosen, the Appendix A). Even if a satisfactory model for the system is "pop" technique remains unsuitable for dynamic analysis of catheter in vivo, since it requires catheter tip that the available. Because catheter systems can collect clotted, or otherwise be degraded in use, a bubbles, the become satisfactory test result from a pop test performed prior to insertion may false sense of security. This q'uestion concern raises the be give of whether excitations applied elsewhere in the system may elicit response from which the appropriate transfer function may a a be deduced. There have been no studies done to answer this question, although Gardner (1981) has described a technique which is created by claimed to be an acceptable excitation: a flow step opening and then present releasing the monitoring setups, where the flush flush valve source is in located most at the transducer end of the fluid column. 1.4.2 Indirect Techniques A number of indirect distortion-measuring techniques have been described. These all assume a second-order model and attempt to determine, from the patient location and magnitude of the examined the magnitude pressure resonance. of the signal Brower signal itself, et the al (1975) spectrum after preconditioning (bandpass filtering and differentiation). presence of a peak in the preconditioned spectra with distortion, and approximate formulae given the degree of damping. components of the Doherty incoming (1981) pressure normalized magnitude and phase in a was associated for determining determined signal, the and regression Fourier applied equation coefficients were optimized to detect resonance within a critical range. Jackson et al (1978) used linear of a complex certain range of frequencies was taken to indicate whose predictive pair pole the certain analysis (a correlation technique) to model the spectrum input signal: the presence The zt of the witnin a presence of resonance. These indirect techniques all evolved because of the need to perform dynamic analysis on the catheter system and to eliminate the need for manual intervention by hospital personnel, either in testing the system or in compensating the response. These advantages over direct methods make indirect techniques extremely desirable. On the other hand, the validity of these indirect techniques rely heavily on both the assumed catheter system model and on the assumed spectrum of the patient waveform. If the blood pressure power spectrum appears "resonant" in the sense of having a local maximum, as may happen in recordings from the smaller arteries (the arterial system itself behaves as an assemblage branched transmission lines), then these techniques erroneous results. An additional problem, although one grown smaller as the cost of computation has computational complexity of indirect analysis. decreased, Although can of give that has is the initial 14 experiments using these techniques appear promising, they Iave not yet been subjected to extensive testing using the full range of clinically observed waveforms. 1.5 Compensation Techniques Various approaches have been taken to compensate the frequency response of catheter systems. These can be divided into two classes, mechanical and electrical. Mechanical compensation may be considered a problem in impedance matching, although researchers regard it as merely increasing the some damping coefficient. Electrical compensation involves active filtering of the signal. We will now examine each of these methods in more detail. 1.5.1 Mechanical Compensation By adding additional damping to the hydraulics, system which previously produced highly distorted a catheter waveforms be made to have a much wider useful bandwidth. Damping by a constriction at the patient end of the catheter has for many years. A set of (1974), experiments by LaPointe adding been and can used Roberge using needle valves as resistance elements, has confirmed the utility of the technique, and Latimer (1968) has justified it in terms of matching the source and line impedance of an acoustic transmission line. Unfortunately, the large amount of constriction neccessary to achieve appreciable damping makes this technique unsuitable for use with flush devices, and it is rather sensitive as well. A more promising technique is parallel damping, reported by van der Tweel (1957) and Crul (1962). This can be described as another form of impedance matching, this time matching the load (transducer) impedance to the There are commercial devices now available damping. One which we have observed is which is placed in transducer. Gardner parallel (1981) the with the valve) perform Sorenson fluid in series parallel Accunamic, line order to with create parallel impedance. The impedance match obtained adjusting this device to impedance. at the has described this device, in which fixed compliance (bubble) is placed in resistance (needle to line minimize step a an by response a variable adjustable empirically overshoot is An advantage of these mechanical compensation techniques is crude but nonetheless fairly effective. that, by effectively increasing the damping to flatten out the resonant peak, they tend to extend the useful range of the system out to approximately the resonant frequency. This can amount to twice the usable bandwidth of the uncompensated system. Moreover, no exotic electronics or processing techniques are required. The chief disadvantage of mechanical compensation is that it requires an external step input to observe when critical achieved. Relying on the patient waveform to adjust damping the is damping is a dubious procedure at best. 1.5.2 Electrical Compensation If an approximate transfer function for the catheter system is known, inverse filtering (convolving the output signal with network having the inverse transfer function) can greatly a extend 16 the bandwidth. Melbin and Spohr (1969) describe an analog circuit to perform inverse filtering. M1ore recently, Brower et al (1975) and Ciccolella (1976) have described digital filtering to perform the same function. If the transfer function approach is to low-pass is not known, filter the signal, the most common with a high enough to retain the significant harmonics of the pressure signal. This is frequency lower than the assumed resonance but cut-off the method most manufacturers include in the monitors at present. Frequently, simultaneously however, - the these resonant conditions peak cannot overlaps be an met appreciable portion of the signal spectrum. Aggressive low-pass filtering (12 Hz cutoff and below) has been practiced by some manufacturers an attempt to prevent resonances occuring at higher from causing systolic overshoot, but at cost the in frequencies of extremely limited bandwidth. Low-pass filtering can be useful in preventing high-frequency artifact from appearing in the output signal, it is extremely limited in its having low resonant frequencies. ability to compensate but systems 17 CHAPTEi 2 MODELING OF THE CATHETER SYSTEM A theoretical understanding of the transducing system is an important step predict the changes in system blood towards pressure being characteristics able to under various conditions (e.g. altering of component stiffness, tubing length, the presence of occult bubbles or leaks). system model, it is possible to Under specify an the appropriate most desirable characteristics for catheter, tubing, and transducer in terms of producing a faithful reproduction of the In pressure waveform. addition, an accurate model for the system may suggest methods of compensation of the frequency response involving additional components, such as impedance matching devices. This section will develop a general model for well-established theory of the wave transducing propagation system in lines, and then proceed to establish conditions under using the transmission which model may be simplified to a lower order lumped-parameter the system with little loss in accuracy and large gain in ease of analysis. 2.1 General Model - Mechanical/Electrical Analogies There can be little doubt as to the validity of a transmission line model for the fluid-filled pressure tubing. The presence of phase delay, attenuation, 'and acoustic impedance have been experimentally demonstrated by many researchers, but pernaps most elegantly by Latimer and Latimer (1969), who determined values for wave speed and attenuation at a number of resonant and antiresonant frequencies. It is interesting to that the tubes was rather for pulse wave transmission in the arterial tree. The principles are theory of acoustic wave originally developed not for transmission catheter in elastic systems largely the same but the catheter system is note in but fact easier to analyze due to the limited number of reflecting sites, consistent internal diameters and better-understood wall properties. Figure 2-1(a) shows a physical model for the of transducer system, represented by codstant internal diameter coupled to a simplest liquid-filled a transducer type tube of through fluid-filled dome. An increase in.pressure initiated at the a left causes liquid to flow to the right through the tubing and which in turn causes a deflection of diaphragm. This deflection is sensed by a strain electrical signal is amplified and the transducer gauge and processed the to dome, resulting produce a shown in pressure recording. An electrical model for this mechanical system is Figure 2-1(b). The tubing and transducer dome/diaphragm will each be examined in turn. In each infinitesimally long segment of the tubing, fluid motion has associated with it friction due to shear stresses in the fluid and inertia due to the mass and velocity of the fluid. There are also compliances associated with the wall and (to a lesser extent) in the storage of potential energy. fluid Finally, itself, the wall tubing leading to exhibits viscoelasticity which causes energy losses due mechanical to hysteresis effects. These physical "line constants" are replaced by the