OBJECTIVE Without reference, identify basic facts about the clinical applications of pulse oximeters with at least 70 percent accuracy. Purpose Provides continuous, noninvasive monitoring of patient oxygenation Provides rapid indication of a patient’s changing level of oxygenation Determines arterial blood oxygen saturation (SpO2) using spectrophotometric oximetry principles • Spectrophotometric Oximetry is the measurement of blood oxygen concentration using an instrument that makes measurements based on the comparison of light being output to the amount received. The more light received the less that was absorbed by the blood. Determines pulse rate using plethysmographic techniques Principles Differential light absorption is used to determine the percent of oxygen saturation of hemoglobin in arterial blood Hemoglobin - an iron-containing compound found in red blood cells that carries oxygen from the lungs to the body tissues Two different wavelengths of light Red Infrared Emitted from a probe Passes through a pulsating arterial bed Fingertip Earlobe Forehead To a photodetector • Light absorption Absorption characteristics » Oxygenated hemoglobin absorbs less than deoxygenated hemoglobin at the red wavelength » They are more similar at the infrared wavelength Each pulse of arterial blood causes small arteries and arterioles to expand and contract, this varies the amount of light absorbed by the arterial blood A portion of the passing light is absorbed by tissue constituents » Venous blood » Muscle » Cartilage » Bone The absorption due to these constituents is constant allowing the microprocessor to eliminate them from the calculations Plethysmographic technique Plethysmogram corresponds to the patient's pulse waveform This signal is used to calculate the patient's pulse rate, and is determined from the peaks of the arterial blood waveforms • Sensor probes Transmittance probes - light from each LED is passed through the measurement site to a single photodetector on the opposite side Reflectance probes - light scattered along the tissue surface is collected by a photodetector adjacent to the LED In both probes, LED's alternately pulse on and off This allows differentiation between red and infrared light at the photodetector OBJECTIVE Without reference, identify basic facts about the clinical applications of patient monitoring systems with at least 70 percent accuracy. Intended Purpose -To watch or monitor a patient's vital signs and display waveforms and/or numerical data Vital signs Heart rate and ECG When the heart contracts an electrical signal can be detected by electrodes placed on the patient's chest and extremities This electrical signal can be plotted as a function of time, and the resultant waveform is referred to as an electrocardiogram (ECG) • Pulse • Temperature • Blood pressure Systolic - the peak blood pressure felt in the circulatory system Diastolic - the resting pressure of heart Mean - average of the systolic and diastolic Respiration Arterial O2 saturation Cardiac Output Airway CO2 Concentration The medical and nursing staff analyze these physiological parameters To reveal changes in patient's condition To determine proper treatment Typical locations Acute care units Intensive care units Cardiac care units Surgery and recovery units Description of Monitoring Systems Bedside monitor mainframe Has a display and possibly a printer Can be switched with any bedside monitor Has ports for different modules or different modules are built in Modules of a bedside monitor ECG module Amplification of the cardiac signal is performed within the ECG module Displays electrical activity of the heart by waveform presentation on CRT Can display oHeart rate - determined from the R to R interval oHeart irregularities Heart rate alarms Tachycardia - heart rate is too fast Bradycardia - heart rate is too slow Blood pressure module To monitor patients when fluids are being lost; in burn victims or major surgery To monitor patients when fluids are gained; during infusions of blood or other fluids To detect hypotension (low blood pressure) which can lead to vascular collapses because of hypovolemia (low blood volume) To detect hypertension (high blood pressure) which can overload the heart because of hypervolemia (high blood volume) Temperature module (body temperature) Usually measured by means of a thermistor probe Probe is inserted in mouth, armpit, or rectum Respiration module Methods » Most common - impedance pneumography - measurement of the change of impedance across patient's chest during respiration Pressure sensitive capsules placed on abdomen to detect body movement Thermistor near mouth or nose to measure change in temperature between inhaled and exhaled breaths May also include an apnea alarm Carbon dioxide (CO2) module • Measures CO2 concentration at the