Palltronic Flow Check II Unit Simple verification of flow measurement instruments ®

advertisement
USD 2643
Palltronic® Flow Check II Unit
Simple verification of flow measurement instruments
Features
u Actual flow rate is automatically
calculated and shown on the display
Benefits
u No need to calculate flow rate using
conversion table
u
A precise pressure sensor determines
the atmospheric pressure
u
Even small changes in height or weather
conditions are detected
u
Robust, sanitary stainless steel design
makes it ideal for use in the
pharmaceutical industry
u
The unit can be used with all integrity
test instruments in laboratories or
cleanrooms
u
Easy to use Stäubliu connections
u
u
Internal Pall Gaskleen® filter protects the
capillary in order to ensure maximum
service life
The unit can be easily and quickly
connected to and disconnected from
filter integrity test instruments
u
Internal filter protects against particles
and liquids destroying the capillary
Description
Specifications
The Palltronic Flow Check II unit enables the user to perform a
fast, simple check on a filter integrity test instrument to confirm
that the instrument is measuring the air or nitrogen gas flow
correctly. A ‘Flow Check’ test can be performed as often as the
user requires, in order to ensure complete confidence in the
integrity test instrument. The user simply connects the Flow
Check II unit to the integrity test instrument and performs either
a Flow Check test (Palltronic Flowstar integrity test instruments
only), Forward Flow or Diffusion test. The flow measured by the
integrity test instrument is then compared with the calibrated
flow of the Flow Check II unit.
Dimensions
180 x 150 x 90 mm (6 x 5 x 3 in.)
The gas flow rate through a capillary is dependent on the
differential pressure across the capillary. The inlet pressure is
controlled by the connected integrity test instrument, as pressure
above the actual atmospheric pressure. The atmospheric
pressure is dependent on the weather situation and local height
above sea level. The Flow Check II unit has a built-in pressure
sensor. When the button on the side of the unit is pressed, the
atmospheric pressure is measured and the unit displays the
actual flow rate. There is no need to use a table to calculate a
correction factor.
Weight
2 kg (4.4 lbs)
Battery:
The unit include a standard battery (1.5 V battery type AA
(LR06/AM-3). The battery is replaced during annual service
and calibration.
Scope of Supply
Each Palltronic Flow Check II unit is supplied with a calibration
certificate, Instructions for Use, pneumatic tubing and
Stäubli* fittings.
The Palltronic Flow Check II unit is available in three configurations
- which version should be used depends on the flow range of the
integrity test instrument being checked. For the FC02M and
FC02H models (see table below), a volume must be installed
between the unit and the instrument.
Ordering Information
Part Number
FC02
Flow Rate Range
8 ± 1 mL/min
Upstream Volume
Requirements
0.5 to 2 meter hose
FC02M
90 ± 10 mL/min
0.5 to 2 liter vessel
FC02H
900 ± 15 mL/min
2.5 to 6 liter vessel
Visit us on the Web at www.pall.com/biopharm
e-mail us at biotech@pall.com
United States
800.717.7255 toll free (USA)
516.484.5400 phone
516.801.9548 fax
biotech@pall.com e-mail
Europe
+44 (0)23 9230 3303 phone
+44 (0)23 9230 2506 fax
BioPharmUK@europe.pall.com e-mail
International Offices
Pall Corporation has offices and plants throughout the world in locations such as: Argentina, Australia,
Austria, Belgium, Brazil, Canada, China, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea,
Malaysia, Mexico, the Netherlands, New Zealand, Norway, Poland, Puerto Rico, Russia, Singapore, South Africa,
Spain, Sweden, Switzerland, Taiwan, Thailand, the United Kingdom, the United States, and Venezuela. Distributors
in all major industrial areas of the world.
The information provided in this literature was reviewed for accuracy at the time of publication. Product data may be
subject to change without notice. For current information consult your local Pall distributor or contact Pall directly.
© 2009, Pall Corporation. Pall,
, Palltronic and Gaskleen are trademarks of Pall Corporation. ® indicates a
trademark registered in the USA. Filtration. Separation. Solution. is a service mark of Pall Corporation. uStäubli is a
trademark of Stäubli A.G.
10/2009, PDF, UK GN09.3737
USD 2643
SM
Download