Uploaded by xin ouyang

Breas PV101+, 102+ Ventilator - Service manual

advertisement
PV 101+/PV 102+ Service Manual
TABLE OF CONTENTS
TABLE OF CONTENTS
1
INTRODUCTION............................................................................. 1–1
1.1
About the PV 101+ and PV 102+ ............................................................1–1
1.1.1
1.1.2
1.1.3
1.1.4
1.2
About this manual ...................................................................................1–2
1.2.1
1.2.2
1.2.3
2
Intended use of the PV 101+ ................................................................ 1–1
Intended use of the PV 102+ ................................................................ 1–1
Design and function .............................................................................. 1–1
Service personnel’s training requirements ............................................ 1–1
Scope .................................................................................................... 1–2
Intended audience................................................................................. 1–2
Icons...................................................................................................... 1–2
MAINTENANCE SERVICE INSTRUCTIONS ................................. 2–1
2.1
Purpose of the maintenance ...................................................................2–1
2.2
Introduction .............................................................................................2–2
2.2.1
2.2.2
2.2.3
2.2.4
2.2.5
2.3
Preparing inspection................................................................................2–4
2.3.1
2.3.2
2.3.3
2.3.4
2.3.5
2.4
Verifying the components and the installed software............................ 2–4
Initial recording...................................................................................... 2–4
Markings................................................................................................ 2–4
Information from the patient/user .......................................................... 2–4
Validity of the documentation ................................................................ 2–4
External inspection ..................................................................................2–5
2.4.1
2.4.2
2.4.3
2.5
Service check schedule ........................................................................ 2–2
Safety precautions ................................................................................ 2–2
Service record ....................................................................................... 2–2
Inspection equipment and tools ............................................................ 2–3
Replacement parts ................................................................................ 2–3
Visual inspection for external damage and wear .................................. 2–5
Checking the power supply connection................................................. 2–5
Short function check.............................................................................. 2–5
Internal inspection ...................................................................................2–6
2.5.1
2.5.2
2.5.3
2.5.4
2.5.5
2.5.6
2.5.7
2.5.8
2.5.9
2.5.10
Doc. No. 003259 En
Opening the casing ............................................................................... 2–6
Cleaning the inside of the machine ....................................................... 2–6
Inspecting the internal wiring................................................................. 2–6
Securing the components ..................................................................... 2–6
Checking the power supply ................................................................... 2–6
Checking the electrical safety values .................................................... 2–6
Calibrating the pressure sensor ............................................................ 2–6
Calibrating the flow sensor.................................................................... 2–6
Calibrating the high pressure limit......................................................... 2–6
Reassembling the casing ...................................................................... 2–6
Issue: X-1
BREAS MEDICAL
TOC – 1
TABLE OF CONTENTS
2.6
Final inspection before delivery .............................................................. 2–7
2.6.1
2.6.2
2.6.3
2.6.4
2.6.5
2.6.6
2.6.7
3
4
5
Function check ...................................................................................... 2–7
Checking the low pressure leakage alarm (PV 102+)........................... 2–7
Checking the trigger function ................................................................ 2–7
Checking the pressure and rate ............................................................ 2–7
Replacing the filters............................................................................... 2–8
Checking the accessories ..................................................................... 2–8
Adjusting the settings for the patient ..................................................... 2–8
REMOVING/REPLACING THE MAIN COMPONENTS ................. 3–1
3.1
Tools ....................................................................................................... 3–1
3.2
Removing and reassembling the upper casing ....................................... 3–2
3.3
Removing and replacing the CPU board ................................................ 3–3
3.4
Removing and reassembling the main assembly ................................... 3–4
3.5
Removing and replacing the blower unit................................................. 3–6
3.6
Removing and replacing the power board .............................................. 3–7
3.7
Removing and replacing the MDA board ................................................ 3–8
CALIBRATION INSTRUCTIONS.................................................... 4–1
4.1
Requirements.......................................................................................... 4–1
4.2
Equipment............................................................................................... 4–1
4.3
Calibrating the pressure sensor .............................................................. 4–2
4.4
Verifying the pressure calibration............................................................ 4–3
4.5
Calibrating the flow sensor...................................................................... 4–4
4.6
Calibrating the high pressure limit........................................................... 4–5
PARTS DRAWINGS AND PART-NUMBER LISTS ....................... 5–1
5.1
Overview – Main components................................................................. 5–2
5.2
Parts drawings ........................................................................................ 5–4
5.2.1
5.2.2
5.2.3
5.3
Parts drawing – Upper and lower casing .............................................. 5–5
Parts drawing – Main assembly ............................................................ 5–6
Parts drawing – Blower unit and patient air outlet................................. 5–7
Part-number lists..................................................................................... 5–8
5.3.1
5.3.2
5.3.3
5.3.4
6
PV 101+/PV 102+ Service Manual
Part-number list – Upper and lower casing ........................................... 5–8
Part-number list – Main assembly......................................................... 5–9
Part-number list – Blower unit and patient outlet .................................. 5–10
Part-number list – Miscellaneous .......................................................... 5–11
FUNCTIONAL DIAGRAMS ............................................................ 6–1
6.1
Pneumatic diagram ................................................................................. 6–1
6.2
Functional block diagram ........................................................................ 6–2
TOC – 2
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
7
ELECTRONICS............................................................................... 7–1
7.1
Electronics – Main components ..............................................................7–2
7.2
Functional block diagram ........................................................................7–4
7.3
Internal wiring diagram ............................................................................7–5
7.4
Circuit board descriptions........................................................................7–6
7.4.1
7.4.2
7.4.3
Electrical safety measurements ..............................................................7–8
7.6
Circuit diagrams – Circuit boards ............................................................7–9
7.7
7.8
Circuit diagram – CPU board 1, rev 2 ................................................... 7–10
Circuit diagram – CPU board 2, rev 2 ................................................... 7–11
Circuit diagram – CPU board 3, rev 2 ................................................... 7–12
Circuit diagram – CPU board 4, rev 2 ................................................... 7–13
Circuit diagram – CPU board 5, rev 2 ................................................... 7–14
Circuit diagram – Power board 1, rev 3................................................. 7–15
Circuit diagram – Power board 2, rev 3................................................. 7–16
Circuit diagram – MDA board 1, rev 3................................................... 7–17
Circuit diagram – MDA board 2, rev 3................................................... 7–18
Circuit diagram – MDA board 3, rev 3................................................... 7–19
Component location drawings – Circuit boards.......................................7–20
7.7.1
7.7.2
7.7.3
7.7.4
Component location drawing – CPU board........................................... 7–21
Component location drawing – Power board, upper side ..................... 7–22
Component location drawing – Power board, under side ..................... 7–23
Component location drawing – MDA board .......................................... 7–24
Component lists – Circuit boards ............................................................7–25
7.8.1
7.8.2
7.8.3
Component list – CPU board ................................................................ 7–25
Component list – Power board.............................................................. 7–26
Component list – MDA board ................................................................ 7–29
FAULT TRACING ........................................................................... 8–1
8.1
9
CPU board ............................................................................................ 7–6
Power board.......................................................................................... 7–6
MDA board ............................................................................................ 7–7
7.5
7.6.1
7.6.2
7.6.3
7.6.4
7.6.5
7.6.6
7.6.7
7.6.8
7.6.9
7.6.10
8
TABLE OF CONTENTS
Fault-tracing table....................................................................................8–1
APPENDICES ................................................................................. 9–1
9.1
Engineering change history – PV 101+...................................................9–1
9.2
Engineering change history – PV 102+...................................................9–1
Service Record for BREAS PV 101+/PV 102+
9.3
Returning products to BREAS.................................................................9–5
Service Report for BREAS PV 101+/PV 102+
Doc. No. 003259 En
Issue: X-1
BREAS MEDICAL
TOC – 3
TABLE OF CONTENTS
PV 101+/PV 102+ Service Manual
10 PV 101+ OPERATING MANUAL
11 PV 102+ OPERATING MANUAL
TOC – 4
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
1
INTRODUCTION
INTRODUCTION
This chapter gives an overview of the PV 101+ and PV 102+ devices and the service
manual.
WARNING!
This product must be:
• subjected to regular service, maintenance and control and any applicable upgrades,
in accordance with BREAS service instructions;
• repaired and/or modified in accordance with BREAS service manuals, technical bulletins and any special service instructions, by service technicians that have been
authorised after BREAS PV 101+/PV 102+ service training, or have an equivalent
technical knowledge on medical respiratory devices.
Deviation from these service instructions may lead to risk of personal injury!
1.1
About the PV 101+ and PV 102+
The PV 101+ and PV 102+ are intended for use in homes, while travelling or in institutions. They are bi-level devices, which means that the expiration pressure (EPAP) can
be set lower than the inspiration pressure (IPAP).
1.1.1
Intended use of the PV 101+
The BREAS PV 101+ device is intended for treatment of obstructive sleep apnoea.
1.1.2
Intended use of the PV 102+
The BREAS PV 102+ device is intended for treatment of obstructive and central sleep
apnoea, hypopnea and other forms of sleep-related breathing problems. The back-up
rate is adjustable, as well as the rise time, which makes it possible to customise the time
for the pressure to change from EPAP to IPAP.
1.1.3
Design and function
The PV 101+ and PV 102+ are based on a blower which is driven by an electronically
controlled servomotor. The microprocessor-based electronics calculates, from the set
pressure, rate, rise time etc., the correct motor speed. The set pressure and trigger sensitivity is continuously monitored.
An audible alarm is immediately given in the event of a power failure.
1.1.4
Service personnel’s training requirements
Service personnel working with the PV 101+ and PV 102+ devices should have medical/technical training and a good knowledge of the construction and function of respiratory devices. Authorisation by BREAS PV 101+/PV 102+ service training is
recommended.
Always contact your BREAS representative if you have any questions or if
training is required.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
1–1
INTRODUCTION
PV 101+/PV 102+ Service Manual
1.2
About this manual
1.2.1
Scope
All the maintenance checkpoints and the additional service actions for the PV 101+/
PV 102+ are described in this manual. It contains all documentation required for maintaining and servicing the machines, such as enlarged drawings, wiring diagrams, component location guides and so on.
BREAS Medical reserves the right to make changes to the products and/or the contents
of this manual without any prior notice.
1.2.2
Intended audience
This service manual is intended for service technicians who have medical/technical
training and who have a good knowledge of the construction and function of respiratory
devices.
The service manual is NOT intended for clinic personnel or patients, who can find
all the information they need in the PV 101+ and PV 102+ Operating Manuals.
1.2.3
Icons
Icons are used in this manual to highlight specific types of information. The meaning of
each icon is explained in the table below.
Icon
Explanation
Warning!
Risk of death and serious personal injury.
Caution!
• Risk of minor or moderate injury.
• Risk of equipment damage, loss of data, extra work or unexpected results.
Note
Information that may be valuable but is not of critical importance, tips.
Reference
Reference to other manuals with additional information on a specific topic.
1–2
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
2
MAINTENANCE SERVICE INSTRUCTIONS
MAINTENANCE SERVICE INSTRUCTIONS
This chapter describes all the routine checks and additional service instructions for the
PV 101+ and PV 102+.
WARNING!
This product must be:
• subjected to regular service, maintenance and control and any applicable upgrades,
in accordance with BREAS service instructions;
• repaired and/or modified in accordance with BREAS service manuals, technical bulletins and any special service instructions, by service technicians that have been
authorised after BREAS PV 101+/PV 102+ service training, or have an equivalent
technical knowledge on medical respiratory devices.
Deviation from these service instructions may lead to risk of personal injury!
For information about fault tracing, parts drawings, board schematics and so on, refer to
the respective chapter in this service manual.
The patient and/or care provider should follow the function checks described in the
chapter “CHECKS BEFORE USE” in the PV 101+/PV 102+ Operating Manual.