analogous electrical symbols in Figure 2-1(b), where: R' = resistance/unit length due to viscosity of the fluid; L' =.inertance/unit length due to the mass of the fluid; C' = compliance/unit length due to compressibility of the fluid and stretch of the tube walls; G' = conductance/unit length due to energy loss in the tube walls; dx = incremental length. These so-called "line constants" are in fact all frequency dependent to some extent as the theoretical development next section will show. The transducer dome and exhibit similar-. resistive, inertial, wall of diaphragm loss, and treating also elastic effects, but the short length of the transducer relative wavelengths we will be considering justifies the to the these as lumped elements. In summary, then, we have inertial and resistive effects associated with the fluid. These present a longitudinal impedance to flow and thus are represented as series elements. The compliance and wall loss are properties of the plastic materials used, present a represented as transverse parallel infinitesimal sections characterized by the impedance elements. then to An produces telegraph flow, and thus infinite sum of these transmission line a equations engineering theory. Prior to determining the of frequency are electrical response of this model, however, we will need to develop the equations calculate val-ues for the line constants. to dome d i annbrqan ''~ " DI plastic tubing Pin transducer (a) R' dx L'dx R'dx L'dx R'dx dx L'dx G'dx (b) Figure 2-1. Modeling the catheter system Rtr C'dx Ltr Ctr 21 2.2 Theoretical Calculation of Line Constants 2.2.1 Longiud nal Ipedance The development of (R and L) theory describing laminar oscillat.ory fluid flow through narrow tubes has been made by Lambossy and Womersley (1956). The most significant result of this (1956) .theory is the prediction of a "skin effect" phenomenon, which causes increase in resistance and decrease in inertance at an high frequency due to alteration of the fluid velocity profile. across the tubing cross-section. If the tubing is assumed to be rigid, straight, and of circular cross-section Womersley shows that Q= r j - ej (1) where Q = volume flow A = amplitude W = circular frequency = 2rf Aejwt= pressure gradient = - .p dp/dx = fluid density a = r1/TU u = p/p = kinematic viscosity = Womersley coefficient (dimensionless) J0 and J1 are the zero and first-order Bessel functions of the tubing are complex argument j3/2 ; j = T = phase shift of Tr/4 radians Although the requirements of rigid and straight 22 not szric;ly met by 2ctheter systems, many authors (Latimer, Jager) have applied these equations to calculate the longitudinal impedance per unit length Z' with good results. Following the analysis of Jager et al (1965): Z' = dp/dx = j 2J1 {j Saj r(Jooj 3/2j 3/2} o (2a) 3/2 if we write this as ZV 2. j L'(w) + R'(w) (2b) where L' and R' denote resistance/unit length and inertance/unit length, respectively, then P 2 S7Tr M{ 1- 2 0a _ 2J ; aj 312Jo MI'o= modulus 1 For the case familiar of - steady ~1o= phase flow these sinle irr4Mio S1i - (3a,b) 2J1 j 31/2Jo0 equations reduce to the Poiseuille equations for R' and L': L = 4/3 W _8 11 R' W 7r (4) P2 7r The significance of the W'omersley calculations for R' and L' is that effective as the frequency inertance decrea ses increases. In the limit drops to 75% of its of of d.c. oscillation and the infinite value, and is increased, effective frequency, the the resistance the resistance inertance becomes 23 infinite. A plot of dimensions used in R'(w) this and study L'(Q) is vs. shown w in for the Figure tubing 2-2. The viscosity and density of saline have been taken as 0.01 poise 0.998 gm/cm 3 respectively (approximate values at 20 nsr degrees centigrade). 2.2.2 Transverse impedance (C and G) A calculation of compliance and wall loss requires detailed knowledge both of the physical properties of the plastic used for the pressure tubing and the mode(s) of wave propagation fluid and in the wall. These are quite generally specify precisely. Equations exist which specify in the difficult the of a tube of uniform cross-section as a function of the to compliance internal and external radii, Young's modulus and the Poisson ratio for the tubing material. These assume a linear, isotropic medium with no losses, but may be applied to plastics so long as the results are not expected to be quantitatively precise. Using the equation for the compliance of a thick-walled tube (see reference 25) we have: C' : 2nr 1+r /r 2 E 1-rf/r 2 ,,, i + s (5) e where r. = internal diameter; re = external diameter; E = Young's modulus; s = Poisson's ratio. No such simple equation is known describing the wall loss G, and 24 'i SL'(Pa-sec/m x10, 4) Ii 12 12 l,,,.r L'(0) c S_ i X1 0 'O R' 5 (Pa-sec/m4 ) - m 3 3 ( R'(O) I 11 Ii3 41 M 7 Ii11 Figure 2-2, R' and L' as a function of frequency 11M vw1 in any case the properties of a given sample of plastic are generally not so well specified as to allow direc; calculation of compliance and wall loss. In other studies similar to this one, experimental measurements are invariably substituted for in the calculation of transverse (1949) impedance. Hansen theory has found that G' is proportional to frequency, implying a m3chanical hysteresis loss per cycle, but it is not known whether this generally true for plastic materials. More research in this is clearly needed but is beyond the scope of the We will use experimentally models, and assume G'=O. determined As long as present values for G'/wC' is C' is area study. in small, our this assumption should cause negligible error in predicted location of resonant frequency. 2.3 Transmission Line Formulation 2.3.1 Telegraph Equations and Propagation Constant We will take the following circuit representation to be an adequate model of the transducing system: Ps s Transmission Line k=propagation constant Z=characteristic impedance Ctr • Rtransducer= 0 L transducer-O o0 where the parameters Rtr and Ltr have been set to zero. This is justified when the transducer dome radius (or effective radius iin the case of a non-cylindrical dome) is much larger than the tubing radius, due to the strong inverse dependence of both R and L on the radius. This condition is nearly always met in catheter systems. The telegraph equations governing this line are: dP/dx = -(R'+jwL')Q Q = -(G'+jwC')P where P represents pressure solution which is the waves: (6a) sum (6b) and of Q is forward flow. and If we backward assume a traveling 27 P(x,t) = P+e -kx + Pe +kx }e jwt (7) then we can solve for k: k = = V(R'+jwL')(G'+jwC') + j : m/LvV' C'/(V+j)(W+j) ( (8) where S= attenuation constant B = phase constant :WVL"C' 0o V = R'/wL' W = G'/wC' squaring the above equation: k2 =c 2 _ +2jaB -2 = a21(VW-1)+j(V+W) (9) separating real and imaginary terms and solving, we have: a = 0.5F5BV+W) (10) a = Bo/F (11) where (12) T{(1+V1) (1+Wz)+I-VW} F is the correction factor, important primarily at very frequencies, in the form derived by Latimer and Latimer low (1968). This correction factor will rarely need to be used in this study, since by the first resonant frequency F will nearly reached its high-frequency limit of 1. It is, have always however, theoretical interest because if the condition V = W occurs of (i.e. 28 then F = 1 and a Heaviside "distorotionless line" is R'/L'=G'/C'), obtained, with constant attenuation and phase shift proportional to frequency. This condition, unfortunately, is never encountered in practice except at isolated frequencies, because R', L', G', and C' are all frequency dependent to some extent. 2.3.2 Characteristic impedance The characteristic impedance Z of the transmission line is defined as Z - Q dP/dx where Q is the volume rate of flow and dP/dx the pressure differential. Z can be expressed in terms of V and W as follows: (13) (R'+jwL') (G'+j wC') SL'iC'/(V+j)/(W+j) = IZI Z je (14) 2 +1)}.2s (15) e = 0.5(cot-1V - cot-1W) (16) Zo = Zo{(V 2 +1)/(W -L'/C' alternatively, Z may be expressed in terms of resistance reactance: Z = :Z:(cose + jsine ) 2.3.3 Boundary Conditions and Reflection Coefficient The solution for P(x,t) and Q(x,t) can be expressed as (17) and 29 P(x) = P+e- xe j- Q(x) = -- {P+e -cxe -jIx - Pe +cax e+jix } e + Pe (13a) 1 (18b) Z where the dropped have we equations. The parameters a and length and phase shift per dependence ejw t time from the 8 denote the attenuation per unit unit length respectively of the transmission line. To solve equation (18) we need to specify the conditions at each end of the transmission line. We have a zero-impedance pressure source and represented as a pure compliance. The a load (the boundary assumed transducer) conditions boundary are therefore P(x=O) = P + P = PO Q(x=l) Pefining (19a) = jwCt{P +e -k+Pe+kI the reflection = Z{P e-k; -Pe coefficient +k r =P+/P_ } (19b) and solving equation (19): P+=Po -21 1+fe 21 ; P =P +Fe-2k £ and P(x) = +e-21va!- (e 1+re) -kx + 'e-2kekX) (20) where 1-jWC r= tr Z (21) 1+j CtrZ At the transducer, the pressure is Poe-kk - P(k) = l+re - 2k (1+r) (22) I a relation whicn cescribes the attenuation and phase shift of the pressure wave from input to output as a function of frequency. 