IAEA Training Material on Radiation Protection in Radiotherapy Radiation Protection in Radiotherapy Part 10 Good Practice including Radiation Protection in EBT Lecture 2: Dosimetry Dose in radiotherapy Is the therapeutic agent Is high - radiotherapy means putting as much dose into the target as possible Carries some risk of severe complications Must be delivered very accurately Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 2 Required dose accuracy Depends on steepness of the dose response curve 5% difference in dose make a 15% difference in tumour control probability in head and neck patients - this is clinically detectable Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 3 Delivery of dose within +/-5% Sources of uncertainty: Absolute dosimetry/calibration Relative dosimetry (%depth dose, profiles, output factors) Treatment planning (estimated uncertainty of the order of +/- 2%) Machine performance on the day (+/- 2%) Patient set-up and movement (+/- 3%) Not much room for error in dosimetry... Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 4 Objectives Understand the principles of beam calibration Appreciate the objectives of clinical dosimetry Identify methods for in vivo dose verification on patients undergoing external beam radiotherapy Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 5 Contents 1. Calibration 2. Clinical dosimetry Beam data acquisition Phantom measurements In vivo dosimetry 3. External audits Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 6 Absolute and relative dosimetry Absolute dosimetry is a technique that yields information directly on absorbed dose in Gy. This absolute dosimetric measurement is also referred to as calibration. All further measurements are then referenced to this standard geometry i.e. relative dosimetry is performed. In general no factors are required in relative dosimetry since it is only the comparison of two dosimeter readings, one of them being in reference conditions. Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 7 1. Calibration Determine absolute dose in Gy at one reference point in the beam Determines the beam on time or the number of monitor units required to deliver a certain dose Very important - if this is wrong, everything will be wrong In the BSS framework part of the optimization of medical exposure Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 8 Optimization of protection in therapeutic exposure BSS appendix II.18. “Registrants and licensees shall ensure that: (a) exposure of normal tissue during radiotherapy be kept as low as reasonably achievable consistent with delivering the required dose to the planning target volume, and organ shielding be used when feasible and appropriate; ... (e) the patient be informed of possible risks.” Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 9 Important note on optimization 1. The dose only to normal tissues shall be kept as low as reasonable achievable 2. In practice, the dose to the target in radical radiotherapy shall be as high as possible to maximize the chances of tumour control The two requirements may be seen at times as incompatible - the key lies in the term “reasonable” What is “reasonable” is a decision which the patient and the clinician must make Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 10 Important note 1. The dose only to normal tissues shall be kept as low as reasonable achievable 2. The dose to the target in radical radiotherapy shall be as high as possible to maximize the chances of tumour control In practice usually the second objective takes precedence in radical treatments - if the tumour cannot be controlled, there is no point sparing normal tissues... One must still protect normal tissues as much as possible... Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 11 Mis-calibration is an important contributor to accident in EBT Calibration of beams Accidents due to mistakes in the determination of the dose rate caused overdosage to as many as 115, 207, 426 patients… by as much as 60% There were other accidents, related to misinterpretation of a calibration certificate, of a reported pressure value for correction, a change of physicist with poor information transfer; a wrong use of a plane-parallel ionization chamber Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 12 Accidents due to calibration mistakes Contributing factors to accidents Lack of understanding of beam calibration, certificates, conversion factors and dosimetry instruments… lack of training and expertise within radiotherapy physics Lack of redundant and independent determination of the absorbed dose (mistakes were not detected) Lack of formal procedures for communication and change of personnel Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 13 Accidents due to calibration mistakes Contributing factors to accidents In one of the cases, for 22 months, there was no verification of the beam; the physicist was devoted to a new accelerator and “ignored” the Co-60 unit (There was a lack of revision of the staff needs when a new accelerator was installed) Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 14 BSS appendix II.19. “Registrants and licensees shall ensure that: (a) the calibration of sources used for medical exposure be traceable to a Standards dosimetry laboratory; …” Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 15 The IAEA/WHO SSDL Network Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 16 Traceability of calibration Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 17 Traceability National Strategy Frequency established by the Regulatory Authority If there is no SDL in the country, the national strategy should include institutional arrangements to facilitate quick import/export and additional arrangements among several countries Redundancy in the calibration of new sources and beams Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 18 BSS appendix II.19. “Registrants and licensees shall ensure that: ... (b) radiotherapy equipment be calibrated in terms of radiation quality or energy and either absorbed dose or absorbed dose rate at a predefined distance under specified conditions, e.g. following the recommendations given in IAEA Technical Reports Series No. 277 [20]; …” Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 19 Calibration Determination of the dose at a reference point - correlation of treatment time or ‘monitor units’ with absolute dose Absolute dosimetry required: Dose = const * Detector Signal Const. must be well known and fundamental: Calorimetry Ionometry W/e Chemical dosimetry g Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 20 Calibration protocols Calibration is a complex process requiring an expert in radiation oncology physics There are many protocols which can provide guidance international (e.g. IAEA TRS 277 or TRS 398) national (usually developed by the national medical physics association) – e.g. AAPM TG 21, AAPM TG 51, DIN 68, ... Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 21 Calibration protocols It is essential to follow ONE protocol It is essential to follow the protocol BY THE LETTER there is no room for error... Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 22 Forms are available Radiation Protection in Radiotherapy Very helpful for guidance Available for most protocols Here shown for IAEA TRS 398 Part 10, lecture 2: Dosimetry 23 Calibration protocols There has been a development from protocols based on calibrations at the national standard lab in air as air KERMA or exposure, to calibration in terms of absorbed dose to water… This development has been in parallel at the IAEA and many national associations (e.g. AAPM) Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 24 Move to absorbed dose to water calibration Follows improved capability of national standard labs Same in US by moving from AAPM TG21 (1983) to AAPM TG51 (2000) Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 25 Which protocol to use? Depends on how the ionization chamber has been calibrated in the standards lab. If one has an air KERMA calibration factor (NK) or an exposure factor (NX), TRS-398 CANNOT be used… If also the dose to water factor (NDw) can be provided by the laboratory, TRS 398 can be used. Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 26 Advantages of absorbed dose calibration The exposure/ KERMA way Easier for the user Less factors required Get NDw directly - only conversion for beam quality required Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 27 A note on calibration The process is beyond the scope of the present course Calibration is a very critical process Calibration (in particular using exposure/KERMA formalism) is complex (>10 factors) It should always be checked by an independent person Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 28 A second note: Calibration can link the absolute dose to a variety of different reference conditions It is essential to know what your reference conditions are. (They are typically linked to the treatment planning system in use) Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 29 BSS appendix II.19. “Registrants and licensees shall ensure that: ... (e) the calibrations be carried out at the time of commissioning a unit, after any maintenance procedure that may have an effect on the dosimetry and at intervals approved by the Regulatory Authority.” The maximum interval in practice for re-calibration is 1 year - less if there is any indication of problems Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 30 Absolute dosimetry Can be done in principle using calorimetry, chemical dosimetry or ionization chambers For radiotherapy practice all protocols are based on ionization chambers Farmer type chamber Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 31 Tools required for calibration In practice a Farmer type ionization chamber - air volume 0.6cc for photons and high energy electrons Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 32 Plane parallel chamber Required for low energy electrons (<5MeV) and recommended for electrons with energy below 10MeV due to steep dose gradients PTW Markus chamber Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 33 Plane parallel chamber 2mm Adapted from Kron in VanDyk 1999 Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 34 Ionization Chamber reading require correction for Air pressure: require an accurate barometer for calibration purposes an error of 10 mBar will give an error of 1% in the calibration Temperature: accurate thermometer an error of 3 degrees centigrade will give an error of 1% in the calibration Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 35 Calibration Records BSS appendix II.32. “Registrants and licensees shall keep and make available, as required, the results of the calibrations and periodic checks of the relevant physical and clinical parameters selected during treatments.” Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 36 2. Clinical dosimetry In the context of BSS, dosimetry has two components: a) dose measurements (dealt with in the present lecture) and b) dose planning discussed more extensively in the fourth lecture of part 10 Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 37 There are multiple objectives for dose measurements in radiotherapy practice Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 38 Roles of clinical dose measurements in radiotherapy Data collection for treatment planning in general Data collection for individual patients Dose verification Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 39 Clinical Dosimetry BSS appendix II.20. “Registrants and licensees shall ensure that the following items be determined and documented: ... (b) for each patient treated with external beam radiotherapy equipment, the maximum and minimum absorbed doses to the planning target volume together with the absorbed dose to a relevant point such as the centre of the planning target volume, plus the dose to other relevant points selected by the medical practitioner prescribing the treatment; …” Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 40 In radiotherapy practice: This means dose measurements are required as as dose determination for the treatment of individual patients input for treatment planning systems Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 41 Dose measurement for individual patients In vivo dosimetry Determination of output for electron cutouts or compensators