end of an exhaled breath • Two types Capnograph » Measures the increase and decrease in CO2 during each inspiratory / expiratory cycle » Displays both CO2 waveform and numerical data Capnometer Continuously measures CO2 Displays only numerical data When connected to a patient monitor, becomes a capnograph Pulse oximeter module Noninvasive and continuous means of monitoring percent of O2 saturation (SaO2) of arterial blood Reduces the need for arterial puncture and blood gas analysis Physiological monitors are often equipped with 2 types of alarms System faults Loose electrodes Defective electrodes Physiological parameters have exceeded the limits set by the operator OBJECTIVE Without reference, identify basic facts about the clinical applications of electrocardiograph units with at least 70 percent accuracy. Intended Purpose To detect the electrical activity of the heart and produce an electrocardiogram (ECG) which is a graphic record of voltage versus time To diagnose cardiac abnormalities To monitor patient's response to drug therapy To revel major changes in heart rate and cardiac rhythm (ECG disturbances) Pericarditis - inflammation of the sac containing the heart Atria and ventricular hypertrophy enlargement f the walls due to obstruction Myocardial infarctions - coagulation in the muscular tissue of the heart resulting from obstruction of circulation Ventricular fibrillation Asystole Electrolyte concentrations and acid base balance Increased metabolic activity Drug reactions Hypoxemia - low oxygen content in the blood Hypothermia - low body temperature ECG monitors typically measure and display up to three physiological parameters Electrocardiogram (ECG) Heart rate Body temperature or respiration Elements of the ECG P-wave - represents depolarization of both atria Begins with electrical impulse from the SA node Impulse spreads in wave-like fashion, stimulating both atria Both atria depolarize (contract) and produce electrical activity PR Segment Electrical impulse from atria passes to the AV node There is a 1/10 second pause allowing blood to enter the ventricles The AV node is depolarized Duration - .12 - 2.0 seconds Measures from the onset of the P-wave to the onset of the QRS complex QRS complex Represents the electrical impulse as it travels from the bundle of HIS into the bundle branches into the Purkinje fibers and into the myocardial cells (causing ventricular contraction) The depolarization of the ventricles Duration - .08 - .12 seconds Consists of: Q-wave First down stroke of the QRS complex Not always present R-Wave - first upward deflection of the QRS complex S-Wave - first downward stroke after the R-wave ST segment Used to identify myocardial infarctions Serves as the isoelectric line from which to measure the amplitudes of other waveforms J-point - junction between the QRS complex and the ST segment T-wave - represents the repolarization of the ventricles Principles of Operation The heart rate is determined by the R to R interval of successive QRS waves Electrocardiographs record small voltages (about 1mv) that appear on the skins surface as a result of cardiac activity by using various electrode configurations discussed earlier Multichannel electrocardiographs Operate similarly to single channel units in that the user selects certain similar functions: Automatic or manual lead switching Signal sensitivity Chart speed Unlike signal channel units, multichannel units have some advantages: Record 3 or more leads simultaneously Tracings can be held in memory Tracings are printed out in a one-page format Modes Manual mode » User selects three leads to be recorded » Unit traces signal from these leads until others are selected Automatic mode » Each standard 12 leads are recorded for a preset time period » Switching from one lead to another occurs automatically » Some units can be programmed to record tracings from any lead sequence Units with integral timers and selectable chart speeds can also be programmed for » Stress testing » Trending » Rhythm monitoring Semi-automatic mode » Recorder scans through first lead group Semi-automatic mode » Recorder scans through first lead group » Then switches to observe mode o Allows user to preview next lead group for signal quality before recording o User restarts recording of next group manually • Lead hold feature Overrides programmed timer Allowing longer recording time for a particular lead Sensitivity setting User selectable Determines size of the recorded ECG waveform ECG signals that become too large and produce a waveform that goes over scale (arrhythmic beat) » Most units will automatically switch to lower sensitivity setting immediately Other units allow user to choose between recording » Affected channel at lower sensitivity » All channels at lower sensitivity » Waveform as is, followed by rerecording