Supporting & Servicing Excellence GC-MS Gas Chromatography-Mass Spectrometry An Hybrid technique which couples the powerful separation potential of gas chromatography with the specific characterization ability of mass spectroscopy. Overview • • • • • • • • GC History What is GC Key Components Separation Process GC Theory Carrier Gas Injectors Columns GC History • Development of GC (1941) by Martin and Synge • Theory of Capillary GC (1957) by Golay • Capillary GC Instruments (1977) • Fused Silica Capillary Columns (1979) What is GC? • GC is a Separation Technique • Sample is usually a complex mixture we require to separate into constituent components. • Why: usually to quantify some or all components e.g. Pharmaceuticals, Environmental pollutants, etc • Occasionally as a qualitative tool What is the sample? • Usually a mixture of several components • Sample usually introduced as a liquid • Components of interest (analytes) usually in low concentrations (<1% to ppb levels) • Samples dissolved in volatile solvent Comaparison: GC & HPLC HPLC GC • non-volatile samples •volatile & thermally stable • thermally unstable compounds •rapid analysis • macromolecules •good resolution • inorganic and ionic samples • More complex interface to Mass Spec . •easily interfaced to Mass Spec Key components of GC • Hardware to introduce the sample • Technique to separate the sample into components • Hardware to detect the individual components. • Data Processing to process this information. Basic Block Daigram! Separation Process • Sample is introduced into system via hot, vaporising injector. • Typically 1ul injected • Flow of “Carrier Gas” moves vaporised sample (i.e. gas) onto column • Column is coated with wax type material with varying affinity for components of interest • Components are separated in the column based on this affinity. • Individual analytes are detected as they emerge from the end of the column through the Detector. Example Chromatogram (Capillary) :2 WI :4 WI c:\star\examples\level4.run File: c:\star\examples\level4.run Channel: A = TCD Results Last recalc: 25/07/1993 18:35 3.210 mVolts 4.463 1.474 500 5.320 2.038 0.541 Detector Response 0.754 2.853 1.113 750 5.562 250 0 Inject Point -87 1 2 3 4 5 Minutes Time 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. Analysis of Halogenated Pesticides -HCH -HCH -HCH Heptachlor -HCH Aldrin Heptachlor epoxide Endosulfan I 4,4’-DDE Dieldrin Endrin 4,4’-DDD Endosulfan II 4,4’-DDT Endrin aldehyde Endosulfan sulfate Methoxychlor Endrin ketone 4 10 7 • 11 6 8 12 13 9 14 1 5 3 15 16 2 2ppb in Water 17 18 Chromatogram GC Step by Step • Carrier Gas • Injector • Column – Capillary – Stationary Phase • Detectors – Mass Spectrometer Carrier Gas Inert Helium Choice dictated by detector, cost, availability Pressure regulated for constant inlet pressure Flow controlled for constant flow rate Chromatographic grade gases (high purity) Column Types Capillary Columns Length: 10m to 100m Diameter: 180um, 250um, 320um & 530um I.d Packed Columns Length: <2m Diameter: 1/8” & ¼” OD Typical column flow rates • Capillary Column Flow – 250 um 1 ml/min – 320 um 1.5 ml/min – 530um up to 2.0 ml/min Purpose of Injection • Deposit the sample into the column in the narrowest band possible • The shorter the band at the beginning of the chromatographic process - tall narrow peaks • Gives maximum resolution and sensitivity • Therefore type of injection method and operating conditions is critical in obtaining precise and accurate results Splitless injector Design Graphite/Viton Seal •Reduced Sample Contact “Unique”Dual Split Vent design •Improved Precision Large Internal Volume •Minimum Solvent Tailing •More Efficient Sweep Shortened Capillary Guide •Minimal Cold Spots •Minimal Upswept Volume Cross Section of PTV Injector Modern Temperature Programmable Injector (Varian 1079) Programmable Temperature Vapourising Injector Split & Splitless Injection • Most common method of Injection into Capillary Columns • Most commonly misunderstood also! • Same injector hardware is used for both techniques • Electronically controlled Solenoid changes Gas Flow to determine Injector function. Split Injection • Mechanism by which a portion of the injected solution is discarded. • Only a small portion (1/1000 - 1/20) of sample goes through the column • Used for concentrated samples (>0.1%) • Can be performed isothermally • Fast injection speed • Injector and septa contamination not usually noticed Splitless Injection • Most of the sample goes through the column (85-100%) • Used for dilute samples (<0.1%) • Injection speed slow • Should not be performed isothermally • Solvent focusing is important • Controlled by solenoid valve • Requires careful optimisation On Column Injection • All of the sample is transferred to the column • Needle is inserted directly into column or into insert directly above column o Trace analysis o Thermally labile compounds e.g Pesticides, Drugs o Wide boiling point range o High molecular weight Large Volume Injection • To enhance sensitivity in Envoirnmental applications. • Uses 100µL syringe: Inject up to 70 µl • Very