Chapter 25 HPLC High-Performance Liquid Chromatography 1 Typical HPLC 2 Application Microdialysis 3 Monitor Aspirin in Blood 4 Increased Efficiency • In any method the object is to increase mobile phase / stationary phase interaction. • Decrease particle size. – Better packing – slower flow. 5 6 How Does the Smaller Size Help 7 Smaller is Better • N ~ 3500 L (cm) / dp (mm) • Smaller particle size leads to – Higher plate number – Higher pressure – Shorter optimum run time 8 9 Size Considerations 10 What is a bar? • A measure of pressure. • One bar = 1.01325 Atm • One bar is 100 000 Newtons/m2 • Under water you gain about one atm for each ten meters in depth. 11 HPLC Columns 12 Stationary Phase / Support • Support is the scaffolding that the stationary phase sits on. • Support - Microporous Silica – – – – Solvent can get inside Surface area of 100’s m2 / gram Silica degrades above pH 8 so keep pH below. Special supports have been developed for higher pH 13 Active Sites Generally Bad 14 Active Sites Can Lead to Tailing 15 We Can Modify the Surface 16 Bonded Phases • SiOH + ClSi(CH3)2 - R = Si-O-(CH3)2 - R • Change surface polarity – – – – – – Amino (CH2)3NH2 Cyano (CH2)3C#N Diol (CH2)2OCH2CH(OH)CH2OH Octadecyl (CH2)17CH3 Octyl (CH2) 7CH3 Phenyl (CH2)3C6H5 17 New Technology Monolithic Silica Columns 18 Micrographs 19 Elution • Competition for the surface. • Which substance has the greater surface affinity – a solvent with a higher affinity will push the analytes along. 20 21 The Eluotropic Series 22 Isocratic and Gradient Elution • Isocratic - same composition • Gradient - composition changes over time • Method Development. • A mixture of compounds. Separated with acetonitrile (B) and aqueous buffer (A) • 1) benzyl alcohol 2) phenol 3) 3’,4’-dimethyloxyacetophenone 4) benzoin 5) ethyl benzoate 6) toluene 7) 2,6-dimethoxytoluene 8) omethyoxybiphenyl 23 24 25 26 Too Long - Over Two Hours! • We can do a gradient. Examine when we get the best separations and then change composition over time. 27 28 Separation Design 29 Separation Design 30 Practical Issues • Solvents – Must be very pure, lack UV absorbing species – Should be filtered – Guard columns should be used to protect column from strong absorbing compounds – Solvents should be degassed – bubbles and oxygen (sparging) – Normal phase solvents should be 50% saturated with water – Gradients in reversed phase can require 10 – 20 column volumes to return to starting conditions • Add 3% 1-propanol to each solvent and you will cut this to 1.5 empty volumes 31 Reducing Waste Solvent (Save some money – be a hero) • Shorter columns with smaller particles • Switch from 4.6 mm to 3.0/2.0 mm id columns • In isocratic systems – use an electronic recycler. 32 Quality Assurance • Inject a QC sample each day to insure that you have consistent peak shapes and retention times. – Keep a log of your column performance 33 Symmetric Band-shape • Asymmetry Factor A/B should rarely be worse than 0.9 to 1.5. • Tailing a bigger issue and fronting generally. – – – – – – Amines interacting with support active sites. Add 30 mM TEA Acidic compounds Add 30 mM Ammonium Acetate Or for mixes or unknowns – 30 mM triethylammonium acetate. Persistent problem – add dimethyloctylamine or dimethyloctylammonium acetate These take a long time to wash on on changing mobile phases. 34 Other issues • Voids can develop at the column inlet. – Repack with fresh stationary phase to get rid of this problem. (Might want to get new column) • Columns should be washed to get rid of salts and strongly adsorbed compounds • Frits should be cleaned. Back wash or replace • Samples should be dissolved in mobile phase or a weaker solvent. 35 36 Overloading • Care should be taken not to overload the injection on the column. – Inject 10x less and see if peaks look any better. • Reversed Phase can deal with 1 to 10 mg sample per gram of silica. (About 10 cm on a 0.46 mm column) • Injection volume – < 15% of the peak volume at baseline 37 Minimize Dead Volume • Minimize connection tubing • Ensure that fittings are proper matches. 38 Injection and Detection • Pumps – Piston Type under program control. Up to 400 bar (40 MPa or 6000 psi). Gradients made by proportioning valves. 39 40 Injection • Sample loop filling either done manually or by and autoinjector 41 Detectors Mass or Concentration Types 42 UV- Vis (Photometric) Variable Wavelength Detector 43 Photodiode Array All wavelengths at once 44 45 46 Refractive Index Universal Detection with major issues. 47 Evaporative Light Scattering • • • • • Solutes less volatile than mobile phase Light scattered from mass of analyte. Poor linearity – polynomial calibration Good with gradients. No solvent front. Same buffers as used with mass spec. Acetic acid, formic acid and TFA, ammonium acetate, diammonium phosphate, ammonia or TEA. 48 49 Electrochemical • Analytes that can be oxidized or reduced – Phenols, aromatic amines, peroxides, mercaptans, ketones, aldehydes etc. • Can be very sensitive but are difficult to work with. 50 Method Development 51 Eluent Strength Nomogram 52 Gradient Separations • Same purpose as temperature or pressure programming in GC. Speed up analyses. • Increase the eluent strength as the run progresses. That is increase the amount of organic phase as the run progresses in reversed phase methods. 53 Dwell Time • Time it takes for the composition change at the pump to reach the head of the column. • Important on method transfer • Can be determined by running a gradient of 0.1% acetone (detect at 260 nm) without a column. Start gradient at time 0 and see when the acetone starts to be absorbed. 54 55 Since gradients require a post equilibrium time then perhaps isocratic will work better. • What is the span of all the peaks. That time span is Dt – If Dt/tg > 0.25 then use a gradient. If less you can make an isocratic system. (See next slide) 56 57 Optimize Gradient • Run a broad gradient. (5% to 100% over 40 to 60 min) • Eliminate gradient before first peak and after last peak. Run over same time. • Cut time if above works well to save time. 58 59 60 61 62