2.1
Purpose of the maintenance
The PV 101+ and PV 102+ are designed to give the users many years of trouble-free
assistance, provided that preventive maintenance is carried out at the intervals specified
in this manual. Well-performed maintenance will considerably increase the life of the
machine.
It is also important to check any peripheral equipment at the same time as services are
carried out.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
2–1
MAINTENANCE SERVICE INSTRUCTIONS
2.2
PV 101+/PV 102+ Service Manual
Introduction
Before you start the maintenance service, read the safety precautions and make sure
you have a new service record and all the necessary equipment, tools and replacement
parts at hand.
2.2.1
Service check schedule
The schedule below lists the recurrent service checks of the PV 101+/PV 102+’s components and the intervals at which they should be performed. The reference numbers
refer to chapters and sections in this manual that contain more detailed instructions.
A complete maintenance service, as described in this chapter, must be carried out at least every 12 months.
Interval
Service check
See
section
Every 12 months
Complete maintenance check
2
Check the electrical safety limit values.
7.5
Calibrate the pressure, flow and high pressure limit
sensors.
4
Replace the blower unit.
3.5
Calibrate the pressure, flow and high pressure limit
sensors.
4
Every 15,000 operating
hours or when required
2.2.2
Safety precautions
Follow the safety precautions below when working with the PV 101+/PV 102+:
•
Do not work on the device with the casing removed and any power supply connected unless the instructions in this manual clearly say so.
•
Always use caution when working with the device connected to the mains and the
casing removed.
•
Make sure that all precautions to prevent electrostatic discharge (ESD) have been
taken. Follow all the regulations regarding ESD.
•
Explosive gases and liquids must not be kept or used near the PV 101+/PV 102+.
The PV 101+/PV 102+ Operating Manual contains an extended list of safety precautions.
2.2.3
Service record
See section 9, “APPENDICES” for the BREAS service record.
•
2–2
Copy the service record and use it for noting the service checks while performing
the yearly service.
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
2.2.4
MAINTENANCE SERVICE INSTRUCTIONS
Inspection equipment and tools
Before servicing the PV 101+/PV 102+, make sure you have the following equipment at
hand:
•
Test lung
•
Multimeter
•
Measuring equipment for pressure 0–70 mbar
(Thommen HM 28 digital manometer is recommended, BREAS part no. 001934)
•
Flow meter
•
Screwdrivers TX10, TX25, Phillips 1
•
Box socket wrenches M3, M4
•
Allen key 2, 5x25
For the equipment required when calibrating the device, see section 4.2.
2.2.5
Replacement parts
The following parts should be available when servicing the device:
Description
BREAS part no.
Patient tube
000245
Patient air filter, washable (2 pcs)
000996
Patient air filter, disposable (if used)
001259
Air filter, internal cooling
001121
Blower unit assembly, 15.000 hours
003110
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
2–3
MAINTENANCE SERVICE INSTRUCTIONS
PV 101+/PV 102+ Service Manual
2.3
Preparing inspection
2.3.1
Verifying the components and the installed software
•
Check section 9.1, “Engineering change history – PV 101+” or section 9.2, “Engineering change history – PV 102+” to find out what changes have been made to
the device and from which serial number they were implemented.
If in any doubt about the circuit board revisions or upgrades that may have been made
that have not been recorded, check the component designations on the circuit boards.
2.3.2
2.3.3
2.3.4
2.3.5
Initial recording
1
Copy a new service record (see section 9, “APPENDICES”).
2
Identify the PV 101+/PV 102+; note the model and serial number and any inventory number, if applicable, on the service record.
3
Check earlier actions and events recorded on any previous service records.
4
Note the current patient settings.
5
Note the total number of operating hours.
Markings
Check that the following markings are legible:
•
Make, model designation, serial number
•
Warning texts
•
Any inventory markings
•
Any other texts on the external equipment and accessories
Information from the patient/user
Before starting the service, check the following with the patient/clinic:
•
Has the PV 101+/PV 102+ functioned without any problems? If not, what were
they?
•
How does the patient/care provider check the function of the device? How often?
•
How often are the filters changed?
•
How many filters will be required before the next service?
•
Other observations?
Validity of the documentation
1
Check that the documentation used by the patient is valid and up to date.
2
2–4
Check if any modification or revision of the PV 101+/PV 102+ needs to be done at
the same time as the service.
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
MAINTENANCE SERVICE INSTRUCTIONS
2.4
External inspection
2.4.1
Visual inspection for external damage and wear
1
Clean the cover of the PV 101+/PV 102+ using a mild detergent.
2.4.2
2.4.3
2
Check for visible damage to the cover and other components.
3
Check that all external parts are tightened and secured.
Checking the power supply connection
1
Check the plug, power cord and mains power socket.
2
Make sure the strain-relief clamp for the power cord is not damaged.
3
Check the power supply to the machine.
4
If the PV 101+ does not operate from the mains power, check the fuses.
Short function check
1
Connect the power cord to the power socket of the PV 101+/PV 102+.
2
Connect the patient circuit.
3
Start the PV 101+/PV 102+ and check that it functions normally.
4
Disconnect the power cord and check that the power failure alarm sounds. Reconnect the power cord. The alarm should be silenced.
5
Switch on the PV 101+/PV 102+. Upon starting, all the LEDs on the front panel
should be lit and the trigger settings should be indicated on the bar graph. The
entire bar graph should then become yellow for 2 seconds before the device starts
functioning normally.
6
(PV 102+ only) Switch on the PV 102+ and wait for 15 seconds for the leakage
alarm to be activated.
7
Switch off the device.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
2–5
MAINTENANCE SERVICE INSTRUCTIONS
2.5
PV 101+/PV 102+ Service Manual
Internal inspection
Disconnect the power supply before removing the casing of the machine.
2.5.1
Opening the casing
•
For instructions, see section 3.2, “Removing and reassembling the upper casing”.
2.5.2
Cleaning the inside of the machine
•
Remove any dust or dirt that has collected in the machine.
2.5.3
Inspecting the internal wiring
•
Inspect all wiring and the connectors for damage. Make sure that no wires or
cables are pinched.
2.5.4
Securing the components
•
Make sure that all the internal components, such as the motor unit, the circuit
boards, the connectors and so on, are properly secured.
2.5.5
Checking the power supply
1
Check that the power inlet socket is undamaged and that it is properly fastened.
2
Make sure the power board is properly fastened.
3
Check the wiring to and from the power board.
2.5.6
Checking the electrical safety values
•
For instructions, see section 7.5, “Electrical safety measurements”.
2.5.7
Calibrating the pressure sensor
•
For instructions, see section 4.3, “Calibrating the pressure sensor”.
2.5.8
Calibrating the flow sensor
•
For instructions, see section 4.5, “Calibrating the flow sensor”.
2.5.9
Calibrating the high pressure limit
•
For instructions, see section 4.6, “Calibrating the high pressure limit”.
2.5.10 Reassembling the casing
•
For instructions, see section 3.2, “Removing and reassembling the upper casing”.
2–6
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
MAINTENANCE SERVICE INSTRUCTIONS
2.6
Final inspection before delivery
2.6.1
Function check
1
Connect the power cord to the power socket of the PV 101+/PV 102+.
2.6.2
2
Connect the patient circuit.
3
Start the PV 101+/PV 102+ and check that it functions normally.
4
Disconnect the power cord and check that the power failure alarm sounds. Reconnect the power cord. The alarm should be silenced.
5
Switch on the PV 101+/PV 102+. Upon starting, all the LEDs on the front panel
should be lit and the trigger settings should be indicated on the bar graph. The
entire bar graph should then become yellow for 2 seconds before the device starts
functioning normally.
6
(PV 102+ only) Switch on the PV 102+ and wait for 15 seconds for the leakage
alarm to be activated.
7
Switch off the device.
Checking the low pressure leakage alarm (PV 102+)
1
Adjust the pressure settings to:
2
2.6.3
Checking the trigger function
1
Adjust the pressure settings to:
2
3
2.6.4
IPAP
10 mbar
EPAP
10 mbar
Make sure that the alarm LED is lit and that an audible alarm is heard.
IPAP
10 mbar
EPAP
4 mbar
Connect the patient tube without a mask.
Block the air outlet on the tube. When you remove the blockage, a triggered
breath should be initiated.
Checking the pressure and rate
Pressure
The measurement should be made using a blocked patient circuit with leakage provided
(see figure below).
1
2
Adjust the pressure settings to:
IPAP
10 mbar
EPAP
10 mbar
Measure the pressure and check that it is correct (tolerance: ±4% of the maximum
value and ±8% of the current pressure value).
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
2–7
MAINTENANCE SERVICE INSTRUCTIONS
PV 101+/PV 102+ Service Manual
Pressure gauge
mbar
5 mm leakage
Patient circuit
Plug
Rate
3
Adjust the settings to:
IPAP
10 mbar
EPAP
4 mbar
Back-up rate 8 BPM (PV 102+ only)
The PV 101+ back-up rate is fixed to 6 BPM.
4
2.6.5
2.6.6
Replacing the filters
The patient air filters are located in the filter holder under the carrying handle of the
device. The internal cooling filter is located in the lower casing.
1
Replace the washable grey air filter.
2
Replace the disposable white filter, if used.
3
Replace the internal cooling filter from underneath the machine.
Checking the accessories
1
Inspect the patient circuit. Replace it if necessary.
2
2.6.7
2–8
Measure the rate and check that it is correct.
Inspect any other accessories.
Adjusting the settings for the patient
•
Adjust all the patient settings to the prescribed values.
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
3
REMOVING/REPLACING THE MAIN COMPONENTS
REMOVING/REPLACING THE MAIN COMPONENTS
This chapter describes how to install, remove and/or replace the main components of
the PV 101+/PV 102+.
3.1
Tools
Before disassembling the PV 101+/PV 102+, make sure you have the following tools at
hand:
•
Screwdriver TX10
•
Screwdriver TX25
•
Screwdriver Phillips 1
•
Box socket wrench M4
•
Box socket wrench M3
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
3–1
REMOVING/REPLACING THE MAIN COMPONENTS
3.2
PV 101+/PV 102+ Service Manual
Removing and reassembling the upper casing
Disconnect the power supply before removing the casing of the machine.
1
2
To remove the casing:
Remove the mains power cord.
Turn the PV 101+/PV 102+ upside down. Remove the four screws placed in the
four holes underneath the lower casing and the two screws placed underneath the
handle.
6 × TX25
3
Lift up the upper casing. Disconnect the CPU board’s ribbon cable from the MDA
board. Put the upper casing aside.
To reassemble the casing:
•
Reassemble in reverse order.
3–2
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
3.3
REMOVING/REPLACING THE MAIN COMPONENTS
Removing and replacing the CPU board
To remove the CPU board:
1
Remove the upper casing (See section 3.2).
2
Turn it upside down and remove the three Torx screws placed underneath the
CPU board.
3
Remove the CPU board.
3 × TX10
To reassemble the CPU board:
•
Reassemble in reverse order. Fit the components to the holes for the display
panel controls and screw the CPU board back in place.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
3–3
REMOVING/REPLACING THE MAIN COMPONENTS
3.4
PV 101+/PV 102+ Service Manual
Removing and reassembling the main assembly
All the main internal components – blower unit, power board and MDA board – are
mounted on the base plate.
To remove the main assembly:
1
Remove the upper casing (See section 3.2).
2
Turn the PV 101+/PV 102+ upside down. Remove the M4 nut that fastens the
blower unit rubber mount.
3
Remove the five screws on the lower casing that hold the base plate.
2
1 × M4
Compound
nut
3
5 × TX10
3–4
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
4
REMOVING/REPLACING THE MAIN COMPONENTS
Press the rear panel side of the lower casing outwards to release the connectors
and lift out the main assembly.
To reassemble the main assembly:
•
Reassemble in reverse order. Make sure the connectors on the rear panel are
properly fitted in their holes.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
3–5
REMOVING/REPLACING THE MAIN COMPONENTS
3.5
PV 101+/PV 102+ Service Manual
Removing and replacing the blower unit
To remove the blower unit:
1
Remove the upper casing (See section 3.2).