slow injection with injector temperature a few degrees below solvent boiling point, split open, flow at about 150 mls/ min • Solvent vents out of split vent, thus concentrating the analytes • Close split • Fast temperature ramp to top column temperature +20°C • Column programming as per sample requirements Columns Material of Construction • Metal (1957) • Glass (1959) • Fused Silica (1979) • Aluminium Clad (1984) • Inert Metal (1990) Capillary Column Characteristics • Length (10M - 50M) • Internal Diameter (0.1mm - 0.53mm) • Liquid Stationary Phase • Film Thickness (0.1um - 5um) • Polarity (Non-polar - Polar) Stationary Phases Choice of phase determines selectivity Hundred of phases available Many phases give same separation Same phase may have multiple brand names Stationary phase selection for capillary columns much simpler Like dissolves like Use polar phases for polar components Use non-polar phases for non-polar components Column Bleed Bleed increases with film thickness Polar columns have higher bleed Bleed is excessive when column is damaged or degraded Avoid strong acids or bases Adhere to manufacturer’s recommended temperature limits Avoid leaks Choosing a Column • Internal Diameter • Film Thickness • Length • Phase Internal Diameter, Smaller ID’s • Good resolution of early eluting compounds • Longer analysis times • Limited dynamic range ID Effects - larger ID’s • Have less resolution of early eluting compounds • Shorter analysis times • Sufficient resolution for complex mixtures • Greater dynamic range Film Thickness Amount of stationary phase coating Affects retention and capacity Thicker films increase retention and capacity Thin films are useful for high boilers Standard capillary columns typically 0.25µm 0.53mm ID (Megabore) typically 1.0 - 1.5µm Column Capacity The maximum amount that can be injected without significant peak distortion Column capacity increases with : film thickness temperature internal diameter stationary phase selectivity If exceeded, results in : peak broadening asymmetry leading Length effects - isothermal analysis • Retention more dependant on length • Doubling column length doubles analysis times • Resolution a function of Square Root of Length • Gain 41% in resolution • Is it worth the extra time and expense?- Length effects - programmed analysis • Retention more dependant on temperature • Marginally increases analysis times • Run conditions should be optimised Summary - Effect of ID, Film Thickness, and Length • • • • • ID Film Thickness Length Choice based on Gain in resolution is capacity and resolution • Thick film for low not double Use 0.25mm for MSDs boilers Isothermal: tR L Use 0.32mm for • Thin film for high Programmed: tR is split/splitless & DI boilers more dependent on Use 0.53mm for DI & • Thicker films for larger temperature purge & trap ID's Detectors Overview • Basic Mass Spectrometry Theory • Types of Ionisation - Electronic Ionisation - Chemical Ionisation • Interpretation of Mass Spectra • Ion Trap Theory • Components of the Ion Trap Ion Trap Mass Spectrometry Basic Mass Spec.Theory • Mass Spec. is a Microanalytical Technique used to obtain information regarding structure and Molecular weight of an analyte • Destructive method ie sample consumed during analysis • In all cases some form of energy is transferred to analyte to cause ionisation • In principle each Mass Spectrum is unique and can be used as a “fingerprint” to characterise the sample • GC/MS is a combination technique that combines the separation ability of the GC with the Detection qualities of Mass Spec. Basic GCMS Theory(1) • Sample injected onto column via injector • GC then separates sample molecules • Effluent from GC passes through transfer line into the Ion Trap/Ion source • Molecules then undergo electron /chemical ionisation • Ions are then analysed according to their mass to charge ratio • Ions are detected by electron multiplier which produces a signal proportional to ions detected Basic GCMS Theory(2) • Electron multiplier passes the ion current signal to system electronics • Signal is amplified • Result is digitised • Results can be further processed and displayed Types of Ionisation • Electron impact ionisation • Chemical Ionisation Definition of Terms Molecular The ion obtained by the loss of an electron from ion the molecule The most intense peak in the MS, assigned 100% Base peak intensity M+ Symbol often given to the molecular ion Radical cation +ve charged species with an odd number of electrons Lighter cations formed by the decomposition of Fragment the molecular ion. ions These often correspond to stable carbcations. Electron Ionisation(1) • Sample of interest vaporised into mass spec • Energy sufficient for Ionisation and Fragmentation of analyte molecules is acquired by interaction with electrons from a hot Filament • 70 eV is commonly used • Source of electrons is a thin Rhenium wire heated electrically to a temp where it emits free electrons Electron Ionisation Electron Ionisation • The physics behind mass spectrometry is that a charged particle passing through a magnetic field is