Instrument Science Report STIS 96-015 ON-ORBIT FLAT FIELDS AND ABSOLUTE CALIBRATION OF STIS R. C. Bohlin, D. J. Lindler, S. Baum January 1996 ABSTRACT The Space Telescope Imaging Spectrograph (STIS) is a second generation HST instrument with three 2-dimensional detectors: a far-UV CsI MAMA, a near-UV CsTe MAMA, and a visible/near-IR CCD. STIS will obtain science data in both imaging and several spectroscopic modes, so that flat field corrections and absolute photometric calibrations are required for many different instrumental configurations. The basic calibration philosophy is to obtain the bulk of the flat field data from the internal calibration sources, while the deviation of the internal lamps from the external OTA illumination at the focal plane (i.e. “vignetting”) will be measured by external sources in selected modes. Some of these external sources will be standard stars and will also define the absolute photometric sensitivity with the required ~1% internal consistency for all supported science modes and grating settings. 1. INTRODUCTION For a general overview of STIS, see Kimble (1995), Baum, et al. (1995), and Lindler (1995). The calibration requirement in the STIS CEI Specification, Part 2 for the high frequency, pixel-to-pixel flats is to obtain a S/N of 100 per resolution element. The biggest problem with obtaining these high S/N flats is the global count rate limitation of 300,000 counts/s for the STIS MAMAs. The MAMA detectors can be operated in a high resolution mode with 2048x2048 pixels, while the default format is 1024x1024 pixels. Unless otherwise noted, a pixel will mean the default 1024x1024 pixel size. The MAMA resolution element is 2x2 default pixels; so for the best case of uniform illumination with 0.3 counts/s per pixel, the minimum time to get 10,000 counts per resolution element or 2500 counts in each pixel is 2.3 hours. The procedure to obtain MAMA flats in the laboratory is discussed by Ebbets (1994,1995a,1995b,1995c), who tabulates the required flat field calibrations. On orbit, the CCD flats will be determined from independent integrations. However, to avoid excessive 1 exposure time with the internal lamps, most MAMA flats are “manufactured” to indirectly determine the flat field correction file (see 2.2.3). Our expectation is that the high frequency pixel-to-pixel P flats change slowly with wavelength. A grid of measured, P flats with 10% spacing in wavelength means that any intermediate wavelength is within 5% of a flat. If the wavelength of a required flat is within 5% of the wavelength of an existing P flat, then the assumption is that an adequate (<1% added uncertainty) flat can be interpolated from the bracketing flats or that the nearest flat in wavelength is adequate. Table 1 lists the STIS spectral modes that require flats; and Table 2 lists the STIS camera modes requiring flats. The difference #TOT-#PFLAT in Table 1 is the number of MAMA configurations that use manufactured P flats, rather than measured flats. Table 1. STIS Spectral Modes and Flat Field Requirements DET MODEa WL-RANGE CsI 1.1 G140L 1150-1700 1.2 G140M 1150-1700 1.3 E140M 1150-1700 1.4 E140H 1150-1700 1.7X3 X140M 1150-1700 1.7X4 X140H 1150-1700 2.1 G230L 1650-3100 2.2 G230M 1650-3100 2.3 E230M CsTe CCD Å/expb #TOTc #PRIMEc #PFLATc #STARc 1 1 0 0 18 12 4 2 1 1 0 0 11 3 0 0 1 1 0 0 11 3 0 0 1 1 0 0 90 23 19 6 2 1650-3100 800 6 2 0 0 2.4 E230H 1650-3100 267 26 6 0 0 2.7X3 X230M 1650-3100 800 2 2 0 0 2.7X4 X230H 1650-3100 267 6 6 0 0 2.1B G230LB 1750-3000 1 1 1 1 2.2B G230MB 1650-3100 12 11 12 1 ck 3.1 G430L 3050-5550 1 1 1 1 3.2 G430M 3050-5550 286 15 10 15 1 ck 4.1 G750L 5500-11200 4500 2 2 2 2 4.2 G750M 5500-11000 572 16 11 16 1 ck totals 154 93 57 11 55 210 210 150 a. G-First order grating. E-Echelle grating. X-Cross dispersed grating. b. Wavelength range per grating setting for the cases of #PRIME>1. 2 c. #TOT is the total number of allowed grating position settings, each of which require an L flat measurement. #PRIME is the number of steps to get complete spectral coverage with 10% overlap between steps. #PFLAT is the number of positions at which initial high frequency P flat field data will be obtained with the internal lamps. #STAR is the number of settings to be calibrated with a standard star at 10-20 positions along slit. The ck indicates a check of the vignetting assumptions. Table 2. STIS Camera Modes Require 15 Flat Fields DET MODE FILTER BANDPASSa COMMENT CsI 1.6 25 MAMA 1150-1700 clear F25LYA 1216,85 L-α F25SRF2 1280-1700 SrF2 longpass F25QTZ 1450-1700 Quartz longpass 25 MAMA 1150-3150 clear F25SRF2 1280-3150 rejects L-α F25QTZ 1450-3150 Quartz longpass F25CN182 1820,350 medium bandwidth F25CN270 2700,350 medium bandwidth F25CIII 1909,70 CIII] F25MGII 2800,70 MgII 50CCD 2000-10000 clear F28X50LP 5500-10000 optical longpass F28X50OII 3740,80 [OII] F28X50OIII 5007,5 [OIII] CsTe CCD 2.6 3.6,4.6 a. Bandpass is in the form of a range of wavelengths for broad filters and in the form of center, width for narrower filters. 2. PIPELINE CALIBRATION FILE REQUIREMENTS 2.1 Overview The general strategy for STIS flat field and flux calibrations is to separate the process into 4 parts, which can be produced and updated asychronously over time. The four files that comprise these calibrations are: 3 2.1.1 LFLTFILE and PFLTFILE The pixel-to-pixel flat, P or PFLTFILE, corrects the detector net response R for the high frequency non-uniformities and leaves the low and intermediate scale instrumental sensitivity variations unchanged. After application of the P flat calibration, the residual lower frequency sensitivity variations across the detectors are removed by the calibration image L, which is stored in a CDBS file named LFLTFILE. Both files are normalized to unity; and the product PL should have a S/N=100 per resolution element, i.e. S/N=50 per pixel in the default 1024x1024 MAMA format. If the internal krypton, deuterium, and tungsten lamps emit a pure continuum and provide a perfect simulation of the OTA illumination at the focal plane, then the response to these lamps RL can fully determine both L and P. The high frequency variation P and the