Assessment of dose distribution in complex treatments (e.g. IMRT) Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 42 Dosimetry as part of commissioning of equipment In the past this has been more the determination of unknown dose rather than verification, however, these days most beam parameters are within tight specifications and known prior to commissioning. Commissioning affects both: Treatment units Treatment planning Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 43 Treatment unit commissioning Aspects: Safety Verification that specs are met Other bits and pieces required for planning Many protocols and guidelines available Usually done using a water phantom and slab phantoms Radiation Protection in Radiotherapy Significant time commitment - however, access is usually not a problem Part 10, lecture 2: Dosimetry 44 Tools for commissioning Mainly scanning water phantom Determine all properties of all radiation beams depth dose, TPR profiles wedges blocks,... Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 45 Phantoms In radiotherapy the term "phantom" is used to describe a material and structure which models the radiation absorption and scattering properties of human tissues of interest. There are many different phantoms for a variety of purposes available in radiotherapy dosimetry. Phantoms are an essential part of the dosimetric process. Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 46 Commissioning of treatment planning Non-dose related components Photon dose calculations Electron dose calculations Brachytherapy Data transfer Compare lecture 4 Special procedures in the present part 10 Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 47 Typical dosimetric accuracy required (examples) From AAPM TG53 Radiation Protection in Radiotherapy Square field CAX: 1% MLC penumbra: 3% Wedge outer beam: 5% Buildup-region: 30% 3D inhomogeneity CAX: 5% Part 10, lecture 2: Dosimetry 48 Typical accuracy required (examples) The required accuracy depends on situation and purpose Uncertainty has two components: dose uncertainty AND spatial localization uncertainty Radiation Protection in Radiotherapy Square field CAX: 1% MLC penumbra: 3% Wedge outer beam: 5% Buildup-region: 30% 3D inhomogeneity CAX: 5% Part 10, lecture 2: Dosimetry 49 Requirements for dosimetry Required accuracy depends on situation and purpose Uncertainty has two components: dose uncertainty AND spatial localization uncertainty Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 50 Clinical Dosimetry is not only applicable to the tumor BSS appendix II.20. “Registrants and licensees shall ensure that the following items be determined and documented: ... (e) in all radiotherapeutic treatments, the absorbed doses to relevant organs….” Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 51 Dose measurements in phantoms Phantoms mimic radiological properties of patients Different complexity from slabs of tissue equivalent material to anthropomorphic phantoms Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 52 Examples for phantoms Slab phantom for consistency measurements IMRT verification phantom Small water phantom for calibration Anthropomorphic head phantom Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 53 Phantoms are available to mimic all aspects of patients and all types of patients Radiation Protection in Radiotherapy Example: Pediatric phantom and CT scans of the phantom Part 10, lecture 2: Dosimetry 54 …but no phantom mimics everything Therefore, one must be aware of the limitations of each material and phantom This means also other - and often cheaper - materials can be used to test a particular property of the radiation beam. Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 55 Clinical Dosimetry BSS appendix II.21.: “In radiotherapeutic treatments, registrants and licensees shall ensure, within the ranges achievable by good clinical practice and optimized functioning of equipment, that: (a) the prescribed absorbed dose at the prescribed beam quality be delivered to the planning target volume; and (b) doses to other tissues and organs be minimized. Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 56 Minimization of dose to normal tissues Optimization of beam direction Beam shaping using blocks or MLC Conformal therapy = conform high dose envelope to target region Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 57 Optimization of beam direction Avoid critical structures - e.g. spine in a lung cancer treatment lung in breast radiotherapy, Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 58 Blocks to avoid lung Radiation Protection in Radiotherapy Could be customized or prefabricated Part 10, lecture 2: Dosimetry 59 Customized shielding Depends on approach and radiation quality Multi Leaf Collimator for megavoltage photons Radiation Protection in Radiotherapy Eye shields for superficial radiation beams Part 10, lecture 2: Dosimetry 60 Organ specific shielding Radiation Protection in Radiotherapy Scrotal shields for megavoltage photon treatments Suitable for scattered radiation not primary beam Part 10, lecture 2: Dosimetry 61 Verification of dose for individual patients Each patient is different Each (radical) treatment is different Routine verification of single fields, e.g. electron output factors, compensator factors Now increasingly verification of full 3D dose distribution for complex high-tech treatments, e.g. IMRT, HDR brachytherapy Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 62 Dosimetry for special procedures Difficult to model in treatment planning Unusual geometry Good patient data is not available Rare occurrence Examples: Most of brachytherapy Total Body Irradiation (TBI) Total electron skin irradiation (TBSI) Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 63 Total body irradiation (TBI) Target: Bone marrow Different techniques available 2 lateral fields at extended FSD AP and PA moving of patient through the beam Typically impossible to do a computerized treatment plan Need many measurements Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 64 TBI: one possible patient position Rice bags Radiation field at >3m FSD; collimator rotated Placed all around body to achieve two distinct separations