DOSIMETRY FOR MEDICAL APPLICATION OF IONIZING RADIATIONS: Calibration requirements and clinical applications Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Olivera Ciraj-Bjelac, Milojko Kovacevic, Danijela Arandjic, Djordje Lazarevic Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Department Laboratory for Radiation Measurements Belgrade, Serbia ociraj@vinca.rs Content Metrology and calibration requirements Clinical application Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Global trends in medical exposures Dosimetric quantities and units Dosimetry in diagnostic radiology Medical exposure to ionizing radiation Medical exposure contributes 99% of man-made radiation exposure to humans The concept of risk is used to quantify possible detrimental effects 0.005 0.002 0.002 0.005 mSv Medical Nuclear weapons Occupational Chernobyl 0.61 Atmospheric nuclear tests Total dose from man-made sources of radiation> 0.61 mSv Medical: 0.6 mSv (> 99.97%) Source: United Nations Scientific Committee for Effect of Atomic Radiation (UNSCEAR), 2010 Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Medical exposure to ionizing radiation The role of dosimetry is to determine the amount of radiation received by a person from the radiological examination Dose? Patient dose assessment Establishment of Diagnostic Reference Levels (DRL), optimisation of protection Assessment of x-ray equipment performance Standards of good practice Assessment of radiation detriment Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Dosimetry in diagnostic radiology Global trend 3,6 billion radiological examinations in the period 1997-2007 Significant increase of CT practice: Examination frequency Dose per examination Interventional procedures Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Increase of 50% compared to previous decade Dose to patient 0.02- 0.05 mSv 2 mSv 100CxR 5-20 mSv 400- 1000 CxR 50 chest radiographies= annual natural background radiation dose Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Depends on the examination type Variations for the same type of procedure Dose for the same examination type varies up to 2 orders of magnitude Increased utilization of high-dose procedures CT Interventional procedures Effects Increase of probability for stochastic effects, in particular in the case of the repeated examinations Possible radiation injuries in high-dose procedures Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Problems Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Radiation injuries ICRP 85 10 International Measurements System (IMS) Framework for dosimetry in diagnostic radiology Consistency in radiation dosimetry Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Basic metrology elements Bureau International des Poids et Mesures (BIPM) National Primary Standard Dosimetry Laboratories (PSDL) Secondary Standards Dosimetry Laboratories (SSDL) Users performing measurements Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs International Measurements System (IMS) Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Traceability chain Dosimeters used to determine doses received by individuals Measurements need to be traceable though an unbroken chain of comparisons to national and international standards Traceability is needed to ensure accuracy and reliability Legal and economic implications Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Metrology and traceability The prime function: to provide a service in metrology Designated by the competent national authorities SSDL-Secondary standards, calibrated against the primary standards of laboratories participating in the IMS Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Role of the SSDL Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Journal of the ICRU Vol 5 No 2 (2005) Report 74 Dosimetric quantities in units in diagnostic radiology Basic dosimetric quantity: Air kerma Easy to measure Calibration: Clinical application: Quantities derived from air kerma for different imaging modalities Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Dosimeters calibrated in terms of air kerma Dosimetric quantities Quantities for risk assessment Conversion coefficient for tissue and organ dose assessment Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Basic dosimetric quantities Application specific dosimetric quantities Energy fluence Unit:J/m2 dR da Kerma Unit:J/kg, Gy tr dEtr K dm Absorbed dose Unit:J/kg, Gy en d D dm Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Basic dosimetric quantities Basic dosimetric quantities en K D tr Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Charged-particle equilibrium Absence of bremsstrahlung losses Application specific dosimetric quantities Symbol Unit Equation Incident air kerma Ki Gy Entrance -surface air kerma Ke Gy Ke Ki B Air-kerma area product PKA Gym2 PKA K (x , y)dxdy A Air-kerma length product PKL X-ray tube output Y(d) Gym PKL Kair (z)dz L Gy/As Y (d) Ka (d) / PIt Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Quantity Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Application specific dosimetric quantities: computed tomography Application specific dosimetric quantities: computed tomography Symbol Unit CT air-kerma index Ca,100 (free in air) Gy CT air-kerma index (in standard CPMMA, 100 phantom) Gy Weighted CT air kerma index Normalized weighted CT air kerma index Air-kerma length product Cw Gy Equation C a ,100 1 T 50 50 1 CW C PMMA,100,c 2C PMMA,100,p 3 CVOL CW nCw Gy/As n CVOL PKL Gym K (z)dz NT C W I p CVOL PIt PKL n CVOLj l j PItj j Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Quantity Quantities describing risk Dose-conversion coefficients for assessment of organ and tissue doses c dosimetric quantity normalisat ion quantity Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Organ and tissue dose Equivalent dose Effective dose The Use of Effective Dose (E) E is a risk-related quantity and should only be used in the low-dose range Primary use: Not for: detailed retrospective dose and risk assessments after exposure of individuals epidemiological studies, neither in accidents. In the last cases: organ doses are needed ! ICRP 103, ICRP 105 Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs to demonstrate compliance with dose limits in regulation, for prospective planning of radioprotection Effective Dose in Medical Exposure The relevant quantity for planning the exposure of patients and risk-benefit assessments is the equivalent dose or the absorbed dose to irradiated tissues. E can be of value for comparing doses from different diagnostic procedures similar procedures in different hospitals and countries different technologies for the same medical examination. ICRP 103, ICRP 105 Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs The assessment and interpretation of E is very problematic when organs and tissues receive only partial exposure or a very heterogeneous exposure (x-ray diagnostics) Dosimeters in diagnostic radiology Ionization chambers Semiconductor dosimeters Others Accurate Compact TLD Good energy dependence Energy dependant OSL Design for different application (cylindrical, parallel-plate, different volumes..) PSDL/SSDL Film (radiochromic) Scintillation (kVp meters) user Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Tube voltage 20-150 keV, various A/F combinations, various modalities IEC 61674: Dosimeters with ionization chambers and/or semi-conductor detectors as used in X-ray diagnostic imaging Diagnostic dosimeter: detector and measuring assembly IEC 60580: Dose area product meters Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Dosimetry standards in diagnostic radiology User IEC 61674 Ionization chambers Semiconductor detectors SSDL Ionization chamber of reference class Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Requirements for dosemeters Air kerma: Radiography and mammography Kerma-length product Dosimeters in CT Kerma-area product Radiography and fluoroscopy PPV: kVp meters Frequency: according to national regulations Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Calibrations in diagnostic radiology Calibration in diagnostic radiology Calibrated Quality control Traceability for all beam qualities Auxiliary equipment: electrometers, thermometers, barometers… Environmental conditions Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs SSDL with relevant measurement capabilities General requirements: beam qualities, tube voltage and filtration measurements Dosimeter of reference class (with electrometer) Dosimetry Radiation source Ionization chambers Position system HV supply for monitor and reference class ionization chamber Electrometer X-ray generator, 50-150 kVp, 20-40 kVp Ripple less than 10% for radiography and less than 4% for mammography Beam qualities according IEC 61267 “Shutter” mechanism Filters and attenuators Tube voltage meter (ppv, ±1.5%) Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Equipment Reference class dosemeter Range tube voltage (kV) Intrinsic uncertainty (k=2) Maximum variation of response (%) Unattenuated beam General radiography cylindrical or plane parallel 60-150 3.2 ±2.6 Fluoroscopy cylindrical or plane parallel 50-100 3.2 ±2.6 Mammography plane parallel 22-40 3.2 ±2.6 10 μGy/s10 mGy/s CT cylindrical 100-150 3.2 ±2.6 0.1 mGy/s50 mGy/s Dental radiography cylindrical or plane parallel 50-90 3.2 ±2.6 1 μGy/s10 mGy/s 1 mGy/s500 mGy/s Attenuated beam 10 μGy/s5 mGy/s 0.1 μGy/s100 μGy/s Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Application Type of chamber Range of air kerma rate Spectrum X-ray beam quality: First half-value layer (HVL1) Second half-value layer (HVL2) Homogeneity coefficient: h Tube voltage Total filtration HVL1 HVL2 Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Specification of the x-ray beam Radiation quality RQR Radiation origin Unfiltered beam emerging from x-ray assembly Phantom material No phantom Application General radiography, fluoroscopy, dental radiology Measurements behind the patient (on the image intensifier) RQA Radiation beam from Aluminium an added filter RQT Radiation beam from Copper an added filter CT applications (free in air) RQR-M Unfiltered beam emerging from x-ray assembly Mammography (free in air) RQA-M Radiation beam from Aluminium an added filter No phantom Measurements behind the patient Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Radiation beam qualities (IEC 61267) Typical calibration set up Apertures Focal spot Test point Shutter Additional filtration Monitor chamber Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs X-ray tube window Procedures before calibration (acclimatization, positioning, stabilization…) Calibration procedures (methods, number of measurements, interval between measurements…corrections… Procedures following calibration (uncertainty budget, certificate…) Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Calibration procedures Dosimetry formalism Air kerma: K M Q M 0 N K ,Q0 Reference conditions: set of influencing quantities K M Q M 0 N K ,Q0 ki Influencing condition: quantities that are not subject of mesusremst but have an impact on the result P 273.15 T Air