at lower sensitivity Frequency response Factory-set to detect ECG signals between 0.05 and 100 Hz for diagnostic purposes Electrical interference also occurs within this range producing artifacts by recording » Muscle movement » Line power frequency To reduce such interference, notch filters can be selected to block these frequencies » Because filters limit frequency response, they can affect diagnoses based on certain details (amplitude) » Therefore, they are not usually used for diagnostic recording ECG tracings Single-channel units require cutting and pasting to achieve standard format Mutichannel electrocardiographs require less preparation time - 3 leads are recorded simultaneously In some mutichannel units, no cutting or pasting is necessary » ECG signals from each lead are stored in memory » Then traced all at once on a single page » Therefore, all 12 ECG tracings are representations of the heart's electrical activity from same heart beat Formatting options One-page formatting Formatting according to number of pages programmed Formatting according to number of heart beats recorded in each lead group Formatting according to number of seconds for overall recording Automatic formatting saves time but an atypical waveform may be missed Some units are capable of producing multiple copies of recent ECGs Data which can be printed at the top of ECG recordings Patient data (entered by alphanumeric keyboard if available) Time Lead identifiers Heart rate Recording parameters » Sensitivity setting » Chart speed » Filtered mode Options Heart rate LED or LCD displays Alarms » Tachycardia » Artifacts » Loose electrodes » Out-of-paper » System fault (leads/electrodes) » Automatic recording Capability of storing rhythm strips for later retrieval Ability to extend recording time if arrhythmia is detected ECG report editing capability "Freeze" capability » ECG will be held and displayed indefinitely » The current trace will be lost or transferred to another channel (printer) "Memory delay" feature - the last few seconds of the trace prior to an alarm will be lost or transferred to another channel or printer Some store, compute, and display trends in data through the use of a microprocessor » Trend plot is a graph of a physiological parameter over a period of time » Heart rate trend is an example Where Multichannel Electrocardiographs are Found Doctor's office Flight medicine clinics Physical exams and standards clinics Intensive care units Coronary care units Emergency rooms Cardiopulmonary labs Special care units OBJECTIVE Without reference, identify basic facts about the clinical applications of defibrillators with at least 70 percent accuracy. Terms and Definitions Arrest - cessation of the electrical activity of the heart Atrial flutter - very rapid (250-150/min) electrical discharge and contraction of the upper chambers (atria) of the heart Cardiogenic shock - shock resulting from the diminishment of cardiac output Fibrillation - irregular, totally disorganized electrical activity of the atria or ventricles or both Defibrillation - electrical termination of fibrillation Cardioversion - the restoration of the sinus rhythm by electrical shock Synchronous - occurring at the same time Stored energy - the energy stored within the defibrillator by the capacitor Joule - a unit of work; the energy expended by 1 amp flowing for 1 second through 1 ohm of resistance Also called a watt second The unit of calibration for the output of a defibrillator Intended Purpose To apply controlled monophasic (single phase) or biphasic (two phase) DC defibrillating pulse to the heart Monophasic One pulse of electricity One direction of current flow between the paddles High amounts of energy needed Biphasic • Current reverses itself • Two directions of current flow • Uses smaller amounts of energy • Can compensate for differences in patient impedance Chest size Tissue density Sync mode To perform elective cardioversion Uses the patient's generated R-wave as a timing reference Most defibrillators also monitor the ECG signal To verify fibrillation To verify effectiveness of treatment Patient signals can be picked up by the external paddles which are transcutaneous leads Paddle Types Standard adult external Anterior/posterior Internal Pediatric Neonatal (newborn) Disposable Modes of Operation External emergency defibrillation Discharge into the patient by pressing the 2 discharge buttons (one on each handle) simultaneously 2,000 to 4,000 volt shock For less than 20 msec Using gels and pastes to improve conductivity between paddles and chest AHA ( American Heart Association) Studies indicate that about 98% of patients can be defibrillated with 300 joules or less Recommends not exceeding 360 joules delivered to the patient Internal defibrillation Energy is delivered directly to the exposed heart