2
Remove the main assembly (See section 3.4).
3
Disconnect the motor connectors from the MDA board.
4
Disconnect the flex tube from the blower unit’s air outlet. Disconnect the red sensor tube from the blower unit’s sensor connector.
5
Remove the remaining two nuts that hold the blower unit’s rubber mounts from
underneath the base plate.
6
Lift away the complete blower unit.
3
3
5
4
2 × M4 Compound nut
5
To reassemble the blower unit:
•
3–6
Reassemble in reverse order.
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
3.6
REMOVING/REPLACING THE MAIN COMPONENTS
Removing and replacing the power board
To remove the power board:
1
Remove the upper casing (See section 3.2).
2
Remove the main assembly (See section 3.4).
3
Remove the blower assembly (See section 3.5).
4
Move aside the patient air and sensor tubing. Disconnect the power cables to the
external DC and mains supplies on the MDA board.
5
Disconnect the DC control cable from the PSU board.
6
Unscrew the two screws that hold the rear part of the power board, and remove
the two nuts under the base plate.
7
Pinch the two securing pins on the front part of the power board, e.g. by using a
screwdriver, and lift up the power board from the base plate.
2 × Phillips 1
5
4
6
7
To reassemble the power board:
•
Reassemble in reverse order.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
3–7
REMOVING/REPLACING THE MAIN COMPONENTS
3.7
PV 101+/PV 102+ Service Manual
Removing and replacing the MDA board
To remove the MDA board:
1
Remove the upper casing (See section 3.2).
2
Remove the main assembly (See section 3.4).
3
Remove the blower assembly (See section 3.5).
4
Remove the power board (See section 3.6).
5
Disconnect the patient air and sensor tubing from the sensor connectors on the
MDA board.
6
Disconnect the external DC cable, the fan cable and the DC control cable from the
MDA board.
7
Turn the base plate upside down and remove the 3 screws that fix the MDA board
to the bottom of the plate.
8
Remove the 4 screws that fix the MDA board to the rear part of the base plate.
9
Lift up the MDA board from the base plate.
4 × T10
8
4 × M3 Compound
nut
6
5
7
3 × T10
3 × M3 Compound nut
To reassemble the MDA board:
•
3–8
Reassemble in reverse order.
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
4
CALIBRATION INSTRUCTIONS
CALIBRATION INSTRUCTIONS
This chapter describes how to adjust and verify the pressure, flow and high pressure
limit sensors of the PV 101+/PV 102+ device.
4.1
Requirements
•
The calibration should be made every 12 months or after 15,000 operating hours.
•
Keep the power cord of the PV 101+/PV 102+ connected to the mains power during the calibration.
Always use caution when working with the device connected to the mains
and the casing removed.
Always check and adjust the patient settings to the prescribed values after
you have calibrated the PV 101+/PV 102+.
4.2
Equipment
Make sure you have the following equipment to hand:
•
Screwdriver for the calibration screws (∅ 1 mm)
Pressure calibration
•
Flow restrictor nipple 6 mm, BREAS part number 000997
•
Flow restrictor nipple 5 mm, BREAS part number 003311
•
Voltmeter
•
Syringe with a y-piece tube
•
Pressure gauge, 0–70 mbar
Flow calibration
•
Flow restrictor 6 mm, BREAS part number 000997
•
Voltmeter
High pressure limit calibration
•
Flow restrictor nipple 4 mm, BREAS part number 003312
•
Fan control unit, BREAS part number 003302
•
Pressure gauge, 0–70 mbar
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
4–1
CALIBRATION INSTRUCTIONS
4.3
PV 101+/PV 102+ Service Manual
Calibrating the pressure sensor
1
Start the PV 101+/PV 102+ device by pressing the I/O button and the + button at
the same time.
2
Remove the casing (see section 3.2). Keep the CPU board connected to the MDA
board.
3
Use the voltmeter to measure the voltage on pin 9 on the analogue/digital output
connector on the rear panel of the PV 101+/PV 102+.
4
Adjust the screw on the P203 (Offset) potentiometer on the MDA board until the
voltage is 1V.
5
Connect the pressure gauge to the syringe and the y-piece tube. Remove the
black tube from the pressure sensor on the MDA board. Connect the free end of
the y-piece tube to the pressure sensor.
6
Use the syringe to create a pressure of 30 mbar ±0.5 mbar.
7
Adjust the screw on the P204 (Gain) potentiometer on the MDA board until the
voltage is 4 V.
8
Remove the y-piece tube from the MDA board’s pressure sensor and reconnect
the black tube.
9
Repeat step 4 to 8 until no further adjustment is necessary.
10
Remove the equipment. Continue with the pressure verification procedure (see
section 4.4).
P203
P204
Pressure
sensor
MDA board
P203 Offset
BREAS MEDICAL
P204 Gain
_
4–2
_
+
+
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
4.4
CALIBRATION INSTRUCTIONS
Verifying the pressure calibration
1
Start the PV 101+/PV 102+.
2
Remove the casing (see section 3.2).
3
Connect a pressure gauge to the patient air inlet and arrange a leakage with a
5 mm flow restrictor (see diagram below).
4
Set both the IPAP and EPAP pressure to 20 mbar.
5
Check that the pressure rises to 20 mbar ±1 mbar. If not, the P204 potentiometer
must be adjusted.
6
Set both the IPAP and EPAP pressure to 4 mbar.
7
Check that the pressure rises to 4 mbar ±0.5 mbar. If not, the P203 potentiometer
must be re-adjusted.
8
If necessary, repeat the pressure calibration procedure (see section 4.3).
9
Remove the equipment and reassemble the casing (see section 3.2).
Pressure gauge
mbar
5 mm leakage
Patient circuit
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
Plug
4–3
CALIBRATION INSTRUCTIONS
4.5
PV 101+/PV 102+ Service Manual
Calibrating the flow sensor
Always calibrate the pressure sensor before you calibrate the flow sensor.
Calibrating the flow sensor first may lead to unreliable results.
1
Start the PV 101+/PV 102+ device by pressing the I/O button.
2
Plug the patient air outlet with a
6 mm flow restrictor nipple.
3
Set both the IPAP and EPAP
pressure to 12 mbar.
4
Remove the casing (see section 3.2). Keep the CPU board
connected to the MDA board.
5
Use the voltmeter to measure the voltage on pin 8 on the analogue/digital output
connector on the rear panel of the PV 101+/PV 102+.
6
Adjust the screw on the P201 (Gain) potentiometer on the MDA board until the
voltage is 2.15 V (±0.15 V).
7
Block the air outlet to completely stop the air flow.
8
Adjust the screw on the P202 (Offset) potentiometer on the MDA board until the
voltage is 0.5 V (±0.05 V).
9
Remove the air flow blockage.
10
Repeat step 6 to 9 until no further adjustment is necessary.
11
Remove the equipment and reassemble the casing.
P201
P202
MDA board
P201 Gain
BREAS MEDICAL
P202 Offset
_
4–4
+
+
_
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
4.6
CALIBRATION INSTRUCTIONS
Calibrating the high pressure limit
1
Start the PV 101+/PV 102+ by pressing the I/O button.
2
Connect a pressure gauge to the patient air inlet and arrange a leakage with a
4 mm flow restrictor nipple (see figure).
Pressure gauge
mbar
4 mm leakage
Patient circuit
Plug
3
Remove the casing (see section 3.2). Disconnect the CPU board from the MDA
board.
4
Connect the manual fan control unit to the X202 connector on the MDA board.
5
Use the fan control unit to start the PV 101+/PV 102+ blower at full speed.
6
Adjust the screw on the P205 potentiometer on the MDA board until the pressure
measured on the pressure gauge is 40 mbar (±1 mbar).
7
Remove the equipment and reassemble the casing.
P205
X202
MDA board
+
Doc. No. 003259 En Issue: X-1
_
P205
BREAS MEDICAL
4–5
CALIBRATION INSTRUCTIONS
4–6
PV 101+/PV 102+ Service Manual
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
5
PARTS DRAWINGS AND PART-NUMBER LISTS
PARTS DRAWINGS AND PART-NUMBER LISTS
This chapter contains enlarged drawings and part-number lists of the PV 101+/PV 102+.
The main components of the device are displayed at the beginning of the chapter.
Some of the listed parts can be ordered as replacement parts from BREAS.
Please contact your BREAS supplier for more information.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
5–1
PARTS DRAWINGS AND PART-NUMBER LISTS
5.1
PV 101+/PV 102+ Service Manual
Overview – Main components
2
1
3
4
5
6
7
8
5–2
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
PARTS DRAWINGS AND PART-NUMBER LISTS
No
Description
1
Upper casing
2
Push-button overlay
On the upper casing
3
CPU board
In the upper casing
4
Blower unit, incl. motor
Mounted on base plate
5
Power board
Mounted to bottom of base plate
6
MDA board
Mounted to rear of base plate
7
Base plate
In the lower casing
8
Lower casing
Doc. No. 003259 En Issue: X-1
Location
BREAS MEDICAL
5–3
PARTS DRAWINGS AND PART-NUMBER LISTS
5.2
5–4
PV 101+/PV 102+ Service Manual
Parts drawings
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
5.2.1
PARTS DRAWINGS AND PART-NUMBER LISTS
Parts drawing – Upper and lower casing
For definitions of the part numbers, refer to the list in section 5.3.1.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
5–5
PARTS DRAWINGS AND PART-NUMBER LISTS
5.2.2
PV 101+/PV 102+ Service Manual
Parts drawing – Main assembly
For definitions of the part numbers, refer to the list in section 5.3.2.
5–6
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
5.2.3
PARTS DRAWINGS AND PART-NUMBER LISTS
Parts drawing – Blower unit and patient air outlet
For definitions of the part numbers, refer to the list in section 5.3.3.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
5–7
PARTS DRAWINGS AND PART-NUMBER LISTS
5.3
Part-number lists
5.3.1
Part-number list – Upper and lower casing
5–8
PV 101+/PV 102+ Service Manual
BREAS
Part no.
Definition
000041
Rubber mount
000488
Absorbent, dwg 10646-1 (T2), outer edge upper casing push-buttons
000489
Absorbent, dwg 10646-2 (T2), outer edge inlet bend
000490
Absorbent, dwg 10646-3 (T2), outer edge inlet
000492
Absorbent, dwg 10646-5 (T2), outer edge outer wall
000494
Absorbent, dwg 10646-7 (T2), outer edge upper casing inlet
000497
Absorbent, dwg 10647-3 (T2) outer edge lower casing outer wall
000499
Absorbent, dwg 10647-5 (T2) outer edge lower casing outer wall
000501
Absorbent, dwg 10647-7 (T2) outer edge lower casing outer wall
000674
M4, wing nut
000678
M4, lock nut
000706
Power cord strain relief
000718
Screw, MRX steel M5 x 12 fzb POZ
000996
Filter PV 101+/PV 102+, washable (2 pcs)
001095
Outer casing, upper
001112
Button for CPU board
001121
Filter 45 x 45 x 10 for cooling fan
001245
Screw, RX-PT 4-20*6 fzb
001259
Filter, disposable (5 pcs)
002490
Screw Mrt M4x14 fzb
002931
Overlay, back panel PV 101+/PV 102+
002960
CPU board PV 101+/PV 102+
002962
Lower casing PV 101+/PV 102+
002963
Screw Rts St 3.5 X 6.5, PV 101+/PV 102+
003198
Overlay, PV 102+ setting panel, Swedish
003199
Overlay, PV 102+ setting panel, English
003200
Overlay, PV 102+ setting panel, German
003201
Overlay, PV 102+ setting panel, Finnish
003202
Overlay, PV 102+ setting panel, Spanish
003203
Overlay, PV 102+ setting panel, Italian
003204
Overlay, PV 102+ setting panel, Dutch
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
5.3.2
PARTS DRAWINGS AND PART-NUMBER LISTS
BREAS
Part no.