2.3.4 Natural Frequencies The transmission line equations (18a,b) may also for the natural resonant frequencies of P(x=O) = 0. Equation (19) the system be by solved setting then becomes Z-tankt = -I/jwCt r (23) a transcendental equation in complex Z and solved in closed form. However, we can find k wh-ich cannot approximate be values for the natural frequencies by making the assumption w>>R'/L' and w>>G'/C' (generally valid at and above the first frequency), allowing us to make the approximations k Z Lc' wCtr resonant jwVL'C' in which case equation (23) becomes r -L'/C r 'C1) i4 which can be solved graphically or numerically frequencies wn" (24) for the natural 31 2.4 Lumped Model Approximation While the transmission line model in theory is probably system, most accurate representation of the catheter several practical reasons why lumped-parameter the there are are more models commonly invoked to explain the resonance phenomenon. First among these is the fact that higher-order (i.e. three-quarter-wave above) resonances almost invariably occur at a frequency the range of significant blood pressure harmonics, and beyond making their presence inconsequential. A second factor tending to minimize the importance of the higher resonances is the Womersley which causes the resistance (and therefore dampinS) markedly with frequency, thus minimizing the to effect, increase height resonant peaks. Thirdly, the presence of trapped air of the bubbles (a common circumstance) introduces large lumped compliances into the system, tending to make a lumped circuit representation more tractable than the corresponding transmission line model. A simple way to approximate the transmission line model with lumped-elements is to take the circuit representation of 2-1(b) but let each section represent a finite length of rather than taking the limit dx->O. Li, Van Figure tubing, Brummelen, Noordergraaf (1978) have performed this analysis for N=1,2, and and infinity, where N is the number of lumped sections. These authors also considered the effect of different element ( 7 vs. inverted-L) on the calculated frequency findings indicate that sizable errors in the configurations response. location Their of the first resonant frequency are introduced for N=1, but that as the 32 lengrth of each section becomnes small wavelength (highest frequency) becomes v.ry good. of relative interest, They also found the to the the shortest approximation configuration, 7 involves slightly less lumping per section, to be more which accurate for a given N than the inverted-L model. We will adopt a slightly different tack in this study. First, will examine the two limiting cases Ctu<<C tu tr and Ctu>>C tu»tri we and demonstrate how each can be represented by a second-order circui; with appropriate correction factors. Then, a simple equation will be constructed which is precise for the introduces only a small calculable error in limiting cases resonant and frequency for intermediate cases where neither compliance dominates. Ctu << Ctr In this case the tubing is rigid compared with the transducer, and the wavespeed is so high that propagation effects may be ignored. A second-order equation is therefore valid, with W0 tr = i/'iLCt (25) Ctu >> Ctr In this case the transducer looks like an open circuit, we may solve equation (24) Wm Ctu = 7/2LC- and for Ctr=O: (26) tu Ctr Let us construct the following equation which will satisfy 33 both limiting cases: WU (27a) 1 = /L((2/w) /LC eq aD .where r Ctu+Ctr) tu tr (27b) z : •Jl-~D Ceq = Ctr + (2/r) (27c) 2Ct u For Ctu(<Ctr: wo -> I/C/L-C tr and for Ctu>>Ctr -> 2"o tu as desired. 4e can separately solve equations (24) and (27) for intermediate values of Ctr. If we define K then we can frequency (1 - plot Ctr/Ctu the wo )/wo relative error as a function of in K. predicted This is natural shown in Figure 2-3. The maximum error is only 2.44%, so the approximation introduced by equation (27(a)) seems acceptable. If the values of Ctr be and Cu are known, F.igure 2-3 can precise used to Thus we have succeeded in reducing the transmission line to determine a correction factor for equation (27(a)). a.0 2.5 g 1.5 1.0 .5 as f izz .5 1.08 1.5 2. 2.5 < = Ctr/Ctu Figure 2-3. Relative error in predicted natural frequency as; a function of compliance ratio K a simple second-order circuit in terms of preserving the location of the first resonant frequency. Equation (27) implies an times the 4/r2 equivalent lumped compliance ;where a fraction total tubing compliance shunts the transducer: Rtu S ;i;i H(jQ) tu7 T r s- r-rnsfer Ltu (4/7r2-)C c tr Ceq eq 4 , Cu + Cr tu tr function = Po(jw) = Pi(jw) 1/(LCeq) -w (28) 2 +(R/L)jw+(1/LCeq) If the resonant frequency of a catheter system is calculated using this lumped model, a useful check on the legitimacy of the assumptions used in constructing the model is tthe to calculate loss term R'(w r )/L'(w r ) to verify the assumption w >>R'/L' (we still assume Dividing equation (3b) by (3a): R'(w) u - pr L'(w) a'tanelu 2 (29) so the inequality Ar >> pr z a 2 tane 'o 0 becomes our check on the model assumptions. (30) G'=0). I 36 .Air Bubbles 2.5 Effect of Trappe It has previously been noted that the presence fluid line remains tne single most coinmon of c-.use pressure monitoring. This is due to the high air of in the low-quality compressibility of air relative to water, causing even very small bubbles to greatly increase the total compliance of the system and thereby the resonant frequency. The problem of including air reduce bubbles our models is exacerbated by the unpredictability of bubble in size and location in the clinical setup, and by the strong dependence of bubble compliance on temperature and pressure. The alteration of the normal fluid velocity profile in the vicinity of a bubble may model. also violate Nevertheless, we the can plane-wave examine assumption specific of our situations which are amenable to straightforward analysis and thereby possibly develop some intuition towards the more general situation. 2.5.1 Compliance of Air Bubbles The compressibility of air depends on temperature, pressure, and molar quantity. The compliance of an air bubble may therefore be expected to vary as pressure waves are propagated in the fluid line. To specify the variation precisely, we need thermodynamic state of the bubble at all times vs. adiabatic compression cycle). We also (e.g. need temperature variation of air solubility in water time constants to know whether pumping of air to know isothermal to and into solution with pressure variation is significant. In the know the associated and this out of study, we will be content to note the primary effect of static pressure on compliance and ignore all higher-order effects. If air is assumed to be a ideal gas, then PV = nRT. For a bubble, we will assume n and T are constant. Then PV = K V or K K (31) If the pressure changes by an infinitesimal amount dP, then the volume changes by an amount dV, with the relation PV = K = (P+dP)(V+dV) = PV + VdP + PdV + dPdV Cancelling like terms and ignoring the higher-order term dPdV, we have dV/dP= -V/P (32) Using the definition of compliance: C = -(dV/dP) and using equations (31) C = and (32), this becomes K/P 2 (33) so we see that the compliance is a strong nonlinear pressure. If the pressure excursions inside range from 700 mmHg to 1100 mmHg (-60 to +340 atmospheric pressure), between 46% and then the bubble the function fluid mmHg may to vary 118% of its value at atmospheric pressure. There are several lessons to be learned from this First, the inclusion of bubbles in the fluid column significant nonlinearities in the frequency response pressure column relative compliance of excursions are present. Second, at analysis. may cause when large higher -static compliance pressures the effective bubble therefore the resonant pressure is increased. Henry increase may frequency et al as (1967), phenomenon, have even suggested checking the detecting bubbles in the fluid column. as if a for the bubble, considerably simplifying analysis. will take in constant 2nalyzing this response means a the excursions are kept small, we may assume third approach which we static the frequency Third, and noting pressures of the system at high and low static smaller becomes of pressure compliance It is tuis systems with bubbles in Chapter 4. Published values of air compressibility temperatures and pressures exist. At twenty at degrees various centigrade and 760 mmHg: dV/dP 1.0126x10 - 5 Pa - 1 and C = V dV/dP (m 3 /Pa) (34) This is the equation we shall use to calculate bubble compliance. 2.5.2 Relationship between Bubble Location and Resonant Frequency We will now consider how to include a bubble of known volume and location in our lumped element model. The inertance and resistance of the bubble will be taken as negligible, and we will assume therefore that the bubble may be modeled compliance shunting the transmission line (we do distinction between bubbles clinging to the wall as not of a lumped make the any tubing and bubbles completely occluding reason to do so). the lumen, although We may now view the system as consisting of two transmission lines in series: the first terminating compliance there may be (the bubble) and compliance (the transducer). the second in in a another lumped lumped The analysis of section 2.4 may now be applied, yielding the following model: (Z-x) Ctu PS where x=distanze from source to bubble and 2 is the total length of the system. The transfer function governing this circuit is H(jw) = A 4-jBw'-Cw 2 +jDw+l a fourth-order equation in w, where A = x(Z-x)CIC 2 L2 B = 2x(X-x)RLC1 C2 C = [x(x-x)R 2 CIC +L(C +xC 2 2 1 )] D = R(xC 1 + C ) 2 c = ( 4 /r 2 )xCtu+Cbu C2 (4/r 2 )(-x) Ctu+Ctr (35) 40 If we assume that Cbu is much larger than Ctu or Ctr, zhn C,>> and the two second-order circuits tend to be decoupled, resonate independently. The resonant section is approximately equal to the bubble is advanced in (increasing x) the primary importance of this result the tubing the of towards the frequency that located in the catheter will not degrade the system as much as a bubble located the up first in as transducer decreases. recognition farther T.h t1:ey , so we see that 1//xLCI resonant is frequency i.e. a The bubble performance system. As a practical aside, we note that when there is suspicion of a bubble causing a low resonant frequency, the begin at the transducer dome and then the catheter. search proceed should generally backwards toward 41 CHAPTER 3 MATERIALS All experiments carried out in utilized a single brand of tVDes of the pressure course extension high-quality catheterization components laboratories as and addition to these components, a by intensive number of interchangeably. from Because of several the they represent many hospital care