All defibrillators are designed to limit the output energy to 50 joules to prevent injury to the heart muscle • Paddles are small (50 mm in dia.) and able to withstand sterilization between uses Synchronized cardio version (Sync Mode) • Uses a discharge of 25-100 joules • Used to correct certain arrhythmias Ventricular tachycardia Atrial flutter • Sync marker will show up on the defibrillator monitor on the R-wave The shock is delivered On the first down stroke of the R-wave detected after both paddle discharge buttons are pressed Must occur within 30 msec of the Rwave Timing is critical because discharge during the T-wave could cause fibrillation OBJECTIVE Without reference, identify at least four out of six basic facts about the clinical applications of invasive and noninvasive blood pressure monitors. Purpose The presence of an ECG signal does not assure effective pumping of blood Automatic electronic sphygmomanometers noninvasively measure and display a patient's arterial blood pressure Terms and Definitions Systolic pressure - the highest arterial pressure of the blood Diastolic Pressure - the lowest arterial pressure of the blood Mean - midway or an average of the systolic and diastolic pressures Pulse pressure - the difference between the systolic and diastolic pressures Korotkoff sounds - sounds heard through the stethoscope as the blood flow changes Hypotension - low blood pressure Hypertension - high blood pressure Hypovolemia - inadequate blood volume Hypervolemia - excessive blood volume Noninvasive - uses a cuff Invasive - uses a catheter and puts patient at a higher risk of infection or emboli Normal pressure - 120/80 mmHg expressed as systolic over diastolic pressures Principles of Blood Pressure Left ventricle of heart contracts • Blood is forced into arteries • Creates a pressure increase, peak of which is called systolic pressure Ventricles then relax Pressure in the arteries decrease as blood leaves arteries and enters capillaries Lowest point the pressure reaches before the next ventricular contraction represents the diastolic pressure Pressure values are recorded in millimeters of mercury (mmHg) Noninvasive Method Peripheral blood flow sounds were correlated to systolic and diastolic pressures by Nicolai Korotkoff in 1905 Uses a cuff wrapped around patient's arm and a stethoscope Cuff is inflated to pressure greater than systolic pressure Patient's artery closes Blood flow stops Cuff pressure is gradually lowered, pressure falls below systolic pressure but higher than diastolic pressure Some blood forces its way through artery Blood flow not normal, the resulting turbulence produce korotkoff sounds Sounds persist until cuff pressure falls below the diastolic pressure and blood flow returns to normal Techniques for Automatic Measurement Auscultatory Same principles as sphygmomanometers, detection of korotkoff sounds Pressures at which korotkoff sounds first begin mark the systolic pressure Pressures at which sounds disappear marks the diastolic pressure Cuff with transducer wrapped around patient's arm Cuff is inflated with transducer positioned against compressed artery Transducer detects korotkoff sounds Enables user to determine both systolic and diastolic values Oscillometric Cuff is wrapped around a patient's arm and inflated Pressure in the cuff is released Sensor located in the monitor detects air pressure fluctuations in the cuff Due to arterial volume changes Occur because blood is pulsing through artery, rather than flowing smoothly Pressure at which oscillations peak correspond to mean arterial pressure (MAP) Unit calculates the systolic and diastolic pressures from the increasing and decreasing magnitude of the oscillations Differential sensor Cuff with dual-head sensor inflates around a patient's arm Cuff pressure is released Sensor against artery detects korotkoff sounds, oscillometric pressures and artifact signals Sensor against the cuff (air bladder) detects only oscillometric sounds and artifact signals The monitor subtracts the two signals leaving only the korotkoff sounds Process should remove some of the unwanted interference signals Light-emitting diode (LED) Cuff bladder - slipped over patient's finger or thumb Cuff is inflated stepwise under microprocessor control Signals from optical sensor determine mean arterial pressure which is sufficient enough to permit the artery to remain open Procedure repeated several times in first minute until volume of finger under cuff is stabilized » Once stabilized, unit is calibrated to patient » Calibration procedure takes place every minute in some units Changes in blood volume causes the optical sensor to send signals to an electropnuematic servo-controlled valve Increases and decreases the cuff pressure A transducer detects the cuff pressure and produces a corresponding electrical