lower frequency variation L of the flat field are separated to save lamp lifetime. L is a heavily smoothed version of RL and is measured for each grating setting with a short exposure of only S/N~10 per resolution element. The P flats require long exposures; but our assumption is that P flats need be measured for only a small fraction of the grating settings, eg. with a spacing of ~10% in wavelength. See 2.2 for the algorithms that precisely define L and P. 2.1.2 DFLTFILE One delta flat, D, with high S/N=100 per-pixel accounts for potential changes on approximately a monthly time scale for each MAMA detector. D=1 for the CCD modes and at time zero when initial L and P flats are determined. These two delta flats are wavelength and mode independent and account for possible internal gain changes in the microchannel plates inside the two MAMA tubes. Thus, the fully corrected count rate image RC is Rc = R/PDL, (1) where R is the net instrumental response in counts/s, corrected in the pipeline for rate nonlinearities and detector dark response. 2.1.3 PHOTTAB After correction of a data image R by the P, D, and L flats, all pixels in imaging modes are reduced to the same absolute sensitivity, while the spectroscopic modes have the same sensitivity everywhere along the slit direction. For example for a slitless spectrum of a point source, the variation of absolute sensitivity with wavelength is independent of position along the slit direction in the focal plane. This absolute sensitivity, S, will be derived from observations of standard stars and is just one number for imaging or a 1-D vector for spectroscopic modes. Therefore, the final calibrated pipeline output image is F = Rc/(AS) = R/(ASPDL), 4 (2) where the flux, F, for point sources centered between the slit jaws is in the HST standard unit of erg s-1 cm-2 Å-1 and A is the aperture transmission for a point source relative to A=1 for a large aperture. If the aperture width varies along the slit, then A must be a function of location along the slit. Since there are some wavelength regions of low sensitivity where F should be set to zero, the inverse sensitivity (1/S) will be computed for use in the pipeline. Details and transformations to surface brightness are discussed below in 2.2.5. The total calibration image 1/(ASPDL) should be calculated “on the fly” in the pipeline in order to avoid frequent redeliveries of a big file for every mode, since the small files S and L may change more often than the big P file and since there is only one big file D for each MAMA. 2.2 Details, Details 2.2.1 Vignetting Since the response to the calibration lamps RL determines the L and P flats, which correct for all of the STIS optical and detector relative sensitivity variations, any deviation of the lamp illumination from the OTA illumination as a function wavelength, λ, and of position (p,q) in the focal plane must be measured by an external source. This vignetting, V(p,q,λ), in the lamp illumination should be nearly unity everywhere because of the clever design of the cal lamp optical system and the precise polish and parallel construction of the slit jaws. In order to avoid small scale variations >1% in the lamp spectrum along the slit, the smallscale polish on the slit jaws must be constant to 1% of the slit width; eg. a 0.5” slit is 140 microns wide at the focal plane, and the slit jaws must be smooth on the 1.4 micron scale when averaged over the MAMA resolution element of 14 microns (0.05”), as projected to the focal plane. To measure the lamp vignetting V and also account for any smooth largescale variation of the slit width or of any neutral density (ND) filters along a slit, a standard star is observed at ~15 positions along the slit in the focal plane. Observations of the star should be slitless or in a wide slit, in order to avoid any uncertainty due to miscentering, jitter, or variation of the PSF along the slit. For a slit with one axis in the focal plane along the p direction near the center q position, q-cent, V for a spectroscopic mode reduces to a function of 2 parameters R L ( p i, λ )/R L ( λ ) V ( p i, λ ) = ----------------------------------------, R * ( p i, λ )/R * ( λ ) (3) where RL(λ) is the 1-D spectrum of the lamp RL(pi,λ) averaged at the ~15 positions pi of the star along the slit and R*(λ) is the similarly averaged stellar spectrum. The value of RL(pi,λ) at each position is an extraction from the lamp image with an artificial slit of 3-4 5 pixels tall. The continuous function V=V(p,λ) needed to correct the lamp images is a smooth 2-D fit to the measured V(pi,λ) at the ~15 pi discrete positions. The lab calibration results may help set the exact number of stellar positions in the range 10-20 needed to properly sample the vignetting along a slit. The V for a second slit lying at the same position in the focal plane will be the same as for the measured slit, if both slits have a uniform width and ND transmission along their lengths, i.e. provided the only difference is a constant width ratio. If the relative lamp illumination along the slit is constant over time at all wavelengths, then V is independent of mode and requires infrequent determinations. The vignetting should be measured in the .1 modes to get full wavelength coverage with one spectrum; however, the cal lamps are too bright for the MAMAs in wide slits. Perhaps, the best approach is to measure V with 0.5 or 2.0” slits at two extreme wavelength settings in the .2 modes with the hope of confirming the expectation that V should be independent of wavelength, i.e. V(p,λ)=V(p). The x.7 slits are orthogonal to the x.1 and x.2 slits and sample the vignetting V(q,λ) at pcent along the orthogonal axis, q, in the focal plane. However, the widest 0.2” slit may be too narrow; or the cal lamps may be too bright on the MAMAs in all of the 1.7x3, 1.7x4, 2.7x3, and 2.7x4 modes. If the 0.2” slit is uniform enough, the V(q,λ) can be fully characterized down to 1650Å with the CCD detector for the tungsten lamps and for the deuterium lamp in mode 2.1B. The star trail technique of Appendix A could eliminate the need for measuring V for these x.7 slits, as