units. additional manufacturers relative stiffness plastics used in these valves, their wide bore, and contribution considered to to the overall significantly system affect length, the two partly system (three were used of the their they In elements were required in the experimental setup. Hydraulic valves and four-way stopcocks) study and chosen because used This tubing ressure tr nsducer. These elements were on the basis of availability and partly typical of small were response. not A pressurized IV bag and standard fluid were used to flush and fill the hydraulic system. Several different pressure sources and flow sources were used as test 'inputs. A pressure amplifier, CRT- display, tape and strip chart recorders, spectrum converter, and computer system dynamics. All facilities of these described in more detail below. were materials used and analyzer, D/A analyze the components are to 3.1 Extension Tubing The tubing used in monitoring kit (HP No. this study was obtainec 14233A) mark:3ed by tubing is constructed of translucent, from a pressure iiew1eýt- Pckard. high-density polyethyle.ne, with an internal diameter of 1.18 mm, outer diameter of 1.93 and length of four feet (1.22 meter). The ends are supplied one male and one female luer fitting. No specifications are available characteristic compliance is measured tubing is relatively stiff compared manufacturer this for in to market. A previous study by Gardner (1981) for commercial pressure tubing lists a section the 4.1.3. This brands of with but of static range mnm, technical tubing, similar The on the compliances 1.6 mm3 /100 mm Hg for six feet of tubing, with an average to 17.1 compliance of 7.4 mm 3 /100 mm Hg/6 ft. In comparison, the measured compliance of the H-P tubing in these units is 2.72 mm 3 /100 mm Hg/6 ft under static conditions and 0.34 mm3 /100 mmHg/6 ft at high frequency. 3.2 Transducers Two transducers were used in these experiments. The first, a Bentley Trantec Model 800, was generally used as the primary test element in the transducing system because it could be modeled a pure compliance over a wide second, a Hewlett-Packard range 1290A, of static exhibited phenomenon at lower static pressures which was pressures. a fluid probably movement of fluid trapped between the plastic diaphragm quartz transducing element. Since this phenomenon as The leakage due and to the manifests 43 itself as a frequency-dependenc compliance at low frequencies), primary transducer compliance (largor apparent the 1290A proved undesirable for thas. experiments. Instead, used as a reference transducer to monitor input to the extension tubing. In this the as a 129a A tha pressure application, was at the where the transducer effectively shunts the pressure source, compliance not an important issue. Specifications for the transducers list maximum compliances of mmHg, 0.04 Bentley and is and H-P 3 /100 mmn 0.15 respectively. 3.3 Flush Bag and Fluid As in a typical hospital setup, the equipped with a means of standard IV bag was filling filled with flushing and a monitoring standard 3ystem with was fluid. fluid A (described below) and pressurized to between 200 and 300 mmHg by means of an inflatable bag holder. A length of pltssic large-diameter and three-way valve connects the flush bag to the rest tube of the system. The standard solution consisted of debubbled water, prepared by vigorous boiling. A small quantity of soap solution was as a wetting agent and all excess air was purged from added the before pressurizing, thus ensuring a minimum of dissolved air the fluid. These precautions, coupled with slow, careful of the hydraulic system, were found to be of of fluid or in filling the resulted in high values of compliance. This system solution, in filling the utmost importance in excluding air from the system. Lack of care in preparation bag the invariably developed 44 after noting tha I:as i4t virtu~aly impossible to completely eliminate trapped air from the system aft-er filling with was not believed to differ significantly from saline normally used in terms of its the saline, physiological inertial and viscous properties. 3.4 Slow/Fast Flush Unit A Sorenson Intraflo flush element was experiment involving the system response to Intraflo and similar units provide a included a high fast for the flush. The impedance channel between the flush bag and the catheter, in an attempt to keep the catheter tip free from blood clots. A parallel low-impedance channel can be opened manually to provide a large bolus of for the same purpose resulting from the (fast release flush). (closure) The damped of this fluid oscillations valve we.re of interest in this study. 3.5 Bench Equipment Two H-P pressure excitation to and amplifiers processed the (model 78503C) resultant provided signals the from the transducers. The amplifiers were modified for this experiment provide a flat bandwidth out to measurements to be made beyond the 100 Hz, normal allowing 12 Hz tubing was provided by, a blood pressure simulator 601). This simulator features a square-wave frequency bandwidth these amplifiers. The reference input excitation to the of pressure (Biotek output to for Model step response measurements, nine selectable pressure waveforms (stored in read-only memory), manual controls, and provision for an systolic external and diastolic electrical simulator dome has two ports which connect to the level input. fluid The column via luer fittings. Frequency measurements were made by applying a source to the external input jack of the white noise approximate was amplitude small-signal limited Biotek to excitation. noise simulator. 10 The white mmHg The RMS to pressure transfer function (magnitude and phase measurement) between the reference and test transducers was determined Spectrum Analyzer. If desired, the stored on an H-P Model 3960 using an amplifier Instrumentation visual record produced on a H-P Model 78172A H-P Model outputs could Recorder Chart 3582A be and/or a Recorder. A diagram of the setup is shown in Figure 4-3. Computational facilities (for modeling primarily of an H-P Model 85 desktop purposes) consisted computer, which has hardwired BASIC as a programming language. The H-P 85 also has an interface bus which allowed it to be used (in conjunction with digital plotter) to accept spectrum analyzer and digital X-Y plotter. to spectral produce For and time high-quality transient data from a the graphics on the circuit analysis, a simulation program SPICE was run on an H-P 3000 series computer. 3.6 Flow Source For direct measurements of tubing compliance as a function of frequency, a specially designed sinusoidal flow generator used. The flow source consisted of a commercial was microliter 46 syringe (Hamilton model 70011) plunger and is accurate which utilizes a tungsten wi re to 0.01 microliter. A custom-madc adapter allowed the syringe to be tightly coupled to the as luer system under test. The syringe barrel was rigidly mointed in an aluminum block. The plunger was coupled to a dc servo motor through a mechanical linkage, bearing, and eccentric cam. This arrangement produced sinusoidal motion of the plunger. be altered between 0 and 0.8 The stroke cf the syringe could microliter by varying eccentricity of the cam with a linear feed screw, The maximum frequency was conservatively set adequate plunger needle and barrel, to limit but the and ;i frequency range of 0.1-14 Hz could be obtained by varying the motor between the a this speed. friction range was for our purposes. 3.7 Tap Generator In order to test the system response to a manually impulse, a crude form of a "mousetrap" tapper built. A schematic representation Figure 3-1. The tapper consists of of the a was tapper platform applied designed is on and shown which in the pressure tubing is secured by two clamps into a milled channel. A tubular steel spring is spring arm passes manually through its retracted resting and position, 'tubing (causing a flow impulse as fluid is deforms the repeated considered a potential problem, the cylindrical a The displaced) returns to rest. Because tubing failure from wrapped with plastic tape to provide released. small spring amount and then taps was arm was of shock 47 absorption. 48 top view latform spring tubing lip plastic side view ~ Figure 3-1. "Mousetrap" tap generator 49 CHAPTER 4 METHODS AIND RESULTS 4.1. Determination of Line Parameters In order to determine the most transducing system, it physical parameters is first which appropriate neccessary best tnhese parameters can then be to characterize components of the system. The measured or inserted for the determine the model the separate calculated the into examination, allowing comparison between the values model frequency under response of the model and of the experimental, system. In some cases fluid inertance L) there was no convenient of technique (e.g. available for direct measurement, and some theoretical calculations were of neccessity substituted for direct (e.g. flow resistance flow) could be R) the measured observation. at value but the zero a.c. In other frequency value could estimated by reference to pulsatile flow theory. The use unavailable, would have been of of a flow transducer, considerable (steady only be theoretical and experimental values for each parameter are compared possible. Clearly, the cases whenever which benefit in was this study, since it would have allowed simultaneous a.c. measurements of / pressure and flow. Nevertheless, the d.c. measurements obtained represent a positive step towards quantitative analysis and verification of pressure monitoring system. the theoretical models for the 50 4.1.1 Resistance Theoretical Calculating a theoretical value for Rtu actually specifying the frequency of interest, has shown the flow resistance to be (see Fig. 2-2). However, at zero Womersley since a requires function frequency (1956) of -frequency Poiseuille's Law applies, and one can write tu - Taking: 8nr 4 rn=0.001 Pa-sec (approximate for water at 20 degrees C) Z.