Definition
003205
Overlay, PV 102+ setting panel, Danish
003206
Overlay, PV 102+ setting panel, French
003207
Overlay, PV 101+ setting panel, Swedish
003208
Overlay, PV 101+ setting panel, English
003209
Overlay, PV 101+ setting panel, German
003210
Overlay, PV 101+ setting panel, Finnish
003211
Overlay, PV 101+ setting panel, Spanish
003212
Overlay, PV 101+ setting panel, Italian
003213
Overlay, PV 101+ setting panel, Dutch
003214
Overlay, PV 101+ setting panel, Danish
003215
Overlay, PV 101+ setting panel, French
003249
Decal, PV 101+ logo
003250
Decal, PV 102+ logo
003261
Sticker “Plus”
Part-number list – Main assembly
BREAS
Part no.
Definition
000071
Mains input socket
000529
Spacer cooling fan PV 101+/PV 102+
000789
Bolt MRX-H-M4x12H fzb
001041
Fan (cooling air PV 101+/PV 102+)
002959
Power board PV 101+/PV 102+
002961
MDA board PV 101+/PV 102+
1310005
(Eribel part no) M3 Compound nut
1310011
(Eribel part no) DSS M3050x8 Spacer
1310012
(Eribel part no) 5 mm DSUB Spacer
1310020
(Eribel part no) MRXZ 3x6 Steel Screw
1310029
(Eribel part no) RXS-Z ST2, 9x9, 5 Screw
1310110
(Eribel part no) Spacer with securing pin, 10 mm
1310114
(Eribel part no) DSS M3050x10 Spacer
1360074
(Eribel part no) Base plate PV 101+/PV 102+
1410006
Mains power fuse, 110–220V, 1A T 5x29 mm
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
5–9
PARTS DRAWINGS AND PART-NUMBER LISTS
5.3.3
5–10
PV 101+/PV 102+ Service Manual
Part-number list – Blower unit and patient outlet
BREAS
Part no.
Definition
000032
Pressure sensor - MPX 10 DP
000640
Y-connector
000675
M4, Serrated washer
000677
Nut, M6M M4 Metal fzb
000789
Bolt MRX-H-M4x12H fzb
001000
Blower cover, upper
001001
Blower cover, lower with silencer
001003
Blower fan wheel
001004
Bushing, blower fan wheel
001005
Patient air inlet
001027
O-ring 7.66 x 1.78
001028
O-ring 38 x 1
001032
LF isolator
001033
Honeycomb material
001038
Flex tube ∅ 22 x 15 mm
001039
Silicon tube, 5 x 8 mm (blue/green)
001052
Silicon tube, 50 mm (red)
001053
Silicon tube, 30 mm (red)
001054
Silicon tube, 125 mm (blue)
001055
Silicon tube, 150 mm (green)
001056
Silicon tube, 70 mm (black)
001074
Spacer DSSM4070x25
001079
Blower casing, upper
001080
Blower casing, lower
001081
Sealing cover for upper casing
001082
Sealing cover for upper blower casing
001083
Rubber bushing
001111
Blower casing outlet 22 mm
001117
Flow sensor AWM2100V
001122
Seal for upper blower casing
001245
Screw, RX-PT 4-20*6 fzb
001665
Shoulder screw M4 4,3
002966
Compound nut M4, PV 101+/PV 102+
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
5.3.4
PARTS DRAWINGS AND PART-NUMBER LISTS
BREAS
Part no.
Definition
003110
Turbine Exchange kit PV 101+/PV 102+
Part-number list – Miscellaneous
BREAS
Part no.
Definition
000245
Hytrel patient tube
000491
Absorbent, dwg 10646-4 (T2) outer edge outer wall
000503
Absorbent, dwg 10648-1 (T2) inner edge upper casing outer wall
000504
Absorbent, dwg 10648-2 (T2) inner edge upper casing round
000505
Absorbent, dwg 10648-3 (T2) inner edge upper casing U
000506
Absorbent, dwg 10648-4 (T2) inner edge upper casing bottom
000507
Absorbent, dwg 10648-5 (T2) inner edge upper casing outer
000508
Absorbent, dwg 10648-6 (T2) inner edge upper casing outer
000509
Absorbent, dwg 10648-7 (T2) inner edge upper casing outer
000510
Absorbent, dwg 10648-8 (T2) inner edge upper casing outer
000511
Absorbent, dwg 10649-1 (T3) inner edge lower casing inner
000512
Absorbent, dwg 10649-2 (T3) inner edge lower casing inner
000522
Absorbent dwg 10665-1 silencer 1001 round
000523
Absorbent dwg 10665-2 silencer 1001 inside
000711
Cable tie 71 mm
000997
Flow restrictor nipple 6 mm
001073
Washer, 12 x 5 x 2
001934
Thommen HM 28 digital manometer
002965
Screw Mrt M5 X 12 Torx, PV 101+/PV 102+
002973
PV 101+ Operating manual, Swedish
002974
PV 101+ Operating manual, German
002975
PV 101+ Operating manual, Finnish
002976
PV 101+ Operating manual, English
002977
PV 101+ Operating manual, Norwegian
002978
PV 101+ Operating manual, Spanish
002979
PV 101+ Operating manual, Italian
002980
PV 101+ Operating manual, Dutch
002981
PV 101+ Operating manual, Danish
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
5–11
PARTS DRAWINGS AND PART-NUMBER LISTS
5–12
BREAS
Part no.
Definition
002983
PV 101+ Operating manual, Portuguese
002982
PV 101+ Operating manual, French
002984
PV 101+ Operating manual, Greek
002985
PV 101+ Operating manual, Polish
002986
PV 102+ Operating manual, Swedish
002987
PV 102+ Operating manual, German
002988
PV 102+ Operating manual, Finnish
002989
PV 102+ Operating manual, English
002990
PV 102+ Operating manual, Norwegian
002991
PV 102+ Operating manual, Spanish
002992
PV 102+ Operating manual, Italian
002993
PV 102+ Operating manual, Dutch
002994
PV 102+ Operating manual, Danish
002995
PV 102+ Operating manual, French
002996
PV 102+ Operating manual, Portuguese
002997
PV 102+ Operating manual, Greek
002998
PV 102+ Operating manual, Polish
003036
EMC-protection, Flow meter
003116
Screw K6sf M4x16mm
003220
Connector
003221
Connector element
003243
PV 101+ Operating manual, Japanese
003244
PV 102+ Operating manual, Japanese
003259
PV 101+/PV 102+ Service manual, English
003302
Fan control unit
003311
Flow restrictor nipple 5 mm
003312
Flow restrictor nipple 4 mm
BREAS MEDICAL
PV 101+/PV 102+ Service Manual
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
6
FUNCTIONAL DIAGRAMS
FUNCTIONAL DIAGRAMS
This chapter contains a diagram of the pneumatic system of the ventilator and a block
diagram of the PV 101+/PV 102+’s functions.
6.1
Pneumatic diagram
The pneumatic diagram shows the pneumatic components of the air circulation system
of the PV 101+/PV 102+.
Sensor 2
Red
Sensor 1
Green
Blue
Black
No.
Description
1
Patient air inlet (through air filter)
2
Blower
3
Motor
4
Silencer
5
Flow sensor (2)
6
Pressure sensor (1)
7
Honeycomb material
8
Reference pressure tube
9
Patient air tube
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
6–1
FUNCTIONAL DIAGRAMS
6.2
PV 101+/PV 102+ Service Manual
Functional block diagram
The block diagram below shows how the electronics are designed and how they are
connected to the other components.
6–2
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
7
ELECTRONICS
ELECTRONICS
This chapter describes in detail the electronics of the PV 101+/PV 102+. It also contains
circuit diagrams and component location drawings for the circuit boards of the machine.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
7–1
ELECTRONICS
7.1
PV 101+/PV 102+ Service Manual
Electronics – Main components
The electronics in the PV 101+ and the PV 102+ comprises a blower unit and three
printed circuit boards: the CPU board, the power board and the MDA board.
1
2
3
4
5
6
8
7–2
BREAS MEDICAL
7
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
ELECTRONICS
No
Description
Location
1
CPU board
In the upper casing
2
Blower unit, incl. motor
Mounted on base plate
3
Power board
Mounted to bottom of base plate
4
MDA board
Mounted to rear of base plate
5
Hour counter
Mounted to MDA board
6
Analogue/digital
connection
Mounted to MDA board
7
Battery connection
Mounted to rear of MDA board
8
Power connection
Mounted to rear of base plate
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
7–3
ELECTRONICS
7.2
PV 101+/PV 102+ Service Manual
Functional block diagram
The block diagram below shows on an overall level how the electronics are designed
and how they are connected to the other components.
7–4
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
7.3
ELECTRONICS
Internal wiring diagram
The wiring diagram displays the internal cabling and connections of the electronics.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
7–5
ELECTRONICS
7.4
PV 101+/PV 102+ Service Manual
Circuit board descriptions
This section contains detailed descriptions and figures of the PV 101+/PV 102+’s circuit
boards.
7.4.1
CPU board
All the push-buttons are mounted on the CPU board together with the LEDs and the
three bar-graphs for pressure display. Also mounted here is the four 7-segment displays
for the rise time and backup rate displays (PV 102+).
The microprocessor with its software is also located here, as well as an internal memory
(DALLAS) for the logging and storage of the patient usage data.
Refer also to the circuit diagrams (See section 7.6.1, 7.6.2, 7.6.3, 7.6.4 and 7.6.5) and
component location drawings (See section 7.7.1).
7.4.2
Power board
The power supply unit is fed primarily with mains power via the mains power inlet to the
internal AC/DC converter. Secondly, the power board can be fed with 12–24 DC via the
external battery input.
Refer also to the circuit diagrams (See section 7.6.6 and 7.6.7) and component location
drawings (See section 7.7.2 and 7.7.3).
The power supply has mains power as soon as the power cord is connected.
7–6
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
7.4.3
ELECTRONICS
MDA board
X201 connector
P203
P204
P201
P202
X202
connector
Flow sensor
Pressure
sensor
The MDA (motor drive assembly) board comprises the driver stage for the blower motor,
the pressure sensor, the flow sensor, the hour counter and the D-shell connector for
serial communication.
The driver stage is built around the U205 control circuit and the Q205–Q210 transistors.
The P21 buzzer is powered by the C221 supercap if the mains power supply should fail.
The pressure sensor, which is of the semi-conductor type, is amplified by U201. The offset and gain can be adjusted using the P200 and P201 potentiometers. The signal is
used for regulating the pressure.
The flow sensor is also of the semi-conductor type, a mass flow meter. The flow that
passes the sensor is proportional to the flow through the flow resistance in the air tube
from the blower. The offset and gain can be adjusted using the P202 and P203 trim
potentiometers. The signal is used for initiating the trigger and the mask-off function.
Refer also to the circuit diagrams (See section 7.6.8, 7.6.9 and 7.6.10) and component
location drawings (See section 7.7.4).
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
7–7
ELECTRONICS
7.5
PV 101+/PV 102+ Service Manual
Electrical safety measurements
Electrical safety measurements must be made in accordance with EN 60 601-1. However, you can make an insulation resistance measurement instead of the voltage test
specified by the standard. The measurements can be made using an automatic electrical safety tester. All tests must be performed in accordance with class II type BF.
Supply voltage
•
Note the power voltage reading.
The voltage must be noted for each service check, as the currents measured are
directly in relation to the supply voltage. This allows all measurements made on the
same machine to be compared with measurements made on different occasions.
Insulation
•
Measure the insulation resistance using a 500V DC power supply.
The most suitable method is to connect the plus lead to the two ventilator power socket
pins and the minus lead to the casing or patient air connector.