well as for the problematic 45° tilted slits. If V(p,λ)=1 or V(p,λ)=V(q,λ), then the full V(p,q,λ) that is needed for the imaging modes might be assumed to be azimuthally symmetric on the CCD. Otherwise, a star must be stepped over the full field to measure V for imaging modes; or more clever observations of astronomical fields or streak earth flats might suffice (Cox et al. 1987). In general, the cal lamps are too bright for the MAMA imaging modes. See section 2.2.2 for more discussion of V and L for MAMA imaging mode. 2.2.2 LFLTFILE The application of the L flats in the pipeline corrects for the low and intermediate frequency relative-response variations in the STIS images. After dividing the observed image by the L, P, and D flats, the absolute flux calibration or diffuse surface brightness calibration should be the same for sources located anywhere in the field of view. The L flats are derived primarily from images of the calibration lamps with their smooth continuum in the dispersion direction, as corrected by the smoothly varying illumination function V along the slit. The following steps describe the transformation of continuum lamp images RL into the L flats that are required for each grating setting. a) Compute the product image of the L and P flats that is normalized to unity and corrected for any vignetting. 6 LP = {RL/RL(λ)}/V, (4) where RL(λ) is the 1-D average spectrum of the lamp spectrum collapsed along the slit direction. Here, RL(λ) is the full average over all of the 1024 pixels that are illuminated, rather than over the ~15 pi discrete positions used above to measure V. The division by RL(λ) removes the cal lamp signature from the flats in the dispersion direction, just as the division by V removes the lamp signature along the slit. For imaging modes, RL(λ) is replaced by the average of the central ~100x100 pixel region of the flat image. b) Remove the high frequency P flat variations from the product LP by rebinning and filtering. L = <LP> = <RL/RL(λ)>/V, (5) where the < > indicates a 21x21 median filter followed by a rebinning over 4x4 pixels to a 256x256 pixel image. The median filter smoothing eliminates data spikes and residuals of emission lines in the lamp spectra. If the slit jaws are parallel, then the STIS optics should be good enough to maintain constant image quality along the slit; and the technique of dividing the lamp images by the average spectrum RL(λ) should remove the effects of any emission lines in the lamp spectra well enough to recover full L flats under the emission lines. However, the pixelization of the image probably prevents the recovery of P flats under the edges of emission lines (cf. section 2.2.3). Since the detector relative QE variations from center to edge may be greater than the typical pixel-to-pixel P flat variation, a finer grid in wavelength would be needed to predict the large scale L flat variations to the same ~1% accuracy as the P flats with their 10% wavelength spacing. Therefore, an L flat should be measured for every STIS filter in the x.6 modes and every long slit grating setting in the x.1, x.2, and x.7 modes. However, since these flats are smoothed versions of the images, the S/N requirement per resolution element is only 10, which reduces exposure times by 100 relative to the P flat observations. The wavelength of a pixel at the end of the long slit in the x.7X3 modes is less than 2% from the wavelength illuminating the same pixel in the x.3 modes, if the alignment of the slit with the echelle orders is within the tolerance discussed in 2.2.3. The Kr and D2 lamps are too bright for the broadband imaging filters on the MAMAs. Therefore, external targets (Cox et al. 1987) may be required, eg. earth, moon, or a nebula stepped around in the field of view to separate the scene lumpiness from the detector nonuniformities. External sources do not require a separate determination of the lamp vignetting. Other possibilities are the UV tail of the tungsten lamps, the ND filters that cover the full field, or the Pt-Ne wavelength cal lamp at 10 milli-amp. If the lamp V is near unity or has simplifying symmetries and if the filters are spatially uniform, then the best way to get L flats for the imaging modes is to integrate the measured set of medium resolution L flats over the filter bandpass in the same way as the 7 camera mode P flats are determined from medium resolution P flats (see below). If the lamp vignetting is asymmetric, then V must be measured by stepping a star over the entire field of view on something like a 15x15 grid. After a correction for absolute sensitivity in the wavelength direction, the medium resolution L flats are properly normalized along both axes, except for the lamp vignetting. If the low or intermediate frequency variations depend strongly on wavelength within the bandpass of a broadband observation, these L flats will have large uncertainties regardless of the shape of the source spectrum that is used to derive the L flat. The L flat images can be stored and used in the pipeline as the tiny 256x256 binned versions. Bi-linear interpolation is used to resample the L flats, as needed for combination with the larger format P and D flats, eg. the product image LP assumes that the 256x256 L image is resampled to the pixel scale of P. If the detectors are stable and the external stimulus is a high fidelity imitation of the OTA, then any L flats generated in the lab calibrations should be appropriate for reduction of science data obtained on orbit. On orbit monitoring for changes in L flats will be derived from the monthly D flat data and any routinely obtained P flat monitoring. Except for the (unlikely?) event that the normalized lamp vignetting V changes with time, any smooth change in lamp output with wavelength will not effect the L flats, since the mean spectrum is divided out. 2.2.3 PFLTFILE From Eq. 4 and 5, R L /R L ( λ ) P=LP/L=LP/<LP> = ------------------------------- . <R L /R L ( λ )> (6) All images used to derive MAMA P flats will be obtained in hi-res 2048x2048 format, since pipeline access to hi-res flats is required to properly flat field echelle spectra, where