=1.22 meter r=0.59 mm We calculate Rtu= 2.56x10 1 0 Pa-s/m'. Experimental To measure Rt, a calibrated pressure source was to one end of the column. The rate of flow through was measured by collecting the effluent in a attached the column flask graduated and noting the quantity of fluid collected in a given period of time. Invoking the hydraulic equivalent of Ohm's Law, we see that R = P/Q where P represents the source pressure and Q the rate of volume flow. Because there is a source resistance associated with flush bag and tubing, the resistance measured by this the method 51 represents the sum of the source and load (tubing) To correct for this, same technique the source without the resistance fluid resistances. (measured column these measurements are shown in Table 4-1. The to the the attached) subtracted from the total measured resistance. The measured in this way is very close by WJas results value of for theoretical Rtu value based on Poiseuille's Law. 4.1.2 Inertance The calculation of fluid inertance is aiso dependent on whether a parabolic flow profile (the Poiseuille assumption) valid or not. The equation for inertance of a cylindrical is tube of constant cross-section, derived in section 2.2, is L = = fluid mass/(cross-sectional area) 2 pZ/ rr 2 where a multiplying factor of 4/3 frequency to account for a parabolic must be flow values for fluid volume, density of water, included profile. tubing at low Inserting radius and length, we find that 3 2 u = 1.088x10 9 Pa-sec /m The value of Ltu at resonance will, in general, lie between this value and the zero-frequency value. Since somewhere we are interested in linearizing our model by choosing values for the line parameters that match the true values at re sonance, we I R tu Q(mm ) Q(mm /sec) 7850 735 .0981 5500 550 .2727 3250 325 .2923 3570 357 .2521 average .2724 +/- .0201 R(bag+tubing)-R(bag only) .1743 +/- .0201 mmHg-sec/mm (2.32 +/- 0.26) *10 10 Pa-sec/m equivalent circuit: Rbag Pbag Table 4-1. Tubing resistance measurements R(mmHg-sec/mm 53 ne :a to know thi e locazion of ;he natural frequency of our model before we can specify a (3a). However, vlue since n; frequency in this equation value (27), (3a) and (27). iterative procedure occurs L using equation determine ei4C a we need to resort to technique to determine a value equations for for So L tu long that as new an natural iterative satisfies convergence both of (unproven in theory), there the are no compliance of anticipated difficulties. 4.1.3 Compliance Theoretical Determining a theoretical value for the plastic pressure tubing is quite difficult, since exhibits viscoelasticity (wall loss) and creep the tubing (plastic flow) which makes the compliance complex and frequency dependent. The classical formula equation (5) may be for compliance cautiously of applied a thick-walled and reasonably valid at higher frequencies, where expected creep wall loss) may be neglected. Substituting in these values: E r h 1 s = = = = = 1.72 0.59 0.40 1.19 0.46 x10 8 mm mm m Pa we can estimate Ctu as Ctu = 3.25x10 m'/Pa tube, to (but be not approximate Exper imental To directly measure Ctu(as well as frequency, the flow generator used. The system was CtG described assembled as as a function in section depicted in 3.6 Figure taking extreme care to exclude air from the system and cause large errors in compliance 4-1, solution measurement. The transducer compliance was directly measured by calculating ratio of pressure applied (see volume Table increment 4-2). The to observed compliance change was average value of 2.68x10 - 1 5 m 3 /Pa manufacturer specifications for this gave some degree of confidence , well transducer. that the in relatively independent of frequency over the frequency range tested, an was tighten all couplings, since even microbubbles in the filling will of with within This extraneous the result sources of compliance had been eliminated. The total compliance of the parallel combination transducer and extension tubing was then measured between and 13.2 Hz, and Ctu calculated of 0.12 as Ctu :C total -C tr where .Ctuis taken to be independent of frequency. A strong frequency dependence was noted (see Figure 4-2) with the tubing becoming increasingly stiff as the frequency was increased. curve-fitting and extrapolation, Ctufor very low and very high frequencies could be estimated. A high frequency compliance 1.95 x10 14 m 3 /Pa By of (the curve has apparently reached a limiting value by 13.2 Hz) was used for this represents resonance, the value all in subsequent the models, since range around frequency and for the purposes of this study we inaccuracies in can tolerate component values at low frequency. One assumption which should be justified at this point, is that the frequency range over which these measurements made is well below the resonant frequency of the assumption is critical because the effect of and reflections near resonance results in systeim. tbinJ a change effective compliance (i.e. the tubing can no longer be as a lumped element). Fortuitously, the compliance tubing reached a limiting value well below the assumed frequency of the system (.fn>50 Hz), were This iLpe n4 in t:he treated* of this natural so treating the tubing as a lumped comrpliance in this measurement appears to be justified. to motor syringe 3 A to amplifier stopper 4 ft tubing transducer Figure 4-1. Setup for direct compliance measurement Stroke (aV) Test transducer only transducer 0.04 mm 0.20 mm Frequency (Hz) P (mmHg) C (mm3/100mmHg) 0.20 112 0.0357 1.25 115 0.0348 1.60 115 0.0348 2.75 115 0.0348 4.50 117 0.0342 0.12 11.6 1.724 0.92 16.4 1.219 1.20 18.4 1.087 1.83 22.0 0.909 3.00 30.0 0.667 4.05 37.4 0.535 5.25 44.0 0.455 13.20 70.0 0.290 + tubing Table 4-2. Tubing compliance measurements I 58 ×I014 +: measured 20.0 19.0 Ctu 18. 0 I (m3/paP7 . 0 16.0 15. 0 14.0 JL II t 7.. ,..2____fl! ! I :i I ' t I I I i , ., f 'I I, ... _. _ I I ~~ L , i brr 13. 0 t~~ L - I ar I i ' 12. 0 11. 0 10.0 I 9.0 8.0 I __ 7.0 6.0 -I f .I_ I 2 3 I _ i 5.0 4.0 3.0 2.0 1.0 0.0 0 1 4 5 6 7 8 9 10 11 12 13 14 15 f(Hz) Figure 4-2. Tubing compliance vs frequency 59 4.2 Determination of Resonant for Frequency Bubble-free the System 4.2.1 Experimental To test the validity of the lumped-parameter model in terms the of predicting the resonant frequency of the experimental frequency response was monitoring compared system, with theory using the measured or inferred values of the line parameters. The system shown in Figure response measured using 4-3 the was assembled, white noise and frequency the source and spectrum analyzer at a static pressure of 50 mmHg. The system was verified as being bubble-free by remeasuring the frequency response at static pressure of 150 mmHg and noting no apparent change in a the frequency response. Because the compliance of an air bubble is strongly the effect of dependent on the applied pressure, increasing the static pressure on a system with bubbles will be a decrease in the effective compliance and thus a shift in resonance to a higher frequency, as has been shown by Henry et al (1967). The gain and phase of the transfer function at both values of static pressure is shown in Figure 4-4. We note that artifact is present at about 56 Hz. This artifact, which is in all spectra presented in this work, is the result noi/se generated within the pressure amplifiers. of This which was of major importance in our study. at a seen carrier noise unavoidable since the amplifiers are being used far beyond design bandwidth, but fortunately did not occur an is their frequency 60 white noise source BIOTEK to pressure amplifier reference trans ducer 3-way valve pressure amplifier 4 ft. extension tubing test transducer Figure 4-3. Diagram of laboratory setup for testing of catheter system 61 It IM I in mmHg as -i• -2 LO6 I_ 5.8 438 omHg 18 L8 -I -18' -273 r-oww -rlo Figure 4-4. Transfer function for the bubble-free system 4.2.2 Theoretical To locate the resonant frequency of The second-order we need to si3.ultaneously solve equations (3) and and Ltu (Cou and 2.68xi0 - 1 5 m/Pa Ctr are assumed constant and at for Rtu 19.5 and , respectively). Starting with an initial guess at the resonant frequency of the model allows and L, (27) model, the two sets of equations calculation can then be of solved iteratively until R and L (and wn) converge. Convergence has been shown to be guaranteed, but we experienced no R not difficulties in computing a convergent solution. The values for R, L, fn, and D (damping coefficient) found by this method are: fn = 45.78 Hz R1tu 5.561 x1010 Pa-sec/m 3 Ltu= 1.265 x10 D Pa-sec2/m 3 = 0.0761 The frequency response of the second-order system with this natural frequency and damping coefficient is shown in Figure 4-5 with the experimental curve superimposed. The agreement is fairly good considering the many approximations involved in constructing the model and the possible errors in measuring the compliance the tubing (extraneous sources of compliance measurements high). Our check on equation (30), becomes r >> 2 tan;o pr 0 the tend assumption to of make low of our loss, 63 287.6 rad/sec >> 43.9 rad/sec which is not a bad approximation. 