Breast board Angle of breast board adjusted for individual patients Couch top Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 65 Issues with TBI In vivo dosimetry essential May need low dose rate treatment Shielding of critical organs (e.g. lung) and thin body parts may be required this can be only for parts of the treatment to achieve the best possible dose uniformity Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 66 Total electron skin irradiation Treat all skin to very shallow depth Different techniques available 4 or 6 fields rotating patient Impossible to plan using a computer Requires many measurements for beam characterization Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 67 Total Body Skin Irradiation Multiple electron fields at extended FSD Whole body skin as target Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 68 Many of these applications also benefit from IN VIVO dosimetry ICRU report 24 (1976): “An ultimate check of the actual treatment given can only be made by using in vivo dosimetry.” Single Quality Assurance Activities: Quality Control Hand calculation of treatment time Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry Check source activity 70 Treatment Verification: in vivo dosimetry Treatment verification Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 71 In vivo dosimetry Checks large parts of the treatment chain at once – one detects if something is wrong but not necessarily what the problem is. Good strategy when things are mostly OK and within tight tolerances Requires resources Can prevent accidents Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 72 Why do in vivo dosimetry Quality Control – Treatment Verification Measure because we don’t know Limitations of dose planning Patient movement Verify dose for the record Critical organs Legal aspects Clinical trials Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 73 Methods for in vivo dosimetry Thermoluminescence dosimetry Semiconductors diodes MOSFETs Exit dose measurements portal films electronic portal imaging devices Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 74 Thermoluminescence Dosimeters Small physical size Tissue equivalence (at least some materials) No cables, high voltage or bias required High sensitivity - wide dosimetric range Cheap, reusable Many physical forms and materials available Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 75 Example for TLD in vivo dosimetry: Lens dose measurements TLD detector 7 mm of wax bolus to mimick the position of the lens under the lid TLD detectors lens of eye arangement in AP or PA radiation fields lens of eye arangement in lateral radiation fields Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 76 Semiconductors diodes MOSFETs Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 77 Features of semiconductors + Small + On-line + Easy to use + Small - versatile + Small - arrays - Temperature dependence Cables needed Generally not tissue equivalent possible Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 78 Documentation of all dosimetric measurements? Absolutely essential Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 79 3. Dosimetric audits No one is infallible… Dosimetry may be a difficult and complex task Defense in depth requires redundant check A fresh look from outside can verify dosimetry Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 80 WHO/IAEA photon dose Dose Quality Audit TLD capsules Level 1 Dose Quality Audit: Dose in Reference Conditions FS 10x10, d5cm Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 81 Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 82 Aim: The dose at reference conditions should be the same all over the world 1Gy 1Gy 1Gy 1Gy 1Gy 1Gy 1Gy Radiation Protection in Radiotherapy 1Gy Part 10, lecture 2: Dosimetry 1Gy 1Gy 83 Participation in IAEA/WHO postal dose quality audit Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 84 Results of IAEA/WHO postal dose quality audit Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 85 IAEA/WHO postal dose quality audit Important result: Centres which have participated previously in the audit are significantly less likely to have a deviation of measured from expected dose. Audits are not only a check but also a tool for improvement... Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 86 Prostate Radiotherapy Are we as sure about the correct dose? CTV dose 2Gy CTV dose 2Gy ? CTV dose 2Gy Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry CTV dose 2Gy 87 Level III dosimetric intercomparisons Use of ‘anthropomorphic’ phantom Check entire treatment chain Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 88 Dosimetric Intercomparison Level 1: Absolute calibration at reference point (e.g. IAEA/WHO postal TLD service) Level 2: Include simple physical phantom to gather additional information (e.g. wedge factors, %DD, profiles) Level 3: Check of whole treatment chain using an anthropomorphic phantom (e.g. TROG study) Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 89 Anthropomorphic phantom CAN travel... ART radiotherapy phantom in TROG study (Kron et al. IJROBP 2002) Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 90 Summary Dose determines treatment outcome and should be controlled within 5% Calibration of treatment units must be traceable to national standards and should be performed by qualified experts following appropriate protocols There are a wide variety of tasks and techniques available for clinical dosimetry In vivo dosimetry and external audits are valuable verifications of dose delivery in a radiotherapy centre Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 91 Where to Get More Information Radiotherapy textbooks Calibration protocols Radiation Protection in Radiotherapy Part 10, lecture 2: Dosimetry 92 Any questions? Question: Please comment from your experience on the advantages and disadvantages of radiographic film as a radiotherapy dosimeter. Radiographic film as a radiotherapy dosimeter Advantages Two dimensional Widely available Relatively cheap Provides a dose record Highly sensitive Radiation Protection in Radiotherapy Disadvantages Depends on developing Not very accurate Could be too sensitive Part 10, lecture 2: Dosimetry 95