the Doppler shift of the spacecraft orbital velocity of ~7 km/s can move the spectrum by two resolution elements (8 hi-res pixels) on the MAMAs. The resampling of the 256x256 denominator image must be to the 2048x2048 size needed in Eq. 6. These measured P flats can be rebinned to 1024x1024 for all cases except manufacturing of echelle mode flats. Since the P flats are derived by dividing the original image RL by the average lamp spectrum and by a smoothed version of itself, both the emission lines and the low order components of the flat field down to the scale of the smoothing length are removed from the pixel-to-pixel flats. However, the pixelization of the image probably prevents the recovery of P flats under emission lines; and the safest assumption is to consider the flats indeterminate for pixels near the discontinuities of the slit edges for any contaminating 8 lines. Whether a P flat with a 10% difference in wavelength is close enough to define the flats under contaminating emission lines or whether additional S/N=100 lamp exposures are required must be determined experimentally. In principle, pixel-to-pixel flats, P, can be obtained directly for each grating setting by using the internal continuum lamps to illuminate a long slit for the spectroscopic first order modes and for the cross disperser in the echelle modes. The lamps are too bright for the broad filters in the camera modes. The cross-dispersed images in modes 1.7X3, 1.7X4, 2.7X3, and 2.7X4 provide echelle flats for the corresponding echelle spectroscopy modes 1.3, 1.4, 2.3, and 2.4, because the echelle orders lie nearly along the slit and perpendicular to the dispersion in the x.7 modes, where the wavelength along the long slit is constant at the value of the center wavelength of the echelle order. In the worst cases of the 1.3 and 2.3 echelle modes, the center to edge coverage in one order is <1% of the center wavelength. An additional 1% wavelength shift between the positions of the same wavelength in the x.7 and x.3 modes would be caused by a misalignment of the x.7 long slits with the echelle spectra by one order at the end of the slit. Because of the long integration time of 2.3 hours minimum and 5-10 hours typically that is needed to obtain MAMA flats of S/N=100 per resolution element, most of the required flats cannot be obtained directly. Instead, for advance planning, the P flat values for each pixel are assumed to be accurate to ~1%, if obtained at a wavelength within 5% of the required wavelength. For the 1.2 G140M mode, P flats must be measured or manufactured for the allowed 12 prime and 6 intermediate grating settings over the central wavelength range of 1173 to 1714Å (Clampin 1995b). The wavelength assignment of any pixel in any of these 18 MSM settings would be within 4.9% of flats measured at central wavelengths of 1229, 1352, 1487, and 1635Å. The standard central wavelength settings of 1222, 1335, 1470, and 1640Å are close enough to these optimum wavelengths to avoid defining new MSM settings. However, since the Kr cal lamp for band 1 has a SrF2 blocking filter, alternate techniques based on thermal vac data to be obtained in mid-1996 must be developed to define STIS flats shortward of ~1280Å. Similarly, the 23 P flats for the 2.2 G230M mode should be adequately determined by internal lamp images in the standard central wavelength settings of 1769, 1933, 2176, 2419, 2659, and 2977, which are close to the optimum 5.3% spacing of 1776, 1970, 2184, 2421, 2684, and 2976, except for the 1970Å optimum case. Whether this baseline set of P flat measurements are too few, too many, or just right must be answered by laboratory and inflight data analysis. In particular, since cathodes near their long wavelength cutoffs are known to be more lumpy in response than at the wavelength of their peak sensitivity, the longest wavelength flats at 1714Å for 1.2 and at 3134Å for 2.2 should be checked for agreement with the standard flat at 5% shorter wavelength at least once in the lab or on orbit. Some investigation is also required to determine whether the unmeasured P flats are best derived by interpolation or from the nearest neighbor in wavelength. These medium resolution flats could also be used to determine P flats for the remaining MAMA modes. However because the dispersion directions are 9 orthogonal, a few corners of the x.7 cross dispersed P flats would have to be made from flats measured at wavelengths differing by up to ~10%. Because the count rates from the Kr and D2 lamps vary with wavelength by ~2 orders of magnitude in the 1.1 and 2.1 modes (MacFeely 1994) and because the Kr spectrum has too many lines to be very useful below ~1250Å, the 1.1 and 2.1 mode P flats must be manufactured from the medium resolution data. Conceptually, even the imaging mode flats could be made by integrating the medium resolution P flat grid over the wide filter bandpasses for each pixel and by choosing the nearest neighbor P flat for the narrow and medium width filters. Whether the 11 P flats for the echelle modes 1.7 and 2.7 or the 11 flats for the imaging modes 1.6 and 2.6 are obtained from the internal lamps, are also manufactured from medium resolution flats, or are derived from earth streak flat data, depends on laboratory confirmation of the validity of the manufacturing technique, how the lab and flight flats agree, STIS lamp operational behavior, internal calibration capabilites when STIS is not the prime instrument, and the stability of the various MAMA P flats. Laboratory data that will address some of these questions should be obtained by early 1996. Since the exposure times are short for all of the CCD modes, direct observation of the lamps will be used to derive all 2.xB, 3.x and 4.x P flats specified in Tables 1 and 2. A minor complication in the manufacturing of P flats from the .2 modes is the lack of data under the fiducial bars of the .05-2” long science slits. These bars of .4 and .86” length comprise only 5% of the image area, which will be flagged as unflatted in the data quality array for science observations that are made before full flats become available. For 1.1 and 2.1 mode observations, slight differences in grating alignment cause flagging of <5% of the science data. If commanding is available, the P flat integrations could be divided with half the time spent at a position that is offset in both axes to get full coverage of the MAMAs under the emission lines and fiducial bars Variability of the P flats in orbit will be routinely checked at one wavelength by the D flats (2.2.4). If any changes are wavelength dependent, the long integration times make updates prohibitive on time scales of much shorter than one year. Perhaps, checks at some wavelengths other than the D flat wavelength should be done on an annual time scale, depending on how much the D flats change and on ancilliary evidence regarding MAMA stability. 