64 "r .0 II /,M I solid dashed 90 1 der model 1.0 I -§0 S(Hz) ~ ~24, -279d PUU PjAlURW JU%/ Figure 4-5. Comparison of second-order model with experimental transfer function 65 4.3 Air Bubble Experiments 4.3.1 Resonant Frequency as a Function of Bubble Location The development of section 2.5 has suggested that the presence of a bubble in the pressure tubing may be represented by a parallel lumped compliance introduced at the appropriate in the transmission simplification line was then model to of the system. theoretically transmission line into two sections, pre- and model each section as a second-order circuit. cascade of two R,L,C circuits fourth-order equation in w (see desirable separate the post-bubble, This ,;hose ;ransfer equation A point and leads to a function is a (28)). The frequency response of this fourth-order system exhibits two resonant.peaks, a dominant (primary) peak at a lower frequency and a secondary peak at a higher frequency. Since from a practical point of it is the primary resonance which is of interest, the main of this theory is that, for a bubble of given size, resonant frequency decreases as the bubble is the advanced view result primary in the tubing towards the load (transducer) end. To test this theory, we measured the frequency response of our standard system as a bubble of known volume was advanced in the tubing by means of controlled flushing. The response function of the system was measured using the 3582A Spectrum Analyzer and noise source as in the previous experiment. First, the system was carefully filled with the standard solution. The frequency response was then measured at static pressures of 50 and 150 mmHg and compared to verify the absence of any bubbles. Next, a bubble 66 of known volume at atmospheric 3 ) was introduced mm pressure (29.0 at the Biotek dome and advanced through the fluid column by slow flushing. The frequency response of the system was Measured the bubble located at various points (20, 40, 60, 80, along the fluid column. The response functions Figure 4-6(a)-(f). To compare these results with section 2.5.2, plotted in Figure 4-6. damping at higher frequencies the theoretical Note response that the tends to and 122 cm) are the with shown in theory of functions are also higher-than-predicted obscure resonant peak in the magnitude plot, but the phase the plot second clearly shows the 180 degree phase shift around the natural frequency. For the tubing line constants, Rtu and Ltu were allowed to vary with the excitation frequency according to equation (3) and the bubble compliance was calculated using equation (34) . HIote that the model tends to overestimate the high frequency resonance and underestimate the damping. There is for the former result. The latter no is obvious most nonnegligible wall loss (viscoelasticity) in our model ignores. The qualitative explanation likely the agreement, due tubing, to which however, is bubble as "decoupling" the fluid line into two sections, each of which may encouraging, for it supports the view of the be approximated by a lumped second-order circuit. I Ilmlt• II in dows a8 t8 -27 no bubble -306 I -j •)Ust (a) J.JuLI; . 2.8 LI 0. -275 PAi WUVJ x= ZO cm (b) Figure 4-6. r lA dashed line: theoretical 4.8 Jiu iru eqermentaLLLal Model and experimental transfer functions 68 CJI I II"fl 4.B .artifact 2.1 LO -273 -USI li J%--rv %WLL (c) 5.8 4.8 10 rtifact LI 1.8 -271 x=-6 0 rAa cm r AM= Mi (d) 1n r Illff%3 4.08 3.8 2,8 artifact / -18 -278 -3Hi 1 -4 x=80 (e) 5.10 4. 3.0 artifact LI Ll -98 -118 -278 -1ZF/C tm rlPim WwIY (f) 4.4 Tap and Flush Experiment The theoretical development of Chapter 2 has suggested that most catheter systems may be approximated by a limited number of lumped sections, the number of sections neccessary on bubble distribution within the system but d-epending perhaps practically limited to two or three. This lumped model has had the feature of simplifying system analysis at the expense of absolute accuracy. In terms of deducing the system response function H(w) from the time response to a known input (typically a pressure step at the patient end of the catheter), many the researchers lumped second-order model to be adequate. have This is shown because, most cases, even a rough knowledge of the location of resonant frequency and damping coefficient may be judge the adequacy of the system or compensation. via impedance even matching. to the Often, first sufficient attempt in to frequency discovery of a lower-than-expected resonant frequency may result in a search for occult air bubbles rather than inverse filtering of the waveform or other electrical technique which pressure requires fairly precise knowledge of the transfer function. The clinical unsuitability of the "pop" technique and other- inputs which require access to the catheter tip led us to examine two alternate methods of exciting the natural frequencies of the system which do not require withdrawal of the catheter. The first method, tapping the extension tubing with a- small device, represents an attempt to provide an impulse mechanical of to the system. The other, approximating a pressure step, pressure is the "flush" technique technique is quite recommended by Gardner convenient in that (1970,1981). the flush This device normally present as part of the monitoring setup. We tested of these excitations on debubbled systems as well is both as those containing a bubble to determine two things: of the (2) Is the primary resonance of the system, as seen from the (1) Is the time or frequency independent response location of the input? normal pressure source (the catheter tip), sufficiently excited by the input to be detected? 4.4.1 Experiment The experimental system set up to answer these questions depicted in Figure 4-7. Two sections of the H-P were connected in series with bubble to be introduced into a the three-way side pressure valve, port of is tubing allowing the valve. This was done because the side port is a typical a three-way site of bubble entrapment and also to prevent the bubble from being swept out of the tubing during the "fast-flush". A Sorenson flush unit was connected between the transducer and normal clinical location), and the Biotek Intraflo tubing pressure (the simulator connected to the other end of the tubing. The system was filled with the standard solution in usual manner, and the transfer function measured at Pstatic = mmHg and again at 150 mmHg to verify the absence of trapped (see Figure 4-8(a)). The response of the system to the the 50 air following four excitations was tnen tested: (1) Biotek square wave (the model excitation) (2) Tap at x=20 cm (3) Tap at x=224 cm (1-x=20 cm) (4) Flush Typical response waveforms are shown in Figure 4-9(a)-(d). The lack of accurate calibration of the force delivered by the tap device, possible motion artifact of the tubing during the tap and flush rigorous procedures, analysis of and these other problems responses serve exceedingly to make difficult. Nonetheless, the following qualitative analysis may be useful: (1) The Biotek step response was artifact free. Determination of resonant from the peaks of the time response was consistently frequency the and relatively most damping easy and matched the spectrum analysis determined previously; (2) The fast-flush response although care had to be taken to was relatively avoid artifact-free, disturbing the tubing during the maneuver. The initial "spike" artifact in the response could not be used for analysis; (3) The tap response invariably produced some high-frequency oscillation which made determination of f n and D much more difficult. It is not known whether this is due to motion artifact of the tubing, phase cancellation resulting from a secondary wavefront (the impulse can propagate in both directions away from the tap), or perhaps even another mode of wave propagation in the 73 trans ns ducer bubble to pressure amp w "pigtail" valve to flush bag Figure 4-7. Diagram of setup for tap and flush experiments 74 in8 II"ilf i window= 2 sec 4,0 ndows 3.8 2.8 L8 -•8 -188 -278 ree system -MLp#W(&) (a) 4.8 318 2.8 L8 -18 -278 phse(de) with bu bble (b) Figure 4-8. Transfer function with and without bubble 75 * mmHg mmHI 160 120 80 40 0 -40 -80 -120 -160 .2 .4 .6 (a) .8 sec mmH 1.0 sec (b) mmHI sec 0 %VFU (c) Figure 4-9. (d) Square-wave, tap, and flush responses for bubble-free system tube wall itself. We also observed that the tap greater in amplitude as the tap site was response moved closer became to transducer, possibly because of the attenuation produced by the wall loss in the line. A considerable change in the responses to these occurred when a bubble was inserted into the side three-way stopcock connecting the two transfer function of system with bubble decrease in resonant exhibits Lhe the predicted lengths appearance of a second frequency peak. The produce a response from which the lower port of of the tubing. The (Figure 4-8(b)) frequency Biotek resonant damping were easy to deduce (Figure 4-10(a)). The (Figure 4-10(b)-(c)) unfortunately are dominated excitations and continued to frequency tap by and responses oscillation at the higher resonance, leading to the tentative conclusion that the tap is an inadequate excitation for the primary resonance. The flush response has components at both the secondary and primary frequency, as is seen in Figure 4-10(d). However, further experimentation revealed that the high frequency response is not due to the flush itself. Rather, it is due to the manner in which the flush valve in the Intraflo closes. The release "pigtail" which opens the valve actually produces a flow in the tubing as the rubber membrane which valves of the impulse the flow reseats. This can be shown by repeating the flush experiment with equal pressures in the flush bag and tubing. Upon release of flush valve, the piston-like action produces the high-frequency response of shown the the rubber membrane in Figure 4-10(e). 77 Comparing Figures 4-10(d) and (e), it appears that most, if not all, of the high-frequency oscillation produced by the fast-flush is due to the mechanics of valve closure in the Sorenson unit and not from the initial conditions set up Unfortunately, other makes of flush device by steady-state were not determine if they were more suitable for this purpose. flow. tested to 78 mmHg I- 0o Sea 8se (b) (a) __ I tap @ x=224 cm =H S160 120 80 40 01 IA . -40 -80 L20 .