The measurements made during the delivery inspection can be used as reference values for measurements made during future services. If no reference values are available,
the value for the insulation resistance should be >20MΩ.
Leakage currents
•
Measure the leakage currents at different parts of the ventilator using an RC circuit
to ground.
The measurements are made partly at normal case (NC) and at the single fault condition (SFC). Reverse the polarity of the power supply and note the highest value.
The leakage currents to ground must not exceed the stated limit values.
Leakage currents from the casings
•
Measure the leakage current of the casing at an unpainted point, for example, the
head of a screw.
Limit values:
NC
SFC
Break neutral for SFC.
< 0.1mA
< 0.5mA
Patient leakage currents
•
Measure the patient leakage currents between the patient connector and ground.
Limit values:
NC
SFC
Break neutral for SFC.
< 0.1mA
< 0.5mA
Leakage currents with mains power supply at the patient connected part
•
Perform this test using an automatic electrical safety tester with this function.
See the safety instructions for the tester.
Limit value:
7–8
SFC
< 5mA
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
7.6
ELECTRONICS
Circuit diagrams – Circuit boards
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
7–9
C109
100n
100n
C134
C114
100n
100n
C135
C115
100n
100n
C138
C128
T6
T5
RES
P54
P55
C100
100n
C101
100n
VCC
1
3
4
5
8
13
14
7
R102
10k
Vcc
V-
16
2
6
15
9
12
11
10
VCC
P23
C102
100n
C103
100n
F1
C104
P57
A12
A13
A14
22pF
VCC
P56
A11
P21
P55
A15
P54
A10
A9
A8
A7
A6
P63
P62
P61
P60
A0
A1
A2
A3
A4
A5
P64
P65
P66
P67
1
26
35
52
53
54
55
56
57
58
59
60
61
62
63
64
27
28
29
30
31
32
33
34
D7
D6
D5
D4
D3
D2
D1
D0
XTAL100
VCC
P60
P61
P62
P63
P64
P65
P66
P67
P50
P51
P52
P53
P54
P55
P56
P57
P20
P21
P22
P23
P24
P25
P26
P27
10MHz
U102
NMI
Vcc
Vss
BA
LIR
R/W
WR
RD
E
Vss
XTAL
EXTAL
MP0
MP1
RES
STBY
A15
A14
A13
A12
A11
A10
A9
A8
A7
A6
A5
A4
A3
A2
A1
A0
HD63C03YRF
36
37
38
39
40
41
42
43
A7
A6
A5
A4
A3
A2
A1
A0
D0
A14
A13
A8
A9
A11
/OE
A10
/CS
D7
44
45
46
47
48
49
50
51
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
U104
D7
A10
A11
A9
A8
A13
A14
29F010
29
28
27
26
25
24
23
22
21
D7
D6
D5
D4
D3
D2
D1
D0
P21
P23
P24
P25
P50
P51
P52
C105
22pF
A0
A1
A2
A14
A15
A0
A1
A2
A3
A13
P53
1k
P54
R107
P55
P56
P57
P60
P61
P62
P63
P64
P65
P66
P67
A14
A13
A12
A11
A10
A9
A8
A7
A6
A5
A4
A3
A2
A1
A0
1
26
2
23
21
24
25
3
4
5
6
7
8
9
10
2
3
4
5
6
7
8
9
1
11
12
13
U106:D
19
18
17
16
15
14
13
12
10
D2
D3
D4
D5
D6
D7
D7
D6
D5
D4
D3
D2
D1
D0
11
D1
4
5
1
2
9
10
D0
PV101/102 Upgrade
U106:B
74HC00
U106:A
U106:C
74HC00
6
3
8
Date
74HC00
Sheet
19
18
17
16
15
13
12
11
14
20
27
22
74HC00
I/O7
I/O6
I/O6
I/O5
I/O4
I/O3
I/O2
I/O0
U105
I1
I2
I3
I4
I5
I6
I7
I8
GND
D7
D6
D5
D4
D3
D2
D1
D0
GND
CE
WE
OE
CLK
OE
GAL16V8_PLCC
U103
A14
A13
A12
A11
A10
A9
A8
A7
A6
A5
A4
A3
A2
A1
A0
DS1644
Title
Axel Olofsson
CPU/Display
cad056a_R3.sch
Number
Drawn by
A3
Size
Filename
CE2
CE3
CE4
CE9
CE8
CE7
CE6
CE5
D0
D1
D2
D3
D4
D5
D6
D7
Rev
of
2002-10-30
2
5
Issue: X-1
Doc. No. 003259 En
U101
C1+
GND
V+
C2+
C1-
C2R2OUT
R1OUT
T1IN
T2IN
DS14C232
R2IN
R1IN
T1OUT
T2OUT
P24
VCC
R103
R104
R105
R106
4*4k7
1
2
3
4
5
6
7
8
9
10
3,3k
M1
MODUL_1
VCC
A7
A6
A5
A4
A3
A2
5
6
7
8
9
10
11
12
13
D3
P53
D0
D4
A1
A0
D5
P52
D6
U100
8
7
6
5
D1
Vcc
WP
SCL
SDA
A12
4
3
2
1
32
31
30
A12
A15
A16
NC1
Vcc
/WE
NC2
D1
D2
GND
D3
D4
D5
D6
14
15
16
17
18
19
20
D2
A0
A1
A2
Vss
24LC01B-I
3,3k
P56
1
2
3
4
R101
P57
10u/6V
10u/6V
P25
+
C111
100n
100n
P50
100n
C129
P51
T1
100n
C139
VCC
100n
C133
VCC
100n
+
C140
+
C110
C108
C132
BREAS MEDICAL
7–10
Circuit diagram – CPU board 1, rev 2
7.6.1
PV 101+/PV 102+ Service Manual
ELECTRONICS
C131
100n
10u/6V
R100
1
11
U109
10
1
2
3
4
5
6
7
8
9
10
M2
S2
D7
LD2
LD100
D6
D7
D6
D7
CE6
CE8
RES
D7
D6
D5
D4
D3
D2
D1
D0
D7
D6
D5
D4
D3
D2
D1
D0
D0
D1
D2
D3
D4
D5
D6
D7
D5
D5
CE9
D1
D2
D3
D4
D1
D2
D3
D4
D0
D0
CE2
3
4
7
8
13
14
17
18
1
11
3
4
7
8
13
14
17
18
1
11
2
3
4
5
6
7
8
9
1
11
2
3
4
5
6
7
8
9
74HC273
1D
2D
3D
4D
5D
6D
7D
8D
CLR
CLK
U118
74HC273
1D
2D
3D
4D
5D
6D
7D
8D
CLR
CLK
U114
74HC574
1D
2D
3D
4D
5D
6D
7D
8D
OC
CLK
U115
74HC574
1D
2D
3D
4D
5D
6D
7D
8D
OC
CLK
1Q
2Q
3Q
4Q
5Q
6Q
7Q
8D
GND
1Q
2Q
3Q
4Q
5Q
6Q
7Q
8D
GND
1Q
2Q
3Q
4Q
5Q
6Q
7Q
8Q
GND
1Q
2Q
3Q
4Q
5Q
6Q
7Q
8Q
GND
LD107
470R
R117
LD105
2
5
6
9
12
15
16
19
10
2
5
6
9
12
15
16
19
10
19
18
17
16
15
14
13
12
10
19
18
17
16
15
14
13
12
DA_S1_8
DA_S1_7
DA_S1_6
DA_S1_5
DA_S1_4
DA_S1_3
DA_S1_2
DA_S1_1
DA_S2_8
DA_S2_7
DA_S2_6
DA_S2_5
DA_S2_4
DA_S2_3
DA_S2_2
DA_S2_1
16Q
15Q
14Q
13Q
12Q
11Q
10Q
9Q
470R
R118
LD106
470R
M3
R-2R (10K-20K)
1
2
3
4
5
6
7
8
9
10
S1
R-2R (10K-20K)
D6
POWER ON/OFF
OK100
- KEY
OK101
FUNCTION
OK102
SLEEP
OK104
+ KEY
OK103
D4
D5
D6
D7
D0
D1
D2
D3
D0
D1
D2
D3
D4
D5
3
4
7
8
13
14
17
18
1
11
2
3
4
5
6
7
8
9
1
11
1D
2D
3D
4D
5D
6D
7D
8D
OC
CLK
U116
OK1
P56
P57
P55
P54
74HC273
1D
2D
3D
4D
5D
6D
7D
8D
CLR
CLK
U117
74HC574
560R
R128
470R
R114
LD102
1k
R109
GND
100k
R111
470R
R110
1Q
2Q
3Q
4Q
5Q
6Q
7Q
8D
GND
1Q
2Q
3Q
4Q
5Q
6Q
7Q
8Q
2
5
6
9
12
15
16
19
10
19
18
17
16
15
14
13
12
10
DP4
DP3
DP2
DP1
LD103
L119-EGW_2
470R
R113
LD101
470R
470R
R115
R108
R116
1Q
8Q
7Q
6Q
5Q
4Q
3Q
2Q
P51
R127
2k2
P25
P51
P50
T1
T6
T5
1
2
5
6
8
9
12
13
10u/6V
C112
+
4093
GND
Vcc
U112
D5
D6
Q104
7
14
3
4
10
11
16
15
14
13
12
11
10
9
+
U119
R
R
WDO
N.C
N.C
WDI
N.C
GND
BC847B
D3
D2
D1
D0
D7
D6
D5
D4
D3
D2
D1
D0
D7
CE7
CE5
1
2
3
4
5
6
7
8
4k7
R112
C136
100n
C113
10u/6V
MAX699CWE
GND
Vcc
Vcc
GND
N.C
N.C
N.C
N.C
R123
GND
Vcc
F
G
A
B
C
D
E
GND
Vcc
F
G
A
B
C
D
E
GND
Vcc
F
G
A
B
C
D
E
GND
Vcc
F
G
A
B
C
D
E
47R
4511
LT
BI
LE
A
B
C
D
4511
U111
LT
BI
LE
A
B
C
D
4511
U110
LT
BI
LE
A
B
C
D
4511
U108
LT
BI
LE
R126
2k2
3
4
5
7
1
2
6
3
4
5
7
1
2
6
3
4
5
7
1
2
6
3
4
5
U107
A
B
C
D
Q103
VCC
1
2
3
4
RN103
RN105
2k2
6
3
5
15
14
13
7
4
1
2
3
4
4027
J
Vcc
CLK
Q
K
Q
S
R
J
CLK
K
Q
S
Q
R
GND
U113
1
2
3
4
8
7
6
5
RN102
1
2
3
4
RN104
RN106
16
1
2
12
11
10
9
8
C117
1n
8
7
6
5
1k
C120
1n
T6
C121
1n
BC847B
Q101
Q100
BC847B
R119
C119
100n
47R
C118
1n
2k2
R124
Q102
8
7
6