2.2.4 DFLTFILE Because internal gain changes in the microchannel plates inside the MAMA tubes are anticipated on relatively short time scales, the flat field correction will be monitored monthly at one wavelength setting with very high S/N. Since this type of detector instability should be wavelength independent, the baseline plan is to measure these delta flats (D flats) for the DFLTFILE pipeline correction at a S/N=100 per pixel and apply the correction to every P flat for the MAMAs. This S/N is twice the S/N of the P flats in order avoid 10 additional noise in the corrected STIS images. Lamp integrations times will be a minimum of 10 hours; and therefore, a wavelength region with nearly constant count rate over each MAMA is required. For example, the medium resolution central wavelengths at 1335 or 1470Å for the 1.2 CsI band and at 2176, 2419, or 2659Å for the 2.2 CsTe band have nearly uniform count rates; and the choice for each band should have the fewest contaminating emission lines. The two designated D flat positions should also have the initial P flat upgraded to D flat S/N to minimize noise in the ratio of D=current/original, which is applied to every MAMA image. If there are lamp illumination or L flat changes from original to current epoch, then the low and intermediate variations must be removed from the D flats in the same way as for the P flats. 2.2.5 PHOTTAB As defined above, the P, D, and L flats are unaffected by changes in the onboard cal lamps, so that the reduced count rate images Rc=R/PDL have the same sensitivity everywhere in the field of view. Since L is defined for slitless spectroscopy of point sources where A=1, the absolute calibration of STIS images is reduced to a one dimensional problem akin to FOS or GHRS. If the absolute flux of a point source must be measured through a slit, then the transmission A of the slit at the location of the star must be known, and the source must be precisely centered on the slit. For FOS, spline nodes are fit to the log of the sensitivity, which is the ratio of the extracted 1-D stellar spectrum to the standard star flux (Bohlin, Lindler, & Keyes 1995). The spline fits should also be appropriate for STIS calibrations but should require few nodes, since the L and P flats should remove all the high and intermediate frequency sensitivity variations. If the MAMA is nearly aligned with the slits, then the high frequency orthogonal MAMA artifacts must be removed from RL(λ) in order to divide by the smooth average lamp spectrum and retain all high and intermediate frequency variations in the product LP of Eq. 4. One solution is to fit RL(λ) with a polynomial after masking the emission lines. The value of RL(λ) in the lines could be recovered by scaling the fit to the peak over the width of each contaminating line. If the contamination is minimal, this fitting technique might be preferable in general. For example, the bumps and dips in Figure 2 with about 0.5% height and ~50 pixel width would require less effort to remove with the L flat than with the sensitivity S. The measure of STIS response to a point source R(λ) is the sum of the count rates above background over all pixels in Rc that lie along the slit and that contribute to the signal from the star. The R(λ) is extracted at every pixel in the dispersion direction. For imaging mode, R(λ) is the sum of the count rates in all pixels that contribute to the stellar signal. The STIS sensitivity for each grating setting is S = R(λ)/(AF), (7) where F is the known flux for the standard star (Bohlin 1995) and where the aperture cor- 11 rection, A, accounts for the loss in transmission through smaller slits (cf. Bohlin & Colina 1995). Since the total count rate at each wavelength does not depend on the PSF, S is the relevent pipeline calibration per pixel in the image, as well as for the extracted spectrum as long as no significant fraction (>1%) of the stellar signal is lost in faint wings. Observations of the standard stars should be done slitless or in wide slits where A=1 in order to avoid any uncertainty due to miscentering or jitter. For imaging, A is also unity. Once the values of A and S are established from observations of standard stars, Eq. 7 is used by the pipeline to produce the flux images, F, in erg cm-2 s-1 Å-1 for all observations. For the medium dispersion modes, the absolute calibrations of the intermediate grating settings are derived by interpolation of the measured sensitivities at the primary positions (1.2p1-1.2p12 and 2.2p1-2.2p19). To achieve this goal, the function S at each primary position is broken into two parts: an absolute sensitivity Sc at the center of the detector and a relative calibration as a function of detector pixel position Sr, where S = Sc Sr and Sr=1 at the center wavelength. The Sc part is a smooth function of wavelength and can be interpolated to ~1% accuracy for grating settings at any arbitrary center wavelength. The Sr part is a family of smooth functions vs. pixel position on the detector, where there is one smooth function for each primary grating setting. The smooth function Sr needed for the arbitrary center wavelength can be linearly interpolated from the two primary position Sr functions that are nearest to the arbitrary center wavelength. See Robinson (1994) for an example of the use of this technique for GHRS. For the echelle modes, the absolute