-1 160 f " ,, ll "I 0 , I, .2 , I , , ,. , .4 .6 (c) ,. I . . I. , 1.0 .8 see Figure 4-10. Square-wave, tap, and flush responses with hnhhl_ midwn v in t-h f11id 1in t HmmH mR! see (d) 8 vwe (e) 8 CHAPTER 5 DISCUSSIOJ We have gone to considerable lengths in this study to develop a lumped-element model for the catheter system that is good approximation approach through to the the more first accurate resonant transmission frequency, experimentally demonstrate its validity. It may and be a line then argued to that this analysis was not really neccessary to achieve our purported objective of finding an in vivo frequency method of measuring response. We feel, however, that the effort put into modeling has at least paid off in developing a rationale for understanding the relative importance of tubing, transducer, and bubble compliances in determining the primary resonant frequency given dimensions. The analysis has also range of component (R, L, and C) of served values for a to system suggest which the of the lumped approximation is a good one. Having justified the use of a lumped second-order represent the catheter system tested in this proceeded to demonstrate how an occult bubble model study, causes we than one frequency. This result is crucial to of the role that location and energy response to tap and flush results in more detail. inputs. resonances our distribution We can now then decoupling of the fluid column into pre- and post-bubble sections, with result that the column may have significant to at the more understanding play examine in the these 81 5.1 Tap and Flush Responses The results of section 4.4 strongly excitation leads to response surprising for two artifacts. reasons: first, suggest This the that is the not impulse tap entirely has to be transmitted through the tubing wall, which may cause longitudinal wave propagation in the wall, temporary narrowing due to relaxation effects, and other wall of the related lumen phenomena; second, the impulse is a signal which requires very large forces to transfer significant energy to the system - the large pressure variations in the tubing make transmission nonlinearities likely to be evident in the response. Even though we an alternate flow impulse source - the flush valve more discovered - which eliminate some of the problems of the "mousetrap" method may (tubing effects) it still seems impossible to circumvent the nonlinearity problems. Another problem with the tap sensitivity. As we have seen, pre- excitation and is post-bubble location excitations give qualitatively different responses. This may possibly be the result of reflection from the impedance mismatch existing at the bubble The and high-frequency impedance mismatch will attenuation tend to limit in the the tubing. amount of energy transferred through the bubble and thus isolate the two halves of the system. The attenuation makes appreciable Moreover, response the energy without is very it difficult using extremely localized to within achieve high the an forces. tubing, intuitively making the lumped-element model seem unsuitable. All of these considerations combine to make the tap an unattractive 82 method of excitation. On the other hand, we have found evidence to the flush excitation (although perhaps not the device tested) may give results technique to justify its use in similar the sugges: particular enough clinical to and circuit shown previous energy below, catheter considerations. which system First, represents -a containing a d.c. "pop" We can using the consider the justify the similarity of the flush and pop responses lumped model flush the setting. that model bubble of our mid-tubing (component values typical of our experimental system): C1 =3.1*10 -13 C2 =6.2 10-15 Vs S V S where we have substituted V and I for P and Q respectively to avoid confusion between electrical charge and hydraulic flow. We can consider the pop and flush excitations as setting certain initial conditions in the line system to decay. This corresponds to the solution to the coupled differential and then allowing homogeneous equations the (unforced) governing fourth order circuit: (L 1 C1 s 2 + R1 C1 s -1)V1 + (L1 C2 S2 + R1 C2 s)V = 0 2 -V 1 + (L 2 C2 s 2 + R2 C2 s +l)V up 2 = 0 the 83 5.1.1 Initial conditions The initial conditions set up by the pop excitation are: hl 1= 2=bIt = o V1 V 2 = Vs which means that all the energy is stored in the compliances. For the flush, we have I=I2=: V s Ri+R 2 Is ; Ibdt= r 0 V2 = 21 Vx dx = Vý/12 0 . V=2fVV 2x 0S x = 7V2/12 S where we have represented the pressure as a the distance x along the fluid column and linear computed function the square of the pressure within each section. The reason of average for this will be evident shortly. Now, we can determine the approximate distribution of energy within the system under each set of initial conditions. The total energy stored in each section is E EC V EL For the pop excitation, there is no flow, independent of location. i and 1,2. the Since C1 >> C2 (commonly the pressure case is even 84 with small bubbles), the majority of the energy in stored in the bubble. Therefore the response is the system is primarily the decay of the left-hand RLC circuit. Since the lowest tends to be associated with this means the bubble, resonance that the response will in most cases be approximately second order. For tihe flush, there is kinetic energy in the and potential energy in the compliances. Using fluid the motion component values shown, V 2L = 0.5LI Ekin= Ek kin kin 2 1 = 0.5 sL R2 4.8xlO-13V 2 s 4 Ept 0.5C V E pot 0.5C2 V 2 pot 1 1 1 =- (1/24)C V 1 s = 1.3x10 (7/24)C 2 Vz = 1.8x10 We see that for the component values 14 s 1 5V chosen (typical system we have studied) the majority of the energy is the fluid motion and not in the compliances. for the stored Therefore in the initial energy stored within each section is approximately equal. 5.1.2 Transient solution Given these initial conditions, transient solution to the ideal pop and we can flush determine inputs. However, this will require recomputing the line parameters R and L, these are frequency dependent. These observed location of and height the were estimated resonant peaks the since from the in the experim:ental data (Figure 4.7(b)) as: R= 1.5x10 1 0 pa-sec/ m 10Pa-sec/rn 3 R = 2.Sx10 L= 3 7.0x108 Pa-secimn L = 6.3x10 8 Pa-secmn3 2 anrd values parameter A transient solution with these equivalent initial conditions determined from the d.c. was determined using the SPICE circuit simulation the circuit program. The pop test initial canditions is shown in Figure 5-1 and for the flush test the transient solution from 0 to 400 milliseconds for in Figure 5-2. expected, As test pop the almost causes entirely low-frequency resonating response, while the flush yields a response that is a of mixture the response, is easy to distinguish, the particularly since oscillations are quickly damped. This is because the damping increases damping is defined as with guarantees that it oscillations, so we frequency loss/cycle, damping and faster oscillation of will may decay to not the faster reasonably be test low- high-frequency resonances. The low-frequency an and however, high-frequency expected, and both because loss/time. The high-frequency the higher resonance than the low-frequency expect the low-frequency response to dominate. This lends some support to our assertion that the flush excitation is a satisfactory input for determining the low-frequency resonance. Rothe and Kim (1980) have observed "86 th.at the flush waveform produced by Intraflo valve to-transducer excited part of system including snap primarily their catheter. the catheter While we of the Sorenson extension-tube- system, have the not not the entire tested entire systems containing long catheters, it seems likely, in view of our test results and analysis, that it is the flow impulse caused by the snap of the Intraflo valve that produces high-frequency oscillations noted in this study and in the the ;irk of Rothe and Kim. The flush method itself appears to be otherwise sound. How may these results be extended to more general siz-Iions (different component values, more bubbles in the tubing)?.e have outlined a general method for attacking this problem although we have only computed a system. It system in solution for one particular should not prove too difficult to model any specified terms of lumped sections in extension of this work would a similar be :nanner. to' determine An a interesting "worst-case" response - one that contains resonances close enough together make the decoupling assumption poor. to 87 decoupled circuit 20.0 DA321-HPPPICE VERS with pressure step at 3e .1 input 12:12P 13JAN82 0.0 0.0 100.0 TIME Figure 5-1. Simulated response to pressure step at input 200.0 - 3 10 deco d u p 1e 1 i rcuitI ..i th flush s•,urce 40.0 0.0 0 0 T IME Figure 5-2. Simulated response to fast-flush 89 5.2 Extraction of Resonant Frequency and There is some reason (i.e. one without valve ;c believe that closure into the a good artifact) designed. What seems to be required is force a bolus of fluid Damnping from Fast-flush a fluid flush exists valve or that column as source can does it be not closes. However, even if such a device is made available, there are still a number of practical considerations to be considered attempting to calculate resonant frequency and damping in from the Foremost among the anticipated difficulties in clinical use flush response. of the fast-flush for determining resonant separating the flush response from the blood Gardner (1981) frequency pressure has demonstrated the use of the waveform. Sorenson in clinic. The flush valve is released during the diastolic of the cardiac cycle where the slowly. A recording determination of of the resonant pressure response frequency is is is changing then the portion the most examined. relatively damping requires a careful separation of the is The easy, superimposed but flush and patient waveforms. We believe that the flush response analysis may be performed by digital computer with high reliability, relieving the hospital personnel of most of the burden of analysis. What is required some method of patient pressure waveform prediction. This is would allow subtraction of the predicted waveform from the superimposed patient and flush waveforms, resulting in a time-domain filtering of the flush response from the combined signal. Statistical 90 methods of waveform prediction exist, so there is believe that this computation could not be no performed. algorithmic devices would have to be designed in the computer calculation reject reason artifacts, order to Other to make particularly the high-frequency oscillation that we have shown may occur even with an ideal flush source. 