5
8
7
6
5
8
7
6
5
R121
MNR34-J5AJ-471_2
8
7
6
5
MNR34-J5AJ-471_2
1
2
3
4
MNR34-J5AJ-471_2
8
7
6
5
8
7
6
5
BC847B
C116
1n
47R
R122
RN107
MNR34-J5AJ-471_2
1
2
3
4
RN100
MNR34-J5AJ-471_2
MNR34-J5AJ-471_2
1
2
3
4
1
2
3
4
RN101 MNR34-J5AJ-471_2
MNR34-J5AJ-471_2
R125
VCC
VCC
VCC
VCC
BC847B
8
16
15
14
13
12
11
10
9
8
16
15
14
13
12
11
10
9
8
16
15
14
13
12
11
10
9
8
16
15
14
13
12
11
10
9
T1
7
1
2
6
P50
D4
P51
LD104
VCC
5
10
9
7
6
4
2
1
5
10
9
7
6
4
2
1
5
10
9
7
6
4
2
1
5
10
9
7
6
4
2
1
C126
100n
C122
100n
S2
CE4
C130
R132
C127
100n
C123
1n
47R
R120
3
8
DP4
3
8
DP3
3
8
DP2
3
8
DP1
LD2
VCC
Filename
Drawn by
A3
Size
1
2
3
4
5
6
7
8
9
10
1
2
3
4
5
6
7
8
9
10
1
2
3
4
5
6
7
8
9
10
1ar
2ar
3ar
4ar
5ar
6ar
7ar
8ar
9ar
10ar
1ag
2ag
3ag
4ag
5ag
6ag
7ag
8ag
9ag
10ag
1ar
2ar
3ar
4ar
5ar
6ar
7ar
8ar
9ar
10ar
1ag
2ag
3ag
4ag
5ag
6ag
7ag
8ag
9ag
10ag
1ar
2ar
3ar
4ar
5ar
6ar
7ar
8ar
9ar
10ar
1ag
2ag
3ag
4ag
5ag
6ag
7ag
8ag
9ag
10ag
OK1
S1
S2
T6
T5
P25
P51
P50
T1
LD2
VCC
P54
P55
P57
P56
21
22
23
24
25
26
27
28
29
30
21
22
23
24
25
26
27
28
29
30
21
22
23
24
25
26
27
28
29
30
cad056a_R3.sch
Axel Olofsson
CPU/Display
Sheet
Date
PV101/102 Upgrade
3Q
4Q
5Q
6Q
7Q
8Q
1Q
2Q
3Q
4Q
LED106
MV59164
5Q
6Q
7Q
8Q
1Q
2Q
3Q
4Q
5Q
6Q
LED105
MV59164
7Q
8Q
1Q
2Q
3Q
4Q
5Q
6Q
7Q
8Q
LED104
MV59164
Number
C124 C125 C107
100n 100n 100n
Title
TDSG1160
LED103
470R
TDSG1160
R131
LED102
470R
R130
TDSG1160
470R
LED101
R129
TDSG1160
LED100
470R
S1
CE3
100n
1k
R133
R134
P25
C137
T5
BREAS MEDICAL
100n
OK1
20
19
18
17
16
1k
2k
3k
4k
5k
10k
9k
8k
7k
6k
11
12
13
14
15
20
19
18
17
16
1k
2k
3k
4k
5k
10k
9k
8k
7k
6k
11
12
13
14
15
20
19
18
17
16
1k
2k
3k
4k
5k
10k
9k
8k
7k
6k
Doc. No. 003259 En Issue: X-1
11
12
13
14
15
16Q
15Q
X100:8
Rev
X100:3
X100:2
X100:13
X100:12
X100:11
X100:9
X100:7
X100:6
X100:5
X100:4
X100:10
X100:1
X100:14
2
of
2002-10-30
11Q
12Q
13Q
14Q
15Q
16Q
9Q
10Q
11Q
12Q
13Q
14Q
15Q
16Q
9Q
10Q
11Q
12Q
13Q
14Q
15Q
16Q
9Q
10Q
11Q
12Q
13Q
14Q
2
5
7.6.2
470k
PV 101+/PV 102+ Service Manual
ELECTRONICS
Circuit diagram – CPU board 2, rev 2
7–11
R145
R144
10k
10k
10k
10k
10k
10k
10k
R135
R146
20k
R147
R136
20k
R137
20k
R138
20k
R139
20k
R140
20k
R141
20k
R142
20k
R143
20k
R148
F1
P60
P61
P62
P63
P64
P65
P66
P67
R149
BI1
BI2
BI3
BI4
BI5
BI6
BI7
BI8
BI9
BI10
T1
R150
Title
Size
PV101/102 Upgrade
Date
2
2002-10-30
Rev
Sheet
CPU/Display
Axel Olofsson
Number
cad056a_R3.sch
A3
Drawn by
of
Filename
2
5
Issue: X-1
Doc. No. 003259 En
BREAS MEDICAL
7–12
Circuit diagram – CPU board 3, rev 2
7.6.3
PV 101+/PV 102+ Service Manual
ELECTRONICS
BI10
BI9
BI8
BI7
BI6
BI5
BI4
BI3
BI2
S2
DA_S2_8
DA_S2_7
DA_S2_6
DA_S2_5
DA_S2_4
DA_S2_3
DA_S2_2
DA_S2_1
20k
R159
20k
R158
20k
R157
20k
R156
20k
R155
20k
R154
20k
R153
20k
R152
20k
R160
R161
R162
R163
R164
R165
BREAS MEDICAL
R166
R151
10k
10k
10k
10k
10k
Doc. No. 003259 En Issue: X-1
10k
cad056a_R3.sch
Filename
CPU/Display
Sheet
Date
PV101/102 Upgrade
Axel Olofsson
Number
Drawn by
A3
Size
Title
2
of
2002-10-30
Rev
2
5
7.6.4
10k
BI1
PV 101+/PV 102+ Service Manual
ELECTRONICS
Circuit diagram – CPU board 4, rev 2
7–13
DA_S1_1
DA_S1_2
DA_S1_3
DA_S1_4
DA_S1_5
DA_S1_6
DA_S1_7
DA_S1_8
S1
R178
R177
R176
10k
10k
10k
10k
10k
10k
10k
R167
R179
20k
R180
R168
20k
R169
20k
R170
20k
R171
20k
R172
20k
R173
20k
R174
20k
R175
20k
R181
BI1
BI2
BI3
BI4
BI5
BI6
BI7
BI8
BI9
BI10
R182
Title
Size
PV101/102 Upgrade
Date
2
2002-10-30
Rev
Sheet
CPU/Display
Axel Olofsson
Number
cad056a_R3.sch
of
Drawn by
A3
Filename
2
5
Issue: X-1
Doc. No. 003259 En
BREAS MEDICAL
7–14
Circuit diagram – CPU board 5, rev 2
7.6.5
PV 101+/PV 102+ Service Manual
ELECTRONICS
CN2:3
CN2:2
CN2:1
100-240V AC
4k7
6
5
2
3
-
R3
1
U1:B
4k7
R4
PWR_CTRL
7
LM324Y
U1:A
CTRL_IN
CTRL_OUT
LM324Y
8k2
+
-
+
Ledig
SG37
R6
9
10
-
+
100n
2
R1
C59
R61
100n
470k
C54
L1
220n
2n2 Y1
C3
8
4k7
R9
+5V
LM324Y
U1:C
11
220uH
C5
12
13
U1:D
4
15k
D3
+VB
14
2k2
R59
+VB
LM324Y
R16
+
-
R15
680R
10uF/16V
+
LM324Y
GND Vcc
U1:E
100n
C2
2n2 Y1
C4
TR1
+VB
TR4
R17
BZX84C5V1
CNY17F-3
U2
470R
2
3 1
8
7
6
5
C6
OPTO_A
1
D1
R2
BZX84C3V6
3k3
C55
4k7
R8
100n
R7
R5
D2
1M/0.6W
680R
BZX84C5V1
C1
R10
R62
4k7
+
C58
R14
R11
4k7
VARI_S14K300
R12
0u33/35V
C9
CN1
C7
C8
OUT
GND
GND
NC
4 6
5
78L05ACM
U9
68nF
IN
GND
GND
NC
BR1
1
2
3
4
-
DB107G
+5V
+
C57
1n/400V
C13
+
BC817-40
Q1
1k
R22
Q2
BZX84C24
D4
BC817-40
Q3
BC807-40
R28
1
2
3
4
5
6
7
REF
NC
Vcc
Vc
OUT
GND
PGND
8k2
R31
UC 3845D
COMP
NC
VF
NC
SENSEC
NC
RC
U3
R32
D5
400V_DC
5k6/2W
14
13
12
11
10
9
8
DIOD_SMC
4n7/400V XY
Monteras ej
C12
100n
1n/400V
1n/400V
R18
3k3
R19
18k
1n/400V
4k7
56R
100u/400V
C15
R24
R25
R26
100k
R13
1k5
1k5
C14
R23
R91
10n
3k9
47R
1n/400V
R27
4k7
270k
47k
8k2
220k/0.6W
R29
9
1
2
13
33541-901
Q4
100k
R102
1k
R34
150R
R33
BC807-40
(9) 11
(7)
(11)
TR2
TRAFO_ETD34
BAS21
D8
D9
+
47k
R99
100k
SPP20N60S5
Q5
G
BC807-40
Q16
100k
R56
Q18
S
D
BC817-40
BYD37M
D10
C22
330p/400V
BYT12PI-1000
R98
C21
10u/50V
Q17
R36
270R/0.6W
12 (10)
5 (3)
7 (5)
10 (8)
AC_PWR
BYD37M
10R
C24
R42
C61
C16
220k
100n/400V
R104
R105
C20
+
R30
2n2
D21
R101
C18
1k5
68R
C17
33u/50V
330p
C66
BZX84C20
100k
+
22k
+
R92
4n7/400V XY
BAS21
C10
R20
R21
+
C11
C26
1000uF/50V
BC807-40
Q13
100k
R53
L2
22uH
+VB
Filename
TP3
+
4
8
-
+
6
5
+5V
100k
R52
DC_OUT
100n
PSU - AC/DC Power
cad055a_R3.sch
CN2:6
CN2:7
CN2:4
CN2:5
03-01-23
1 of
Sheet
Rev
Date
Breas Medical AB
LM393
GND Vcc
U4:C
100n
C28
BC817-40
Q6
D
U4:B
C53
LM393
7
Q7
"PMOS_TO220"
RXE 185
R90
A. Olofsson
Number
+5V
Drawn by
A3
Size
Title
R60
CONNECTED TO PCB SHIELD
R35
22u
R97
D7
R38
R39
0R68/1W
100k
220p/400V
C65
D13
100k
R103
4k7
68k
100n
D6
100n
BYD37M
C60
R37
BZX84C22
R100
33R
R40
G
+
0R68/1W
47k
D12
C29
S
R51
C25
47k
D11
100n
6k8
K1
2k2
100k
C23
R43
R54
R55
C27
R50
R48
R49
3
2
TP5
10k
1000uF/50V
BZX84C10
6k8
10k
4k7
BREAS MEDICAL
4k7/2W
22k
R57
R58
Doc. No. 003259 En Issue: X-1
10k NTC
7.6.6
47u/16V
PV 101+/PV 102+ Service Manual
ELECTRONICS
Circuit diagram – Power board 1, rev 3
7–15
2n2
C19
S3
ISO
ISO
1k
R64
R78
Q10
Q15
R95
100k
R96
MTD20P06HDL
D15
U8
CNY17F-3
2k2
R84
1
3
BZX84C5V1
TR3
TRAFO_ETD29
D18
5
12
10
7
D19
BYD37M
D22
BYT12PI-1000_2
D20
AC_PWR
?