deflected along a circular path on a radius that is proportional to the mass to charge ratio, m/e. In an electron impact mass spectrometer, a high energy beam of electrons is used to displace an electron from the organic molecule to form a radical cation known as the molecular ion. If the molecular ion is too unstable then it can fragment to give other smaller ions. The collection of ions is then focused into a beam and accelerated into the magnetic field and deflected along circular paths according to the masses of the ions. By adjusting the magnetic field, the ions can be focused on the detector and recorded. Chemical ionisation • Used to confirm molecular weight • Known as a “soft” ionisation technique • Differs from EI in that molecules are ionised by interaction or collision with ions of a reagent gas rather that with electrons • Common reagent gases used are Methane , Isobutane and Ammonia • Reagent gas is pumped directly into ionisation chamber and electrons from Filament ionise the reagent gas Chemical Ionisation(2) • First - electron ionization of CH4: – CH4 + e- CH4+ + 2e• Fragmentation forms CH3+, CH2+, CH+ • Second - ion-molecule reactions create stable reagent ions: – CH4+ + CH4 CH3 + CH5+ – CH3+ + CH4 H2 + C2H5+ • CH5+ and C2H5+ are the dominant methane CI reagent ions Chemical Ionisation(3) • Form Pseudomolecular Ions (M+1) – CH5+ + M CH4 + MH+ – M+1 Ions Can Fragment Further to Produce a Complex CI Mass Spectrum • Form Adduct Ions – C2H5+ + M [M + C2H5]+ – C3H5+ + M [M + C3H5]+ M+29 Adduct M+41 Adduct • Molecular Ion by Charge Transfer – CH4+ + M M+ + CH4 • Hydride Abstraction (M-1) – C3H5+ + M C3H6 + [M-H]+ » Common for saturated hydrocarbons EI vs CI for Cocaine analysis • EI Spectrum of Cocaine • Extensive Fragmentation • Molecular Ion is Weak at m/z 303 Methane CI of Cocaine Pseudomolecular Ion and Fragment Ions Proton Affinity • Proton Affinity Governs CI Susceptibility • The higher the affinity the more tightly bound the proton is to the parent species • The greater the difference in proton affinities between the analyte and reagent gas the more energy transferred to the protonated molecule –more fragmentation Interpretation of Mass Spectra(1) Intepretation of Mass Spectra(2) •The MS of a typical hydrocarbon, n-decane is shown above. The molecular ion is seen as a small peak at m/z = 142. •Notice the series ions detected that correspond to fragments that differ by 14 mass units, formed by the cleave of bonds at successive -CH2- units Interpretation of Mass Spectra(3) Interpretation of Mass Spectra(4) •The MS of benzyl alcohol is shown above. •The molecular ion is seen at m/z = 108. •Fragmentation via loss of 17 (-OH) gives a common fragment seen for alkyl benzenes at m/z = 91. •Loss of 31 (-CH2OH) from the molecular ion gives 77 corresponding to the phenyl cation. • Note the small peaks at 109 and 110 which correspond to the presence of small amounts of 13C in the sample (which has about 1% natural abundance). Determining Isotope Patterns in Mass Spectra •Mass spectrometers are capable of separating and detecting individual ions even those that only differ by a single atomic mass unit. •As a result molecules containing different isotopes can be distinguished. •This is most apparent when atoms such as bromine or chlorine are present (79Br : 81Br, intensity 1:1 and 35Cl : 37Cl, intensity 3:1) where peaks at "M" and "M+2" are obtained. •The intensity ratios in the isotope patterns are due to the natural abundance of the isotopes. •"M+1" peaks are seen due the the presence of 13C in the sample. Isotope Patterns 2,Chloropropane •Examples of haloalkanes with characteristic isotope patterns. •The first MS is of 2-chloropropane. •Note the isotope pattern at 78 and 80 that represent the M and M+2 in a 3:1 ratio. •Loss of 35Cl from 78 or 37Cl from 80 gives the base peak a m/z = 43, corresponding to the secondary propyl cation. •Note that the peaks at m/z = 63 and 65 still contain Cl and therefore also show the 3:1 isotope pattern. 1,Bromopropane • The second MS is of 1-bromopropane. • Note the isotope pattern at 122 and 124 that represent the M amd M+2 in a 1:1 ratio. • Loss of 79Br from 122 or 81Br from 124 gives the base peak a m/z = 43, corresponding to the propyl cation. • Note that other peaks, such as those at m/z = 107 and 109 still contain Br and therefore also show the 1:1 isotope pattern. ION TRAP THEORY • Ionize analytes within the ion trap – Use energetic electrons to ionize • Store ions and continue to ionize until the optimum trap capacity is reached – Optimum ion time calculated by software • Increase the voltage on the Ring Electrode of the ion trap to scan ions out in order from low to high mass – This voltage-time relationship called the EI/MS Scan Function • Store the mass-intensity information as a mass spectrum Electron Ionization Happens Inside the Ion Trap Filament Filament Gate Gate Ring electrode Ring Electrode Trapped IonsIons Trapped CARRIER Analytes + He GAS Carrier Gas Mass Spectrum of Toluene Mass Spectrum of Caffeine Mass Spectrum of Glycerin Mass Spectrum of Cholesterol Mass Spectrum of Aspirin