5.3 Anticipated Usage and Clinical Acceptability The issue of clinical acceptance of catheter system measurement/correction devices that require user intervention has been raised (Doherty, 1981). require no user intervention Distortion at all Doherty and by others (Brower 1975, have extremely systems been described attractive, distortion but described have been shown to reliably the by none automatically of estimate the methods distortion the entire range of catheter systems in use and patient waveforms. Moreover, that Jackson et al 1980). The idea of determining and correcting waveform is obviously analysis central assumption methods, that the pressure waveform spectrum pressure underlying does over not these contain local maxima simulating a resonance, is open to. some criticism. Whether a method which requires user acceptable depends in part on how and when the interaction method is is intended to be part of used. Certainly the calculation of dp/dt is an important many catheterization high-fidelity procedures system. frequency response will However, not be that a requires catheter suitable an system for extremely with accurate poor dp/dt 91 measurements even if it is compensated. Mechanical compensation can only extend the frequency range a limited amount, and inverse filtering methods (Ciccolella 1976) cannot be expected to perform well far beyond resonance where the deviation response becomes large. Therefore, from second-order determination of resonant frequency and damping is probably more important to the physician or technician than elaborate compensation methods when measuring dp/dt. preferred Presently, catheter-tip manometers are the instruments for this measurement, and until fluid-filled systems can reliably achieve Hz flat frequency beyond, this will probably remain the response case. to Even 100 and if .debubbled systems become the norm, this will be difficult to achieve given the physics of the fluid column and the need for flexible tubing of reasonable length. Simple pressure bandwidths of dp/dt monitoring measurements, does not making much more attractive. The most common and require the fluid-filled noticeable large systems effect of low resonant frequency and low damping on the pressure signal is systolic overshoot. If the measurement is being made in a central artery or in the heart, systolic overshoot may falsely indicate a valvular lesion or disease, or high peripheral resistance. peripheral artery, the overshoot may falsely trigger alarms. It is with these "suspect" systems that a In a pressure direct method of determining the system resonant frequency and damping is most needed. In the busy clinical environment, a catheter system that seems to be reproducing the will pressure waveform accurately probably not be tested for adequate frequency response using the 92 fast-flush or any other technique. However, pressure waveform (damped or should be able abnormalities to and given resonant-looking), differentiate between catheter-induced a suspicious the fast-flush real blood-pressure distortion reliability. An occasional flush test does not seem price to pay for pressure waveform. higher confidence in the with too displayed high high a blood 93 94 BIBLIOGRAPHY 1. Brower, R.W., et ai: A fully automatic device for compensating for artifacts in conventional catheter-manometer pressure recordings. Biomed. Engng. 10:305-310, 1975. 2. Bruner, J.M.R., et al: Comparison of direct and indirect methods of measuring arterial blood pressure, part 1. Med. Instrum. 15:97-101, 3. Ciccolella, S.A.: 1981. Compensation of fluid-filled catheter response using digital filter techniques. Master's thesis, University of Rhode Island, 1976. 4. Falsetti, H.L., et al: Analysis and correction of pressure wave distortion in fluid-filled catheter systems. Circulation 49:165-172, 1974. 5. Fox, F., et al: Laboratory evaluation of pressure transducer domes containing a diaphragm. Anesth. Analg. 57:67, 1978. 6. Frank, 0.: Kritik der elastichen manometer. Zschr. Biol. 44:445-61 3 , 1903. 7. Gabe, I.T.: Pressure measurement in experimental physiology. In Cardiovascular Fluid Dynamics Vol. 1. Edited by Bergel, D.H.. 8. New York: Academic Press, 1972. Gardner, R.M., et al: Catheter-flush system for continuous monitoring of central arterial pulse waveform. J. Appl. Physiol. 29: 911-913, 1970. 9. Gardner, R.M.: Direct blopod pressure measurement - dynamic response requirements. Anesthesiology 54:227-236, 1981. 10. Geddes, L.A.: The Direct and Indirect Measurement of Blood Pressure. Chicago: Year Book, 11. Hansen, A.T.: 1970. Pressure measurement in the human organism. Acta Physiol. Scand. 12. Hansen, A.T., 19 (Suppl. 68):1-227, 1949. and Warburg, E.: The theory for elastic liquid-con3aining membrane manometers. Acta Physiol. Scand. 19 .(Suppl. 65):306-332, 1949. 13. Henry, W.L., et al: A calibrator for detecting bubbles in cardiac catheter-manometer systems. J. Appl. Physiol. 23:1007-1009, 1967. 14. Jackson, L.B., Jaron, D., and Wood, S.L.: Compensation of fluid-filled catheter pressure waveforms by linear predictive analysis and digital inverse filtering. Proc. Fifth New Eng. Bioengng. Conf., Apr. 15. Jager, G.N., Westerhof, N., 1977. and Noordergraaf, A.: Oscillatory flow impedance in electrical analog of arterial system. Circ. Research 16:121-133, 1965. 16. Krovetz, L.J., et al: Limitation of correction of frequency dependent artefact in pressure recordings using harmonic analysis. Circulation 50:992-997, 1974. 17. Lambert, E.H., aid Wood, E.H.: The use of a resistance wire strain gauge manometer to measure intraarterial pressure. Proc. Soc. Exper. Bio. & Med. 64:186-190, 1947. 18. Lambossy, P.: Oscillations forcees d'un liquide incompressible et visqueux dans un tube rigide et horizontal. Helv. Phys. Acta 25:371-336, 1952. 19. Lapointe, A.C., and Roberge, F.A.: Mechanical damping of the manometer system used in the pressure gradient technique. IEEE Trans. Biom. Engng. 21:76-77, 1974. 20. Latimer, K.E.: The transmission of sound waves in liquid-filled catheter tubes used for intravascular blood-pressure recording. Med. & Biol. Engng. 6:29-42, 1968. 21. Latimer, K.E., and Latimer, R.D.: Measurements of pressure-wave transmission in liquid-filled tubes used for intravascular blood- pressure recording. Med. & Biol. Engng. 7:143-169, 1969. 22. Li, J.K.-J., van Brummelen, G.W., and Noordergraaf, A.: Fluid- filled blood pressure measurement systems. J. Appl. Physiol. 40:839-843, 1978. 23. McDonald, D.A.: Blood Flow in the Arteries. London: Arnold & Co., 1960. 24. Melbin, J., and Spohr, M.: Evaluation and correction of manometer systems with two degrees of freedom. J. Appl. Physiol. 27:749-755, 1969. 25. Roark, R.J., and Young, W.C.: Formulas for Stress and Strain (p. 104). 26. Rothe, C.F., New York: McGraw-Hill, 1975. and Kim, K.C.: Measuring systolic arterial blood pressure. Critical Care Med. 8:683-689, 1980. 27. Shapiro, G.G, and Krovetz, L.J.: Damped and undamped frequency response of underdamped catheter manometer systems. Am. Heart. J. 80:226-236, 1970. 97 28. Shinozaki, T., et al: The dynamic responses of liquid- filled catheter systems for direct measurement of blood pressure. Anesthesiology 53:498-504, 1980. 29. Vierhout, R.R.: The response of catheter-manometer systems used for direct pressure recording. Ph.D. thesis, University of Nijmegen, Holland, 1966. 30. Yanof, H.M., et al: A critical study of the response of manometers to forced oscillations. Physics in Med. & Biol. 8:407-422, 1963. 31. Womersley, J.R.: Method for the calculation of velocity, rate of flow and viscous drag in arteries when the pressure gradient is known. J. Physiol. 127:553-563, 1955. 98 APPEI1DIX A Table of AnalIogous ElecL ri cal and Hydraulic Electrical '-Electrical Voltage Hydraulic H1ydraulic Pressure V Charge (coulombs) Current I=dq/dt (A) Power VI (W) Resistance R=V/I (ohm) Inductance L=V/(dI/dt) Capacitance C=I/(dV/dt) =q/V (F) Kinetic Energy 1/2LI (J) Potential Energy 1/2CV (J) Resistance/m R'=R/I (ohm/m) Inductance/m L'=L/1 (H/m) Capacitance/m C'=C/l (F/m) P=F/A (N/m2 or Pa) Fluid Displacement X=Ax (m3 ) Fluid Velocity Q=dX/dt=Ax' (m3 / sec) Power PQ (J/sec) Resistance R=P/Q (Pa-sec/m 3 ) Inertance L=m/A (Pa-sec 2 /m 3 Compl iance C=X/P (m3 / Pa) Kinetic Energy 1/2LQ (J) Potential Energy 1/2CP (J) Resistance/m R'=R/1 (Pa-sec/m 3 ) Inertance/m L'=L/1 (Pa-sec 2 Compliance/m C'=C/1 (m3 / Pa) A=cross-sectional area of the tube in mA l:length of tubing / m3) Un its 99 APPENDIX B Calculation of fn and D From the Step Response A system is said to be second order if its dynamic can be described by a second-order second-order equation used to differential response equation. -pproxima~e the frequency The response of the catheter system is 2 s +2Ds 2DS + 1 S2 n n P (t) = P (t) o i where o = undamped natural frequency (rad/sec) D = damping ratio (dimensionless) The step response for an underdamped system (D < 1) is P (t) = on 0 0 = I e- sin(/I- 2t + 0 Dn arcsin(/F-D') We can determine D and n (or fn= ) from the step by measuring the time between peaks and the ratio of the response heights of adjacent peaks. For the typical step response shown below: PT -> V-T 4 time 100 the response reaches a maximum (2n+1)/2T when the sine . Solving for. tn and the ratio Yn/Yn+ argument , equals where Yn is the output pressure at time t=tn, we have t (2n+.l) n - -n Yn+1 If we define V = log Yn then V D Irt~T 2fV 2 to solve for fn, we note the time period between maxima. This yields f 1/T r where f is the resonant frequency of frequency is simply fn = f*r/ / = the system. The natural