calibration S is defined as a function of the central wavelength for each order and is a one dimensional vector with a constant value to apply to the image along the orders. The echelle ripple is the correction in the dispersion direction relative to the center of the order (Lindler 1994) and will NOT be removed until after the spectrum is extracted into a flux vs. wavelength vector. For mode 1.4, the coverage is sparse with only 3 primary and 8 intermediate settings, so each of the 11 positions will require a separate calibration observation of a standard star or internal lamp. For 2.4, the 6 primary plus 20 intermediate settings may be dense enough to avoid absolute calibration observations for some of the intermediate settings by applying a modeling technique that is analogous to the medium resolution method. If the internal lamps are stable over time to ~1%, absolute calibration Sa for mode a can be determined from a previously calibrated mode b at the same wavelength without need to expend HST integration time on external standard stars. Rc ( a ) S a θ a = S b θ b = -------------Rc ( b ) (8) where θa and θb are the widths of the respective clear slits and the ratio is just the relative 12 response to the same stable lamp, as corrected for any difference in the P, D, and L flats. The Rc at both the a and b positions in Eq. 8 are averages over the respective slit lengths. Since Sa and Sb are smooth spline functions, low S/N images suffice. However, the lamp output is fixed and there is not enough dynamic range in the slit selection to directly relate the x.1 calibration to the x.4 modes without long integration times. For example, the sensitivity difference between the 1.1 and the 1.4 modes is over 200x. For an extended source with continuum emission, the surface brightness, I(contin), in erg cm-2 s-1 Å-1 arcsec-2 is computed from the calibrated flux image by dividing by the solid angle Ω(slit) seen by each pixel, which is the pixel height times the width of the slit, both in arcsec; the width seen is the actual width/cos(45) for the tilted slits. I(contin) = F A/Ω(slit) (9) For an extended source with line emission, the surface brightness, I(line), in erg cm-2 s-1 arcsec-2 is computed by dividing the calibrated flux image by the solid angle of each pixel Ω(px) and multiplying by the dispersion in Å/px, since the response in any pixel does not depend on the entrance slit width as long as the slit is wider than the 0.024” pixel size. I(line) = F A(Å/px)/Ω(px) (10) The aperture transmission, A, that appears in Eq. 2 for the flux F must be removed for the computation of I, because the energy lost from the PSF in a narrow slit is returned from neighboring nebulosity for a uniform diffuse source (cf. Bohlin 1993). These three calibration constants, A, Ω(slit), and Ω(px), for diffuse sources should be placed in the header of the flux calibrated spectroscopic data image. The dispersion in Å/px should also be specified in the header, especially if the sampling interval for the extracted flux vector is not exactly one pixel. For images without dispersion, divide the flux image F by the pixel size in arcsec2 to get I. Table 3 is from Table 2.4 of Kimble (1995, and private comm. for the updates and correction). 13 Table 3. STIS Plate Scales Modea arcsec/px (along dispersion) arcsec/px (cross-dispersion) 1.1,2.1,1.6,2.6 .0244 .0244 1.2,2.2 .0306 .0290 1.3,2.3 .0358,.0332 .0290 1.4,2.4 .0466 .0290 3.1,4.1,3.6,4.6 .050 .050 3.2,4.2 .054 .050 2.1B .050 .050 2.2B .061 .050 a. STIS IDT nomenclature for modes. See Table 1 for translation to STScI names. Modes 1.1 and 2.1 cover the full UV range of 1150-3100Å and provide the best monitors for both lamp changes and for changes in UV sensitivity over time. If the lamps are stable, changes in STIS sensitivity in a few narrow wavelength regions are monitored with the monthly D flats and less frequently obtained P flat data. However, one star from a set of 45 STIS standard stars should be observed in the x.1 modes ~4 times per year to firmly establish and maintain the absolute calibration, to provide complete wavelength monitoring for changes, and to obtain photometric repeatability statistics through the widest slits that are commonly used for science observations. 3. SUMMARY OF EXISTING LAB FLAT FIELD DATA Three MAMA flats with uniform illumination of the CsTe prototype STF2 have been obtained as part of the detector subassembly testing and are summarized in Table 4. After rotating the images to correct a one deg alignment offset of the MAMA with respect to the dispersion direction, adding two bands of zeros at the locations of the .4” and .86” (17 and 36 px) fiducial bars near the top and bottom of the field of view, and then rectifying for an emission line with a 3 deg tilt with respect to the dispersion, an example of the 1800Å image is shown in Figure 1 and should be a good fidelity simulation of a rectified STIS image of the cal lamp spectrum. Three different width emission lines are simulated: 2” (83px at 1800Å), 0.5” (21 px at 2200Å), and 0.05” (2 px at 2800Å). The instrumental resolution is simulated by smoothing the emission lines and the fiducials with a Gaussian of 2.5 px FWHM. The extracted average spectra RL(λ) shown in Figure 2 must be transformed back to the unrotated and unrectified image coordinates in order to derive flats for the original 1024x1024 image according to Eq. 4. In order to avoid artifacts in the derived L flats, the regions of the fiducials and the emission line edges must be interpolated from 14 the median of 21 pixels of data on either side. The derived L flats and the differences among them are shown in Figure 3. Since actual data exist at the locations of the introduced artifacts, the accuracy of the computed L flat from the simulation can be quantatively evaluated under the emission lines. An example of the ratios of the derived L flats to the actual L flats appears in Figure 4, where the errors at the position of the lower fiducial bar approach ~2%. A sample 1800Å P flat appears in Figure 5, while the ratio of the 2800 to the 1800Å P flat is shown in Figure 6. Table 4. STF2 Flat