signal Arterial pressure waveform can be displayed on a monitor or sent to a printer Advantages of Noninvasive Low risk to patient compared to invasive Simplicity - easy to set up Locations Emergency room Intensive care unit Surgery Recovery OBJECTIVE Without reference, identify basic facts about the clinical applications of infusion pumps with at least 70 percent accuracy. Introduction to Fluid Therapy Body fluid balance depends upon • Fluid intake • Body requirements for fluids Functioning regulatory systems • Cardiovascular system (volume and blood pressure) • Urinary system (Kidneys) • Nervous system • Endocrine system (Hormones) Electrolytes and other chemical substances which keep body fluids in balance Disturbances in fluid balance Dehydration - insufficient body water Causes » Inadequate fluid intake » Vomiting and diarrhea » Fever and excessive sweating Signs and symptoms Thirst Weight loss Low fluid output/concentrated urine Dry membranes in mouth and eyes Treatment Replace fluids orally when mild Replace fluids by standard IV drip when severe Edema - accumulation of excess fluid in body tissue Causes Heart failure or vessel obstruction Protein loss (starvation, burns) Systemic infection/inflammation Kidney problems/renal failure Signs and symptoms Swollen eyelids/puffy face Swollen feet and ankles Abdominal distention Treatment Special IV fluids which pull water from tissue into vessels Low sodium diet • Other disturbances in fluid balance Dysfunction in almost any organ system can affect fluid balance Fluid balance can be affected by certain drugs - hormones, diuretics Fluid Therapy By Intravenous Route IV regular and special use • Provides route for fluids when oral route cannot be used or is not practical GI problems (obstruction, malabsorption) Inability to swallow (oral surgery, unconscious patient) Intolerance to oral fluids (nausea, vomiting) Some medications are more effective when given by IV route (antibiotics) Used pre- and post-operatively GI tract must be at rest for anesthesia related safety purposes (i.e. no food) Gives emergency access to the circulatory system during surgery for fluid replacements, blood, medication, etc Emergency Effects of IV fluids/drugs are realized immediately (within seconds) Gives rapid access to circulatory system Lack of good peripheral circulation in a shock patient causes poor absorption of IM (intramuscular) or subcutaneous injection. Therefore, IV is the method of choice Limitation/problems of IV therapy • Requires special skill • Certain fluids/drugs cannot be given through an IV • Complications associated with venous IV site Infection Tissue breakdown due to drugs or reaction of tissue with IV line Blood clots (embolism) Air embolus (air bubbles in blood stream) Monitoring IV Therapy Drip rate - measured in cc/hr, is the correlation between number of drops/minute with actual volume (cc)/hour Leaves room for error due to multiplier effect Changes due to filter clogging or kinked tube Ratio of drops/minute to cc/hr differs with infusion sets and viscosity of liquid being infused Thicker fluids have slower drops/minute Importance of monitoring • Assure proper amount of drug/fluid is administered • Required to prevent too much or too little fluid intake (edema or dehydration) Infusion Therapy Best way to provide continuous, well controlled infusion • Controlled/preset volume • Delivery is at a constant rate • Can monitor rate, volume, or both depending on model or style Types Controllers Use gravity feed Monitor the rate by counting drops/minute Control rate by pinching infusion line Pumps - use pressure to move fluid from IV bag to the patient Peristaltic (usually not direct volumetric, but counts drops) Piston - known volume per stroke Therefore, rate of stroke controlled Uses - provides continuous, well controlled administration of critical drugs/fluids when continuous levels and/or well controlled volume is required • Children and infants who have a small body volume • Drug therapy Antibiotics (systemic) Cardiac drugs Anticoagulants Insulin Electrolytes • Hyperalimentation - providing the body with nutrition intravenously Advantages of infusion pumps over standard IV methods Usually more accurate/consistent than standard drip method Allows for better measurement of total fluid input when this is critical (edema) Warn the staff when there are problems like extremely high or low pressure in the IV line (occlusion or dislodged line) Disadvantages - always be aware • Requires power, therefore can stop running unexpectedly • If no battery backup, patient's movement is limited • Pumps can build up extremely high pressure in the IV line or patient • Can be inaccurate Too little infused - insufficient Too much infused - overdose, circulatory overload, edema, death Only as accurate as the BMET who calibrates it!