L3
CN2:8
Breas Medical AB
PSU - AC/DC Power
Date
Sheet
Bat/Main
2 of
03-01-23
3
100n
Rev
DC_OUT
BZX84C6V8
D
S
4
ISO
Number
A. Olofsson
C52
BC807-40
100k
G
D
S
C56
cad055a_R3.sch
100u/50V
100k
Q9
Q14
ISO
Q12
STP60NE06-16FP_2
R77
10R
G
Title
Size
100R
10k
A3
100R
R87
Drawn by
R89
10k
R75
C42
100n
R88
1
3
2
Q8
BC817-40
ISO
R79
D16
R93
10k
ISO
C44
OPTO_A
3
1
2
100k
680p
U6
CNY17F-3
L4
R71
R74
+
100n
C50
220p/400V
Filename
+
C51
C35
R68
10k
10u/20V
C47
C63
220p/400V
R66
G
BAS21
Q11
STP60NE06-16FP_2
10R
4n7/50V
47R/1W
47k
16
15
14
13
12
11
10
9
D17
R76
R83
C31
220p/400V
C64
220p/400V
U7
INV
VREF
NI
VCC
E/A
OUTB
CLK
VC
RT
PGND
CT
OUTA
RAMP
GND
SS
ILIM/SD
1k
R86
5
6
4
1k
0R1/1W
CN3
47R
D
1k
R94
R72
R81
R82
R70
1
2
3
4
5
6
7
8
UC3825
C41
R80
C46
4n7/50V
47R/1W
C45
R85
S
100n
6k8
10n
ISO
C43
100k
BAS21
1
6.3A Trög
C39
4k7
100n
68k
R69
12k
C40
680p
2n2
C36
10k
+
R65
C37
R67
100n
47u/6V
56R
100n
10n
C38
2
12-24V DC
1000u/50V
5
4
6
1000u/50V
C34
R63
+
C33
+
C32
2
Issue: X-1
Doc. No. 003259 En
BREAS MEDICAL
7–16
Circuit diagram – Power board 2, rev 3
7.6.7
PV 101+/PV 102+ Service Manual
ELECTRONICS
BYW29-500
D14
R73
470k
0R1/1W
1
OUT+
GND OUT-
Vcc
SENSOR200
3
C271
LM274
GND Vcc
U201:E
100nF
C204
4
4
2
+10V
TRYCKGIV MPX10DP
11
+5V
X206:7
X206:1
X206:4
X206:5
X206:6
X206:2
X206:3
BAS32_2
X207
2
1
BAS32_2
D202
MOLEXBATTIN_VINK
C201
C280
C202
2
P203
IN
OUT
3,3K
1K
3
R255
R201
1
GND
OFFSET
{Value}
U213
5k
3,3K
R200
220uF/50V
C220
3
2
6
5
+
+
-
-
D204
LM274
U201:A
1
LM274
7
U201:B
{Value}
RXE020
S200
BAS32_2
5
1
REL200
TIMER CURTIS701P
GND
IN
4
2
3
4
3
NTC 47k
100R
R202
100R
R214
P204
20k
6
2
4
2
4
8
3
GAIN
BC847B
Q215
IN
OUT
GND FB
ON/OFF
GND
Q203
LM2597M-12
LM2576S-5
1
3
5
U210
IN
OUT
Vbs
FLAG
SD/SS
GND DELAY
FEEDBACK
BC857B_2
Q200
7
1
5
6
D225
270R
R203
270R
R216
D215
330uH
9
10
10R
R292
BYG90-40
BC847B
Q216
BYG90-40
DR206
HM50-331K
330uH
10K
R236
4.7k
8
3,3K
R291
C221
LM274
R208
+
D213
+5V
100uF/25V
U201:C
4.7k
R217
-
+12V
+
BAS85_2
C244
DR209
HM50-331K
+
Monteras ej!!!
R232
1K
OUT
OUT
OUT
OUT
2
7
3
6
BC847B
Q201
TL317C
IN
NC
NC
ADJ
U202
R205
0,47F
1
8
5
4
2
3
13
12
1
-
+
-
+
LM393
1
U206:A
LM274
14
U201:D
2
3
+
3,3K
R233
+10V
R256
10R
R270
1
8
5
4
680R
1K
R230
2,2K
R231
+5V
BC847B
OUT
OUT
OUT
OUT
TL317C
IN
NC
NC
ADJ
U203
R213
Q202
PZ200
OBO-1206C
DSS306-91FZ103N
1,6K
R210
DR205
10K
R235
BAS85_2
D212
R211
Vcc
R
100nF
D200
100nF
Y1
100nF
Y1
+
R319
VR203
C281
220R
3,3K
R234
+
C209
1
2
Mont Ej
+35V
C206
RESET TIMER
R218
R219
4,7uF/16V
TIMER200
C212
X205:6
X205:7
C207
C203
1K
D214
1
1
2
C210
DR208
BAS85_2
2
3
3
1
Filename
2
100nF
Axel O
cad057a_R3.sch
Number
MDA
1
Sheet
of
2003-01-14
Rev
X202:10
X202:9
X200:9
X200:6
X202:1
X202:7
X200:1
X202:2
X202:8
X202:3
X200:7
Date
PV101/102 Upgrade MDA
1nF
C216
1nF
C217
3
DR210
DSS306-91FZ103N
DSS306-91FZ103N
Drawn by
A3
Size
Title
D219
BAS85_2
220R
10uF/6V
D203
DR201
DSS306-91FZ103N
2
7
3
6
R212
+
1,5K
R204
R209
560R
4k7
150k
47k
VR204
SM6T15CA
X205:4
X205:5
100nF
100nF
39k
18k
C245
C211
C205
C254
0,1uF/50V
Mont Ej
C213
+V
1nF
10k
R207
R206
1k
10k
R215
100nF
C261
100nF
C256
C259
100nF
C260
100nF
C257
R293
Mont Ej
C258
100nF
C255
D205
BREAS MEDICAL
SM6T15CA
Doc. No. 003259 En Issue: X-1
C253
100nF
100nF
100nF
3
3
7.6.8
Mont Ej
PV 101+/PV 102+ Service Manual
ELECTRONICS
Circuit diagram – MDA board 1, rev 3
7–17
+12V
+
1
SENSOR201
+10V
+
LM274
14
U214:D
5k1
R312
+10V
R238
10k
U214:A
1
LM274
R311
1k
Q214
RDF16N06LESM
1k
X202:5
5
6
-
+
LM393
7
U206:B
1K
R229
R228
1
+5V
1
3
3
2
3
33uF/16V
C237
DR202
DSS306-91FZ103N
2,2K
DR204
DSS306-91FZ103N
2
2
DR203
DSS306-91FZ103N
1
+12V
D208
D207
D206
SM6T15CA
SM6T15CA
DR207
BL01 RN1-A62
Title
100nF
Number
Axel O
MDA
X200:8
X202:11
X202:12
X200:5
X203:2
X203:1
X200:2
X200:3
1
2003-01-14
Rev
Date
of
Sheet
PV101/102 Upgrade MDA
C264
cad057a_R3.sch
3
3
Issue: X-1
Doc. No. 003259 En
+
-
7
10k
10K
R284
100K
R283
-
+
R237
3
270R
LM358Y
2
R225
270R
R226
+10V
P202
1k
+
-
7
4.7k
U209:B
R223
10K
R282
4.7k
R224
+12V
5
6
C238
0,1uF/50V
R227
X202:6
SM6T15CA
A3
Size
100nF
U214:B
8
LM274
LM274
U214:C
P201
20k
+
-
+
R280
820K
4
C262
1nF
12
13
5
6
R222
1k
9
10
D220
BAS85_2
C219
100nF
U206:C
Vcc GND
C215
1nF
4
5
3
2
6
C246
4u7F 25V
U209:A
1
R275
+5V
8
Drawn by
100nF
C214
C265
R220
+
LM358Y
10K
5,6K
R276
DZ201
BZX84C6V8
11
Monteras ej!!!
R221
3
2
-
R279
4,7K
C218
100nF
U214:E
Vcc GND
10k
100n
C283
100n
+10V
4
R309
2k2
C284
4
C263
Filename
100nF
R310
24.9k
24.9k
C208
100nF
U209:C
Vcc GND
LM393
+
VR202
KTY 81-220
VR201
KTY 81-220
+12V
8
LM274
3,3k
AWM 2100V SENSOR
C231
VR201 Monteras ej!
1uF/35V
C282
100n
LM358Y
Monteras ej!!!
R281
BREAS MEDICAL
7–18
Circuit diagram – MDA board 2, rev 3
7.6.9
PV 101+/PV 102+ Service Manual
ELECTRONICS
C266
C267
+35V
8
+12V
X202:4
2,7k
R257
LM358Y
1uF/35V
4
LM393
GND Vcc
8
+V
C226
U200:C
Vcc GND
LM393
100nF
C252
0,1uF
+35V
BAS32_2
4
4,7K
R243
4,7K
R244
4,7K
R242
C236
330uF/25V
+
+5V
D221
100nF
U207:C
2,2K
10K
Mont. ej!
VR200
C247
3k3
R274
3
R260
8
2
DR212
VR200 Monteras ej
1
C224
R259
Vcc GND
U212:C
100nF
C276
4
10nF
C243
4,7K
3,3K
R241
1,82K
C225
R268
3
1
DR211
2
R269
10nF
X205:2
X205:3
C223
C275
100nF
X205:1
X205:8
+
NE555
1k
6
5
2
3
-
4k7
R313
U207:A
LM393
LM393
5
6
1
7
470K
R318
+
-
BAS32_2
D238
MC33035PY
BRAKE
ENABLECS
DIR
CSIN
60/120
EAIN
EAOUT
GND
CSLIN
OSC
SPEED
AT
BT
CT
BAS32_2
FAULT
AB
BB
CB
U205
Vc
Vcc
SA
SB
SC
VREF
U207:B
+5V
-
+
+
18k
R263
23
7
3
22
12
13
16
10
11
18
17
4
5
6
8
+12V
U212:B
7
LM358Y
15
9
14
21
20
19
2
1
24
+35V
+
4,7uF/16V
C232
4,7uF/16V
C233
C230
0,33uF
10nF
C228
R316
R317
10nF
C200
D210
10R
D235
3,3nF
47K
47K
3,3K
R285
10K
R247
+V
100R
R239
100R
R288
+
+
-
+
47k
R295
BAS32_2
47k
R297
220k
R296
6
5
Q213
BC847B
C242
R301
LM393
7
U200:B
+V
D237
BAS32_2
Q205
BZX84C10
D218
Q206
R300
R299
+35V
RFD16N06LESM
RDF16N06LESM
10K
?
?
R264
BAS32_2
D239
R246
330pF
TRIG Vcc
R
THR
DIS OUT
GND CON
R261
R314
R315
0,47u
12K
15K
1K
50k
P205
+12V
3,3K
R286
BAS32_2
D232
2
3
100R
R289
-
+
Q212
BC847B
10K
R278
D233
C241
1k5
R272
LM393
1
U200:A
82K
R277
BAS32_2
0,22R
10K
C222
BAS32_2
D209
+
C272
R240
100nF
R249
R251
R262
C234
D217
RDF16N06LESM
Q207
Q208
RFD16N06LESM
3
+
-
22R
G
Q217
MTD20N06HDL
R294
+
S
D
+35V
U212:A
1
LM358Y
4,7uF/16V
2
100K
BZX84C10
R298
+12V
330pF
8
6
3
5
R266
R267
10K
4k7
U208
4R7 Mont. lodsida
2
4
7
1
100k
8k2
D223
R302
D224
R271
D211
BZX84C10
10K
R258
BZX84C10
2k2
BAS32_2
R273
1K
R250
10K
3k3
R254
BAS32_2
D234
R253
4R7 Mont. lodsida
100R
R290
3,3K
R287
R308
R307
R306
Q211
BC847B
3*4k7
+5V
DO-214AB(SMC)
D227
C251
C278
Q209
BZX84C10
D216
Filename
Drawn by
A3
Size
Title
1
2
3
4
1
2
3
4
U211
8
7
6
5
1k
R303
C248
+35V
C279
100nF
100nF
Axel O
cad057a_R3.sch
Number
MDA
Sheet
Date
Rev
X202:13
X204:2
X204:1
X201:3
X201:5
X201:4
X202:14
X201:2
X201:8
X201:6
X201:7
X201:1
100nF
1
of
2003-01-14
PV101/102 Upgrade MDA
100nF
+5V
RFD16N06LESM
Q210
+35V
8
7
6
5
BAS32_2
D236
MICRO_PIC12C67
100nF
RDF16N06LESM
10K
+5V
1nF
C269
R304
+5V
D222
1nF
C249
+5V
BZX84C10
1nF
1nF
C270
10k
2k
R305
+12V
D228
D229
R252
DO-214AB(SMC)
D230
C240
DO-214AB(SMC)
D226
DO-214AB(SMC)
D231
330pF
DO-214AB(SMC)
BREAS MEDICAL
DO-214AB(SMC)
1nF
C250
C268
C277
1nF
C274
Doc. No. 003259 En Issue: X-1
C273
10uF 35V
3
3
PV 101+/PV 102+ Service Manual
ELECTRONICS
7.6.10 Circuit diagram – MDA board 3, rev 3
7–19
10K
R265
ELECTRONICS
7.7
7–20
PV 101+/PV 102+ Service Manual
Component location drawings – Circuit boards
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
7.7.1
ELECTRONICS
Component location drawing – CPU board
For component lists, refer to section 7.8.1.