Field Observations Data Set Exposure Time (seconds) Wavelength (Angstroms) Date STF2_HIRES_FLAT_79000_180_031295.DAT 79,000 1800 3/12/95 STF2_HIRES_FLAT_54000_220_031195.DAT 54,000 2200 3/11/95 STF2_HIRES_FLAT_79000_280_031395.DAT 79,000 2800 3/13/95 The statistics of the flat field variations with wavelength are in Table 5. Table 5. Flat Field Statistics in the Three Images 1800Å 2200Å 2800Å Actual Sigma (%) 2.59 2.68 2.88 Minumum 0.93 0.94 0.93 Maximum 1.07 1.07 1.09 Poission statistics (%) 0.47 0.38 0.41 Actual Sigma (%) 1.48 1.45 1.47 Minimum 0.51 0.64 0.64 Maximum 1.14 1.15 1.15 Poission statistics (%) 1.88 1.51 1.64 Actual Sigma (%) 33.4 32.7 33.5 Mimimum 0.17 0.14 0.24 Maximum 2.45 2.65 2.46 L FLATS (256X256 PX) P FLATS (512x512 px) P FLATS (204x2048 px) Since the expected variation from counting statistics is always less than the actual RMS variations, the true detector performance on the bench has been measured. The three low frequency flats show 13-16% variation over the detector from regions of minimum to max15 imum sensitivity. The high frequency P flats for the 512x512 images have an RMS variation per resolution element of 1.5%, while the 2048x2048 mode flats have an RMS of 33%. Because of systematics in the MAMA electronics, the variations in the 2048x2048 mode flats are much worse than the 512x512 variations. One important question is: Can a flat taken at one wavelength be used to correct data taken at a nearby wavelength? Table 6 summarizes the statistics for the ratios of the flats. Table 6. Statistics for the Ratio of Flat Fields Taken at Different Wavelengths 2200/1800 2800/2200 2800/1800 Sigma (%) 1.00 1.76 2.20 Minimum 0.99 0.96 0.97 Maximum 1.04 1.03 1.05 a (%) Sigma 0.64 0.62 0.71 b (%) Sigma 0.21 0.27 0.35 a (%) Sigma 2.84 2.77 2.78 b (%) Sigma 1.50 1.65 1.23 L FLATS (256X256 px) P FLATS (512x512 px) P FLATS (2048x2048 px) a. Includes contribution of counting statistic errors b. Counting statistic error contribution removed. Table 6 demonstrates that the low frequency L flats for each of the three wavelengths are not appropriate for correcting the other two wavelengths (also see Figure 3). The RMS deviations from unity for the ratios are significantly larger than the fraction of a percent scatter expected from the counting statistics. Therefore, a low frequency L flat must be measured at most primary central wavelengths for band 2. The ratios of the high frequency P flats in Table 6 show RMS deviations from unity of less than ~0.3% per resolution element. However, isolated spots (Figure 6) show local deviations of 1-2%, which indicates that a single high frequency flat is insufficient for all wavelengths of band 2. However, the six P flats that are baselined for band 2 in section 2.2.3 may be adequate to achieve flat fielding to 1% per resolution element. 4. CALIBRATION OBSERVATIONS AND EXPOSURE TIMES Since the flat field exposures are INTERNALS and do not consume HST prime pointed observation time, the requirement for a few hundred hours of MAMA integration time on observations of the internal lamps can be achieved in a few months with a duty cycle of 16 ~10 hours per 24 hour day. The specified internal lamp lifetime is 2500 hours. Because of uncertainty in the cal lamp output flux and degradation rates, exposure times and total program time accounting will not be done until after STIS thermal vac testing in mid-1996. In the event that lamps drift badly, some typical exposure time will be estimated for L flat determinations with stars trailed along the slits. 17 APPENDIX A: TILTED 45° SLITS The tilted slits are not supported for cycle 7 but may be calibrated for later HST cycles. For the first order x.1 and x.2 modes, there are 4 slits tilted at +-45° from the dispersion direction to alleviate spacecraft roll constraints, especially in the ecliptic for planets (Kimble 1995, Clampin 1995a). There are only 2 choices of slit width and both choices have a width change along the slit, viz. 0.05 and 0.2” for one slit and 0.6 and 2” widths for the other. In the x.1 modes, the wavelength shifts corresponding to the 12.4” offsets at the edges of the field of view exceed 10% of the central wavelengths. However, the P flats for the tilted slits in MAMA modes 1.1, 1.2, 2.1, and 2.2 should be manufactured adequately from the measured x.2 flats for the orthogonal science slits, since these medium resolution P flats are a set with a 10% wavelength interval at every pixel, which means that every pixel has a measured flat within 5% of any wavelength. Table 7 contains the summary of requirements for full flat field calibration of all allowed configurations of the tilted slits with the first order gratings. Since the totals for these peculiar modes are comparable to the complete list for all other spectroscopic modes in Table 1, a phased implementation of calibrations for subsets of the Table 7 modes is advisable. Table 7. STIS Flat Field Requirements for Each 45° Tilted Slits DET MODE WL-RANGE #TOT #PFLAT #STAR CsI 1.1 G140L 1150-1700 1 0 0 1.2 G140M 1150-1700 18 0 2 2.1 G230L 1650-3100 1 0 0 2.2 G230M 1650-3100 23 0 2 2.1B G230LB 1750-3000 1 1 1 2.2B G230MB 1650-3100 12 12 0 3.1 G430L 3050-5550 1 1 1 3.2 G430M 3050-5550 15 15 0 4.1 G750L 5500-11200 2 2 2 4.2 G750M 5500-11000 16 16 0 totals, each tilt 90 47 8 totals, both tilts 180 94 16 CsTe CCD Notes to Table 7: Summary of allowed configurations for the plus and minus 45° tilted slit pairs (cf. Table 1). 