density correction: kTP 0 P 273.15 T0 Beam quality correction: K Q M Q N K ,Q0 k Q ,Qo Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs i Calibrations of dosemeters for CT Information on chamber response only Size on active volume only assumed Far from real situation Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Traditionally, irradiation of the whole volume Contras: Calibration for CT: air kerma length product Cylindrical chamber, 100 mm Non-uniform irradiation Uniform response RQT 9 (120 kVp, HVL: 8.5 mm Al) Focal spot Aperture Monitor chamber Ionization chamber w da dr N PKL ,Q K w / M dr da Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Calibration for fluoroscopy: air kerma area product N PKA ,Q M KAP Ref. chamber Film 10 cm Cekerevac at al, Poster B3 10 cm Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs In laboratory (SSDL) Field calibration M ref N Kref,Qo kQ Anom Calibration in terms of practical peak voltage X-ray tube voltage measurements Practical Peak Voltage (ppv): Invasive or non-invasive measurements Voltage divider Uˆ pU U wW i 1 n i i i pU wU i 1 i i Property of the whole exposure cycle Related to image contrast Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs n Uncertainty budget Uncertainty of the reference standard Uncertainty of user’s instrument Uncertainty due to calibration set up Uncertainty of the evaluation procedure Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Air kerma: ± 2.7 % Air kerma length product: ± 3.0 % Air kerma area product: ± 15 % Non-invasive tube voltage measuring devices: 2.5 % Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Uncertainty budget Goal: minimal uncertainty Assurance and control of traceability Quality manual: technical details, methods, traceability, uncertainty budget, QC, safety…. Continuous improvements and reviews External peer review/audit Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Quality Management System Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Patient dose assessment Clinical dosimetry Output of the X-ray tube, scaled for exposure and geometry Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Direct measurement on patients or phantoms Indirect measurements on patients or phantoms Quantities a) incident air kerma, entrance surface air kerma and kermaarea product (radiography); b) kerma-area product and entrance surface air kerma rate (fluoroscopy); c) incident and entrance surface air kerma (mammography); and d) kerma-length product (computed tomography) KAP BSF Ke Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Dosimetric quantites Patient Real situation Phantoms Objects that simulate real patients in terms of interaction of radiation with matter Easy to perform Standardized Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Patients and phantoms Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Radiography Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Fluoroscopy Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Mammography Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Computed tomography Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Patient dose levels Uncertainty of clinical dose assessment Use of k=2 for expression of uncertainty of dose assessment Typically >10% and close to 25%* *if correction for beam quality and for individual patient is not applied Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Radiographers taking the x-ray images Determining tube output Calculation of individual patient doses Determining dose to an average patient Form measurements towards risk assessment Conversion of measured quantity into organ doses and effective dose Ratio of the dose to a specified tissue or effective dose divided by the normalization quantity Measured using phantoms or calculated using computer models Voxel phantoms based on images of human anatomy Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Conversion coefficients Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Organ dose assesment ICRU 74 Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Optimization of protection Application of DRLs DRLs will be intended for use as a convenient test for identifying situations where the levels of patient dose are unusually high. Quantities that are easily measured! Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Values of measured quantities above which some specified action or decision should be taken Values must be specified Action must be specified Doses to patients from radiographic and fluoroscopic X-ray imaging procedures in the UK—2005 review. HPA RPD-029, HPA; 2007. Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Diagnostic Reference Levels Diagnostic radiology is major contribution to total dose from man-made sources of radiation Dose measurements: population dose assessment, optimization of practice Application-specific dosimetric quantities (patients, phantoms) Calibration of dosimeters in the conditions that are similar to the clinical environment, in terms of air kerma kerma-area product kerma-length product Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs Re-cap Olivera Ciraj-Bjelac, PhD, research associate, dosimetry and radiation physics Milojko Kovacevic, MSc, Head of MDL, radiation physicist Danijela Arandjic, MSc, PhD student, dosimetry and radiation physics Djordje Lazarevic, MSc, PhD student, dosimetry and radiation physics Dragana Divnic, technician Milos Jovanovic, technician Nikola Blagojevic , technician Vinca Institute of Nuclear Sciences Radiation and Environmental Protection Laboratory www.vinca.rs