General detection limit in natural water (drinking water and pure lake water) is in the low ppb area. Some approximate guiding electrolabile values. Maximum concentrations without a recommended dilution are 500 ppb. Accuracy is 10 % of full range. However, given values might be matrix dependant. Metal to detect Mn(II) Working Electrode Solid silver alloy Zn(II) Solid silver alloy Cd(II) Solid silver alloy Pb(II) Solid silver alloy Tl(I) Solid silver alloy Bi(III) Solid silver alloy Cu(II) Solid gold alloy Ni(II) Solid silver alloy Co(II) Solid silver alloy Hg(II) Solid gold alloy Cr(VI) Solid gold alloy As(III) Solid gold alloy Fe(II)/Fe(III) Solid silver alloy DMG = Dimethylglyoxime. Supporting electrolyte (high purity chemicals are needed) Ammonium chloride buffer (pH 6.4) (or NH4Cl 0.02 M) Ammonium chloride buffer (pH 6.4) (or NH4Cl 0.02 M) Ammonium chloride buffer (pH 6.4) (or NH4Cl 0.02 M) Ammonium chloride buffer (pH 6.4) (or NH4Cl 0.02 M) Acetate buffer pH 6.4) Ep (V)1 -1.45 Detection limit (t = 360 s)2 < 5 ppb Ligand to add -1.05 < 5 ppb -0.65 < 5 ppb -0.50 < 2 ppb -0.75 < 5 ppb Ammonium chloride buffer (pH 3) (or NH4Cl 0.02 M) HCl (0.01 M) -0.35 < 10 ppb +0.25 < 1 ppb Ammonium chloride buffer (pH 9.6) Ammonium chloride buffer (pH 9.6) HNO3 (10 mM) and HCl (10 mM) -0.95 < 2 ppb3 DMG -1.15 < 2 ppb3 DMG4 +0.55 < 1 ppb HCl (0.1 M) +0.25 < 5 ppb HCl (0.2 M) +0.15 < 1 ppb Tri-sodium citrate 5.5 hydrate or NH4Cl (0.02 M) -0.80 < 2 ppb 1) Peak potentials (Ep) might differ in accordance to used reference electrode (RE), (table values against newly made Ag-AgCl/KCl 3 M RE). (Counter electrode is Platinum). 2) For increased sensitivity the deposition time can be extended, but deposition time longer than 1800 seconds is not recommended. Up to ten times detection limit improvement is expected by using square wave voltammetry, to be included in the new version ATMS600 soon to be introduced. 3) By adsorptive differential pulse cathodic stripping voltammetry (deposition potential at – 0.75 V for 60 s). 4) If Ni(II) and Co(II) is present simultaneously in the sample; use a mixture of DMG (0.3 mM), oxine (0.02 mM) and nioxime (0.01 mM). (Maximum number of metals detected simultaneously is five metal, including Mn(II), Zn(II), Cd(II), Pb(II) and Cu(II) (on Solid silver alloy Working Electrode)).