18 #TOT is the total number of allowed grating position settings. #PFLAT is the number of positions that initial high frequency P flat field data will be obtained with the internal lamps. #STAR is the number of settings to be calibrated with a standard star at 10-20 positions along slit. A handle on the vignetting for each of the plus and minus 45° slits can be obtained for the 0.6-2” slits at the wavelength extremes of the 1.2 or 2.2 grating range in the same way as for the normal science slits. (See section 2.2.1). The measurement of the L flats for the 45° slits is not possible with the Kr or D2 lamps in the 1.1 and 2.1 modes because of the 300,000 counts/s global count rate limit. In addition, cal lamp spectra through the tilted slits show both dark stripes due to bars where the size changes and also a bright streak due to overlapping spectra from the failsafe “safety” apertures. Can the x.1 L flats be manufactured from the dense grid of L flats from the x.2 modes for all 3 slit orientations? Since the purpose of the L flat is to remove instrumental differences in response in the spatial direction along the slit and intermediate scale fluctuations in sensitivity in the dispersion direction, the requirement is that any difference in center to edge response to the lamp along the slit is the same to <1% for any differential shift in wavelength at the slit ends. The offset from the center to where the 35.7” long, 45° tilt slits intersect the edge of the 24.9” field of view is 12.4”. The shift in Angstroms at the ends of the 45° slits is (px/arcsec) (Å/px) 12.4” = (Å/px) 510, since the arcsec/px=0.024. These shifts are listed in Table 8 for the relevant MAMA modes. For cases with the same vignetting function, manufacturing of L flats could be investigated. Table 8. Center to Edge Wavelength Shifts for the 45° Tilted Slits Mode 1.1 2.1 1.2 2.2 Å/px 0.60 1.58 .054 .087 Å Shift 310 810 28 44 For the 1.1 and 2.1 tilted-slit and x.7 cross-dispersed L flats, an alternative calibration technique of trailing a star along the slit direction could provide the required L flat data without the need for a separate determination of V. For observations of a point source, the local count rate limit of 50 cts/s applies. To get 100 counts per pixel at the peak response requires 2 sec per pixel, and the whole observation of the 1024 pixels along the slit requires about 2000 s for a calibration star of the proper brightness. The trail rate is 24.9/ 2000 = 0.012”/s and must be constant to 1% on the ~10 px = 20 s time scale for L flat determination. If the jitter file can be synchronized with the trailed data to ~0.1 s, then the illumination variations caused by jitter could be removed on the 0.024” pixel scale. 19 5. REFERENCES • Baum, S., Clampin, M., Hartig, G., Hodge, P., and Kinney, E. 1995, STIS Instrument Mini-Handbook. • Bohlin, R. C. 1993, in Proc Conf. Calibrating HST, p234; also 1994, CAL/SCS-002. • Bohlin, R. C. 1993, FOS Instrument Science Report, CAL/FOS-106. • Bohlin, R. C. 1995, STScI Instrument Science Report on Standard Calibration Sources, CAL/SCS-006. • Bohlin, R. C., & Colina L. 1995, FOS Instrument Science Report, CAL/FOS-136. • Bohlin, R. C., Lindler, D. J., & Keyes, C. D. 1995, FOS Instrument Science Report, CAL/FOS-144. • Clampin, M. 1995a, STIS Instrument Science Report, STIS-008. • Clampin, M. 1995b, STIS Instrument Science Report, STIS-009. • Cox, C. R., Bohlin, R. C., Griffiths, R. E., & Kelsall, T. 1987, Standard Astronomical Sources for HST: 6. Spatially Flat Fields, STScI. • Ebbets, D. 1994, Ball Systems Engineering Report, STIS-CAL-003. • Ebbets, D. 1995a, Ball Systems Engineering Report, STIS-CAL-010. • Ebbets, D. 1995b, Ball Systems Engineering Report, STIS-CAL-011. • Ebbets, D. 1995c, “Prelaunch Calibration Plan for STIS, AV-03”, Ball. • Kimble, R. 1995, “STIS GTO Observers’ Handbook, Rev. A”, GSFC, 5/22/95. • Lindler, D. J. 1995, STIS Science Data Management and Analysis Software Requirments Document”, version 1.0. • MacFeely, K. 1994, Ball Systems Engineering Report, OPT-115B. • Robinson, R. D. 1993, in Proc Conf. Calibrating HST, p291. 20 6. FIGURE CAPTIONS Fig. 1 - Sample MAMA STF2 image at 1800 Angstroms with simulated fiducial bars (black horizontal stripes near the top and bottom) and emission line (vertical white stripe). The image is normalized to unity by dividing by the mean count rate of ~16000 and is geometrically rectified to correct for simulated alignment offsets of the MAMA with respect to the dispersion direction and for simulated tilt of the slits with respect to the dispersion. The rectified image is inserted into a larger image of zeroes to avoid loss of any data. The wedge of black (zeroes) at the top and bottom are from the simulated difference of one degree between the dispersion direction and the MAMA, while the four degree wedges of black on the sides are caused by an additional three degrees of tilt of the slit with respect to the dispersion direction. Fig. 2 - The normalized spectra RL(λ) from the three rectified images after averaging in the direction perpendicular to the dispersion. The original pixel scale has been preserved in order to divide the original spectral images by their averages. The spectra have been linearly extrapolated in the wavelength range where the spatial coverage is not complete. The fine scale pixel-to-pixel noise is <<1%, as required, since the offset between the spatial direction and the MAMA columns removes most of the signature of the detector from RL(λ). If the long slits are aligned with the detector, then some smoothing of RL(λ) is necessary (see section 2.2.5). Fig. 3 - MAMA STF2 low frequency flats for three wavelengths derived according to our algorithm by dividing by the average RL(λ) from Figure 2, binning the raw flat field (HiRes) data to 1024x1024 pixels, filtering with a 21x21 median filter, and binning to 256x256 pixels. The top three images show the computed low frequency flats scaled from 0.93 (black) to 1.07 (white), while the bottom three images show ratios of the low frequency flats with the same intensity scale. Fig. 4 - The ratio of the derived L flat at 1800Å to the actual L flat. The slit causes an error of <1%, because the slit edges contaminate regions that are only the 2.5 pixel width of the STIS resolution. For the fiducial bars, errors approach 2% occasionally, because of the regions without data are 17 or 36 pixels wide. The narrower lower slit is closer to edge, where the smoothing is ineffectual, and shows larger errors in interpolating the L flat across the region of missing data. Fig. 5 - The 1800Å high frequency flat scaled from .95 black to 1.05 white. The regions of the slit edges and fiducial bars that lack measured data are set to unity. Fig. 6 - The ratio of the 2800 and 1800Å high frequency P flats scaled from .98 black to 1.02 white. 21