USD 2643 Palltronic® Flow Check II Unit Simple verification of flow measurement instruments Features u Actual flow rate is automatically calculated and shown on the display Benefits u No need to calculate flow rate using conversion table u A precise pressure sensor determines the atmospheric pressure u Even small changes in height or weather conditions are detected u Robust, sanitary stainless steel design makes it ideal for use in the pharmaceutical industry u The unit can be used with all integrity test instruments in laboratories or cleanrooms u Easy to use Stäubliu connections u u Internal Pall Gaskleen® filter protects the capillary in order to ensure maximum service life The unit can be easily and quickly connected to and disconnected from filter integrity test instruments u Internal filter protects against particles and liquids destroying the capillary Description Specifications The Palltronic Flow Check II unit enables the user to perform a fast, simple check on a filter integrity test instrument to confirm that the instrument is measuring the air or nitrogen gas flow correctly. A ‘Flow Check’ test can be performed as often as the user requires, in order to ensure complete confidence in the integrity test instrument. The user simply connects the Flow Check II unit to the integrity test instrument and performs either a Flow Check test (Palltronic Flowstar integrity test instruments only), Forward Flow or Diffusion test. The flow measured by the integrity test instrument is then compared with the calibrated flow of the Flow Check II unit. Dimensions 180 x 150 x 90 mm (6 x 5 x 3 in.) The gas flow rate through a capillary is dependent on the differential pressure across the capillary. The inlet pressure is controlled by the connected integrity test instrument, as pressure above the actual atmospheric pressure. The atmospheric pressure is dependent on the weather situation and local height above sea level. The Flow Check II unit has a built-in pressure sensor. When the button on the side of the unit is pressed, the atmospheric pressure is measured and the unit displays the actual flow rate. There is no need to use a table to calculate a correction factor. Weight 2 kg (4.4 lbs) Battery: The unit include a standard battery (1.5 V battery type AA (LR06/AM-3). The battery is replaced during annual service and calibration. Scope of Supply Each Palltronic Flow Check II unit is supplied with a calibration certificate, Instructions for Use, pneumatic tubing and Stäubli* fittings. The Palltronic Flow Check II unit is available in three configurations - which version should be used depends on the flow range of the integrity test instrument being checked. For the FC02M and FC02H models (see table below), a volume must be installed between the unit and the instrument. Ordering Information Part Number FC02 Flow Rate Range 8 ± 1 mL/min Upstream Volume Requirements 0.5 to 2 meter hose FC02M 90 ± 10 mL/min 0.5 to 2 liter vessel FC02H 900 ± 15 mL/min 2.5 to 6 liter vessel Visit us on the Web at www.pall.com/biopharm e-mail us at biotech@pall.com United States 800.717.7255 toll free (USA) 516.484.5400 phone 516.801.9548 fax biotech@pall.com e-mail Europe +44 (0)23 9230 3303 phone +44 (0)23 9230 2506 fax BioPharmUK@europe.pall.com e-mail International Offices Pall Corporation has offices and plants throughout the world in locations such as: Argentina, Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea, Malaysia, Mexico, the Netherlands, New Zealand, Norway, Poland, Puerto Rico, Russia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, the United Kingdom, the United States, and Venezuela. Distributors in all major industrial areas of the world. The information provided in this literature was reviewed for accuracy at the time of publication. Product data may be subject to change without notice. For current information consult your local Pall distributor or contact Pall directly. © 2009, Pall Corporation. Pall, , Palltronic and Gaskleen are trademarks of Pall Corporation. ® indicates a trademark registered in the USA. Filtration. Separation. Solution. is a service mark of Pall Corporation. uStäubli is a trademark of Stäubli A.G. 10/2009, PDF, UK GN09.3737 USD 2643 SM