WFC3 Ambient Calibration #2 Results H. Bushouse TIPS May 17, 2007 1 Ambient Cal #2 Overview • WFC3 ambient calibration #2 ran from April 13-19, 2007 – WFC3 in the GSFC SSDIF clean room – Latched into Radial Instrument Alignment Fixture (RIAF) and “CASTLE” optical stimulus – UVIS channel operations only (IR can’t be cooled sufficiently in ambient) – UVIS-2 detector with CCD’s at -54 C (vs. -83 C flight temp) • Goals: – – – – Verify re-integrated instrument operations Verify successful lien closures First look at UVIS-2 in (nearly) flight-like conditions Verify operations of ground support equipment to be used in thermalvac #2 – Train and verify operations of support personnel TIPS May 17, 2007 2 Instrument Preparations for Ambient Cal #2 WFC3 closeout just prior to top cover installation WFC3 top cover installed WFC3 lift from Low Level Positioner (LLP) WFC3 inserted into Radial Instrument Alignment Fixture (RIAF) TIPS May 17, 2007 3 Accomplishments • ~30 SMSs, ~110 hours, 1525 images – – – – – – – – – – – – Optical stimulus verification UVIS-2 detector alignment check System throughput with UVIS-2 Filter ghost check (lien related) CCD electronic crosstalk check (lien related) UVIS grism calibration: throughput, dispersion, and flat fields Internal Calibration Subsystem check (lien related) UVIS-2 bias, dark, readnoise, gain measurements UV flats signal level check Encircled energy check CCD charge injection test First test of CCD Extended Pixel Edge Response (EPER) mode • Excellent job by I&T, Ops, CASTLE & Sci Cal teams TIPS May 17, 2007 4 Thanks are due … • IS’s – S. Baggett, T. Brown, H. Bushouse, G. Hartig, A. Martel, L. Petro, M. Robberto • DA’s – E. Barker, M. Cracraft, B. Hilbert, S. Holfeltz, D. Karakla, J. Kim, K. Lindsay, D. Long, R. Lucas, J. Mack, H. McLaughlin, C. Pavlovsky, B. Shaw, E. Smith, M. Sosey TIPS May 17, 2007 5 Instrument Science Reports • ISR 2007-15: WFC3 Ambient-2 Testing: UVIS Readnoise S. Baggett 11 May 2007 • ISR 2007-14: WFC3 Ambient-2 Testing: Calibration Subsystem Performance S. Baggett 7 May 2007 • ISR 2007-13: UVIS CCD EPER CTE measurements M. Robberto 30 Apr 2007 • ISR 2007-11: WFC3 Ambient-2 Testing: UVIS Gain Results S. Baggett 27 Apr 2007 • ISR 2007-10: WFC3 Ambient Testing: UVIS Dark Current Rate B. Hilbert 26 Apr 2007 • ISR 2007-09: UVIS Channel Filter Ghosts after Filter Replacement T. Brown 26 Apr 2007 • ISR 2007-08: Searching for Scattered Light on the WFC3 UVIS-2 Detector T. Brown 24 Apr 2007 • ISR 2007-07: WFC3 System Throughput on the UVIS Build 2 Detector T. Brown 19 Apr 2007 TIPS May 17, 2007 6 System Throughput with UVIS-2 From STScI Instrument Science Report WFC3-2007-07 (Tom Brown) • • • • UVIS-2 QE better than predicted Response of both CCDs comparable to UVIS-1 chip 1 UVIS-2 curves subject to small revision at proper operating temp Optics throughput appears to have held up well since 2004 TV#1 TIPS May 17, 2007 7 UVIS Detector Throughput UVIS Detector Comparison Comparison (UVIS-1 QE) / (UVIS-2 QE) 1.8 1.6 1.4 1.2 1 0.8 0.6 200 300 400 500 600 700 800 900 1000 Wavelength (nm) Ratio of Higher QE CCDs Ratio of Average QEs • UVIS-1 wins in blue; UVIS-2 wins in red • F225W only important filter for which response differs by >10% TIPS May 17, 2007 8 Excellent Read Noise w/ UVIS-2 • Like UVIS-1, UVIS-2 demonstrates “goal” level read noise of 3 e- rms, vs. spec of 4; very important in UV Gain = 1 e-/DN UVIS-2 (ambient) UVIS-1 (ambient) Amp A 2.95 3.02 Amp B 2.94 2.97 Amp C 2.81 2.86 Amp D 2.90 2.99 Amp A 3.35 3.21 Amp B 3.34 3.25 Amp C 3.26 3.23 Amp D 3.53 3.28 Gain = 1.5 e-/DN S. Baggett TIPS May 17, 2007 9 UVIS Crosstalk Satisfactorily Resolved • Extended source crosstalk (which could result from very low level signals) is now undetectable (<0.1 e-) – Combination of hardware change and timing pattern change – New timing patterns show slightly improved read noise as well (had been concerned they might have higher noise) • No crosstalk in 2-amp readouts (i.e. from CCD to CCD) • In 4-amp readout, small level of crosstalk remains from saturated sources only – New timing patterns slightly improve that level as well – What remains is not a significant issue preliminary analysis by S. Baggett TIPS May 17, 2007 10 UVIS Crosstalk Satisfactorily Resolved (2) Performance w/Hardware Rev & New CCD Timing Patterns Original Problem window-CCD ghost filter-CCD ghost crosstalk source stimulus artifacts TIPS May 17, 2007 11 Internal Cal Subsystem Illumination Improved • Severe glints reduced; see F390M flats below • Count rate specs achieved • As shown previously w/UVIS-1, Deuterium flats much improved in uniformity (2× vs. 5-10×) F390M (inverse) flat before TIPS May 17, 2007 F390M flat now 12 UVIS Grism Throughput Measured UVIS grism is WF/PC-1 spare; very little surviving data Data provided to SOC for ERS planning H. Bushouse TIPS May 17, 2007 13 CCD Charge Injection Working Properly Mean charge injection signal level ~16,000 e-/pix Noise in a single column ~15 e- rms Much greater CTE benefit than from a lamp pre-flash of equivalent noise level TIPS May 17, 2007 14 Improved Filter Ghost Performance Confirmed • Ambient cal measurements confirm the improved ghost performance of the filters replaced in the SOFA • Biggest success stories: F225W (<0.4% vs. 15%), F606W (elimination of point-like ghosts), F280N (0.6% vs. 10-50% in ghost and horrific scattered light flares) Old F225W New F225W TIPS May 17, 2007 15 UVIS-2 Detector Alignment • UVIS-2 aligned at Ball on 30 Jan ’07 – used new, direct, “through the window” alignment transfer from UVIS-1 to UVIS-2 • UVIS-2 within tolerance in 4 degrees of freedom, out in 2 • Additional adjustment of tip/tilt is desirable, if misalignment is confirmed in upcoming TV#2 testing UVIS-2 Alignment Summary Apr '07 (Post-align) Tolerance QX (o) 0.332 ±0.20 QY (o) -0.397 ±0.20 QZ (o) X (mm) Y (mm) Z (mm) -0.032 0.251 0.032 0.276 ±0.500 ±0.300 ±0.300 ±0.850 TIPS May 17, 2007 16 UVIS-2 Misalignment Consequences • At best focus, the performance degradation due to the current level of tip/tilt misalignment is negligible, however … • OTA breathing exacerbates the misalignment, causing a small loss of performance over field, especially at short wavelengths – Slight EE loss in small apertures – PSF change (due to field dependent astigmatism inherent in optical design) TIPS May 17, 2007 17 What’s Next? • IR-1 detector package installation – Received at GSFC yesterday (5/16); installation today • System Level Thermal-Vac #2 – WFC3, with UVIS-2 and IR-1, into SES chamber ~5/29 – Ambient system functionals and alignment check ~6/1-6/5 – Vacuum testing for following ~75 days • 24/7 support from STScI IS’s, DA’s, and Ops – Proto-flight qualification of instrument • • • • Thermal cycling Performance of all thermal systems Full science calibration of UVIS and IR channels CEI specifications verification TIPS May 17, 2007 18 IR FPA QEs 100 Quantum Efficiency (%) 90 80 70 IR3 60 50 IR1 40 IR2 30 20 10 0 200 IR Channel Wavelength Range 400 600 800 1000 1200 1400 1600 1800 Wavelength (nm) FPA 64 FPA 129 TIPS May 17, 2007 FPA 150 FPA 160 19 Backup slides TIPS May 17, 2007 20 UVIS-2 Misalignment EE Penalty OTA SM despace: +3m Nom -3m Detector alignment: Optimal Optimal As-is As-is TIPS May 17, 2007 21 PSF Change with OTA Breathing UVIS-2 CASTLE 633nm UV-11 -3m 0m TIPS May 17, 2007 +3m 22 FPA 160 Dark Current % of Pixels Above Dark Rate 100 90 80 70 60 50 40 30 20 10 0 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 Dark Current (e-/sec/pix) FPA 64 FPA 129 TIPS May 17, 2007 FPA 150 FPA 160 23 FPA 160 Persistence TIPS May 17, 2007 24 FPA 160 Read Noise CDS Noise (e- rms) FPA 64 (IR2) 24 FPA 129 (IR1) 22 FPA 150 (IR3) 24 FPA 160 25 FPA 165 24 TIPS May 17, 2007 25 FPA 165 QE TIPS May 17, 2007 26 SMGT preparations • Revised SMGT Part 2 SMS – – – – Added UVIS Anneal exposures Revised management of tungsten & D2 lamps Corrected UVIS quad filter commanding Duration of SMS is 20.5 hours • Delivered SMS 3/30/2007 for test – SMS ran successfully in VSTIF 4/23/2007 – Science data ftp’ed to STScI for OPUS processing • Revised SOGS Support Schedules for SM4 Functional Test – Science data from Early VEST test ftp’ed to STScI for OPUS processing • Prepared OPUS for processing of VSTIF and Early VEST data – Anomalies identified from February VSTIF run corrected – Software testing and delivery complete by 5/2/2007 TIPS May 17, 2007 27 SMGT Part 2 SMS Content Visit 01 02 03 05 06 04 07 08 09 10 11 12 13 14 15 16 17 18 19 20 27 24 21 22 23 25 26 28 Content/purpose Recovery from SAF E mo de. CR-SPLIT 0 - 7, various apertures, filters. 1-hr exposure, Take Data Flag dropped. CTE; DARK and Tungsten targets. User & predefined sub-aperture; Selected ampl ifiers. On-chip binning; Ampl ifier gain; Occulted target. Selected ampl ifiers. Deuterium lamp , high current. Predefined dither patterns. User sub-aperture. Deuterium lamp , low and medium current. Quad filters. Quad filters. CCD charge injection. Deuterium lamp , low current. UVIS optics alignment and focus. Anneal. Anneal with TEC disabled. Bias exposure. Transition UVIS to IR. Tungsten lamp . RAPID. Transition IR t o UVIS. Dark target. STEP200, STEP400, STEP600. Occulted target. Grating. SPARS10. Various aperture s. Dither patterns. SPARS25, SPARS50, SPARS100. Sub-arrays. Predefined dither patterns. SPARS200, STEP25, STEP50, STEP100. Sub-array. MIF900, MIF1200, MIF1500. IR optics alignment and focus. SMOV special comm andin g for memory dump. Config. SC UVIS UVIS UVIS UVIS UVIS UVIS UVIS UVIS UVIS UVIS UVIS UVIS UVIS UVIS UVIS SC SC UVIS IR UVIS IR IR IR Dur. [hr] 0.11 5.13 1.13 0.39 0.41 0.61 0.15 0.07 0.75 0.13 0.05 0.25 0.23 0.70 0.05 0.79 0.04 0.01 0.13 1.18 0.14 0.43 0.54 0.29 IR IR IR SC 0.20 0.19 0.81 0.08 TIPS May 17, 2007 28 Preparation of SMGT Part 2 SMS Anomaly Subsystem Assignee Status Update UVIS & IR filters (F588NF200LP & F098WF140W) Errors in SIAF aperture specifications APT, IM ESS Closed TRANS OBS/TEL Closed Incorrect reporting: errors for NICMOS; CRSPLIT exp. ANNEAL not used Detector temperature set-points raised for ambient Mosaic dithers generate pointing warnings Tungsten lamp on during readouts Patch SMS for CS 3.0 D limits Error inhibiting 6-stage TEC for ambient tests APT OED Closed N/A IM INS/WFC3 OED Closed Closed APT TRANS IM IM OED INS/WFC3 OED OED Closed Closed Closed Closed TIPS May 17, 2007 29 SMGT VSTIF Results Anomaly Subsystem Assignee Status UVIS full-frame exposures fail in Generic Conversion OPUS/GC DPT Closed Errors in IR dark reference file keyword values CDBS INS/WFC3 Closed Incorrect IR dark reference file selection for subarrays CDBS INS/WFC3 Closed IR subarray format not listed in science image headers KWDB DPT Closed Incorrect WFC3 SIAF values SIAF OBS/TEL Closed Incorrect CCD amp selection for UVIS Quad filters Commanding SCSB Closed Inadequate representation of UVIS Quad filters in use OPUS/GC INS/WFC3 Open Generic Conversion still using old IR readout times OPUS/GC DPT Closed Incorrect PRIMESI keyword values OPUS/GC DPT Closed Incorrect PHOTMODE keyword values CALWF3 INS/WFC3 Closed CALWF3 IR dark frame selection enhancement CALWF3 INS/WFC3 Closed Updated IR dark calibration reference files needed CDBS INS/WFC3 Closed TIPS May 17, 2007 30