U102
C125
C116
C117
C123
C118
C107
C119
C120
C121
C122
C126
C127
C124
1
+
C131
R143 R150
R142 R149
R141 R148
R140 R147
R139 R146
R138 R145
R137 R144
R136 R135
XTAL100
R101
U100
U119
C115
C133
C110
C134
R103
R104
R105
R106
C139
R100
R107
C101 C100
X100
C132
C105
C104
C102
C138
U105
U104
14
1
U101
1
C103
U106
C128
LED104
LED105
R121
Q101
R120
R125
Q102
R124
R122
LD106
C135
R117
R127
R126
R175
R174
R173
R172
R171
R170
R169
R168
R159
R158
R157
R156
R155
R154
R153
R152
C129
LED106
Q104
R123
R182
R181
R180
R179
R178
R177
R176
R167
R166
R165
R164
R163
R162
R161
R160
R151
U116
Q103
C109
LD105
C140
R118
LD107
+
R115
U114
R102
OK104
U118
U115
C113
C136
C137
+
+
U113
U112
OK101
C112
R132
R133
LED100
R113
R114
LD103
R116
R128
LD102
LED101
C114
LD104
LED102
LED103
U117
R129
C130
R130
C108
R111
R110
Q100 R119
R112
LD100
R109
R108
R134
OK103
OK102
LD101
RN101
RN102
RN103
RN104
U109
RN105
RN106
RN107
+
RN100
R131
C111
U110
U108
U107
2960
U111
OK100
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
7–21
ELECTRONICS
7.7.2
PV 101+/PV 102+ Service Manual
Component location drawing – Power board, upper side
For component lists, refer to section 7.8.2.
CN2
R90
L3
Q7
D9
D22
D20
L2
+
+
+
C24
C51
C27
1
R36
C4
TR3
R43
R82
R83
U2
C23
C18
C46
TR2
R32
C15
+
C19
R86
Q12
Q11
C3
C45
0 AL
+
R28
R85
2959
4
D1
+
R40
R37
Q5
R103
C33
+
1
C21
C6
C60
L4
TR1
S3
C13
L1
CN3
R8
C32
+
C1
C12
~
BR1
-
TR4
R14
+ ~
C6
BREAS MEDICAL
CN1
R6
C10
C8
C7
7–22
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
7.7.3
ELECTRONICS
Component location drawing – Power board, under side
For component lists, refer to section 7.8.2.
R15
D3
Q6
R58
R50
TP3
R1
TP5
U1
R51
C5
R52
R57
C55
R61
R5
R9
R16
R11
R12
+
C56
R49
C53
R48
C59
C54
R59
U4
R2
R89
C64
R7
R88
R3
C2
R17
U9
C58
C52
R4
R60
+
R10
D1
D12
R13
R55
R54
C28
D2
D18
C63
R87
C47
C57
C50
C29
R62
C26
C22
C31
R84
D19
R63
+
R79
C37
R67
R78
R75
Q14
R93
R94
1
C44
R64
R95
Q10
R20
Q9
R74
Q8
R96
D11
C43
R70
+
+
R21
D15
C11
Q15
R8
1
Q1
D5
D7
D10
C42
+
R73
R72
Q4
C38
R69
C41
C39
C40
R65
D21 D13
K1
R101
R102
C16
C17
+
C66
Q17
R104
R105
R100
R99
R97
Q18
Q13
R56
R98
Q16
R53
C35
D8
R30
R18
R19
R26
C14
C36
R66
R34
U3
R27
C9
R39
D6
C20
R31
R29
C34
U7
Q2
R35
R33
R71
R68
Q3
R25
R22
R24
D4
D16
R38
D17
R23
R91
R76
C65
C25
R42
R77
R80
R92
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
7–23
ELECTRONICS
7.7.4
PV 101+/PV 102+ Service Manual
Component location drawing – MDA board
For component lists, refer to section 7.8.3.
C230
C249
U208
C228
R264
+
C232
R242
C251
R244
D210
D209
R262
C250
R243
C269
D222
R253
R287
Q211
X201
Q209
Q210
D216
C268
C240
R252
Q207
R286
Q212
R250
Q208
D217
C270
D223
C248
C241
U210
+
+
C265
DR207
VR201
R275
R276
U209
R282
R283
X206
C221
R279
D220
R280
01
C220
D204
R281
C2
R234
R212
U203
+
R213
Q201
D203
D219
DZ201
R233
R292
200
DR201
R256
C210
R293
DR210
DR202
Q215
+
R291
TIMER200
C253
C236
D215
DR205
DR204
DR206
D213
Q216
D212
X200
C262
+
P205
D225
Q203
REL200
R236
DR208
DR203
X204
DR209
R257
R235
D208
D207
C238
C208
P203
G
P204
G
P201
C212
R206
R214
C209
U201
R216
R217
Q214
C205
R203
+
C207
R204
C211
R284
R232
R205
X207
C280
C202
O
D206
C245
C272
Q202
C257
D214
D205
R314
R298
R277
C231
D233
C234
R278
D221
R313
D238
R274
C275
0
R259
S20
C259
PZ200
C254
R300
R299
C244
R241
C224
R308
R307
R306
R263
R260
U212
R301
R239
C222
R261
R219
R218
C281
C278
U207
C226
D234
U205
U213
D235
U211
C277
D236
R316
C276
R240
R317
R302
R288
R289
C279
R304
R305
R290
C242
R285
R303
C200
VR200
C243
D239
R246
R268
+
C233
C225
D224
Q213
R269
R318
D218
C247
R315
Q205
Q206
R247
R249
C204
C203
C206
R207
R310
R309
R215
C255
C213
C237
1
C261
C258
R229
R230
U206
R228
C219
R231
C264
P202
O
R202
X203
C263
U202
R270
R210
R211
R209
R208
X205
C223
C266
R227
C271
U214
X202
C217
C216
R221
R224
C215
+
C214
C274
D230
R222
R223
R225
D229
D228
R220
C260
C282
C218
R200
Q200
C256
C284
VR202
C267
R312
D200
R238
R226
D202
R201
R255
R237
DR211
C283
R311
DR212
C246
14
D227
D231
D226
D237
R266
R254
R258
R251
SENSOR201
C252
U200
R295
D232
R297
R267
SENSOR200
R272
D211
R271
R296
R294
7–24
R265
Q217
C273
R273
+
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
ELECTRONICS
7.8
Component lists – Circuit boards
7.8.1
Component list – CPU board
Reference
Description
Quantity
C100-103, 107-109
CAPACITOR 100 nF X7R
27
C104, 105
CAPACITOR 22 pF NPO
2
C110-113, 140
CAPACITOR 10 uF/6V
5
C116-118, 120, 121
CAPACITOR 1 nF NPO
6
LD100
LED Red 3mm
1
LD101, 106
LED Green 3mm
2
LD102, 104, 105, 107
LED Green L113GDT 2x5mm
4
LD103
LED Red/Green Circular
1
LED100-103
TDSG1160 7-seg.
4
LED104, 105, 106
Bargraph PV 101+/PV 102+
3
OK100-104
SWITCH 3ESH9-10,4
5
Q100-104
TRANSISTOR BC847B
5
R100, 101
RESISTOR 3k3 1%
2
R102
RESISTOR 10k 1%
1
R103-106
RESISTOR 4K7 1%
4
R107, 109, 119, 133
RESISTOR 1k 1%
4
R108, 110, 113-118
RESISTOR 470R 1%
12
R111
RESISTOR 100k 1%
1
R112
RESISTOR 4k7 1%
1
R120-123
RESISTOR 47R 1%
4
R124-127
RESISTOR 2k2 1%
4
R128
RESISTOR 560R 1%
1
R134
RESISTOR 470K 1%
1
R135-143, 151-159
RESISTOR 20k 1%
27
R144-150, 160-166
RESISTOR 10K 1%
21
RN100-107
RN RESNET MNR34-J5AJ-471
8
U100
IC 24LC01 WSOITR
1
U101
IC DS14C232CWM SO16W (232CB)
1
U102
IC HD63C03YRF
1
(U104)
IC PLCC32 SMD
1
U105
IC GAL16V8H-25SC/4
1
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
7–25
ELECTRONICS
7.8.2
7–26
PV 101+/PV 102+ Service Manual
Reference
Description
Quantity
U106
IC 74HC00
1
U107, 108, 110, 111
IC 4511
4
U109, 115, 116
IC 74HC574
3
U112
IC 4093
1
U113
IC 4027
1
U114, 117, 118
IC 74HC273
3
U119
IC MAX699CWE
1
XTAL100
XTAL 10 MHz HC49/4H
1
Component list – Power board
Reference
Description
Quantity
BR1
BRIDGE DB154G
1
C1
CAPACITOR 220nF PME271
1
C2, 20, 26, 28, 29, 34
CAPACITOR 100nF X7R
17
C3, 4
CAPACITOR 2,2nF Y1 DE1210E222M
2
C5
CAPACITOR 10u/16V
1
C6
CAPACITOR 68nF PME271
1
C7, 8, 10, 12, 15
CAPACITOR 1nF/400V
5
C11, 37
CAPACITOR 47u/16V
2
C13
CAPACITOR 100uF 400V
1
C14, 36, 41
CAPACITOR 10nF X7R
3
C16, 25, 39
CAPACITOR 2,2nF X7R
3
C17
CAPACITOR 330pF NPO
1
C18
CAPACITOR 33u/50V
1
C19
CAPACITOR 100nF/400V
1
C21
CAPACITOR 10uF 50V 1MOD RSH
1
C22
CAPACITOR 330p/500V
1
C23
CAPACITOR 220pF 400V
1
C24, 27, 32, 33
CAPACITOR 1000uF 50V
4
C31, 50, 63, 64
CAPACITOR 220pF 500V
4
C40, 42
CAPACITOR 680p NPO
2
C43
CAPACITOR 10u/20V
1
C45, 46, 60, 61
CAPACITOR 4,7nF 500V XY
4
C51
CAPACITOR 100u/50V
1
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
ELECTRONICS
Reference
Description
Quantity
C58
CAPACITOR 0,33u/35V
1
C65
CAPACITOR 22u/10V
1
C66
CAPACITOR 4,7uF/16V
1
CN1
CONNECTOR 3P
1
CN2
CONNECTOR 8P
1
CN3
CONNECTOR 2P
1
D1
DIODE BZX84C3V6
1
D2, 3, 18
DIODE BZX84C5V1
3
D4
DIODE BZX84C24
1
D5
DIODE RS3K SMC
1
D6, 8, 16, 17
DIODE BAS21
4
D7, 10, 11, 19
DIODE BYD37M
4
D9, 20, 22
DIODE BYT12PI-1000
3
D12
DIODE BZX84C10
1
D13
DIODE BZX84C22
1
D14
DIODE BYW29F200
1
D15
DIODE BZX84C6V8
1
D21
DIODE BZX84C20
1
L1
INDUCTOR 220uH HM50-221K
1
L2
INDUCTOR 22uH HM50-220K
1
L3
INDUCTOR 330uH KM5
1
L4
INDUCTOR TSL1112RA
1
Q1, 3, 6, 8, 9, 14, 17, 18
TRANSISTOR BC817-40
5
Q2, 4, 10, 13, 16
TRANSISTOR BC807-40
5
Q5
TRANSISTOR SPP20N60S5
1
Q7
TRANSISTOR IRF5305 TO220
1
Q11, 12
TRANSISTOR IRFIZ44N TO-220
2
Q15
TRANSISTOR MTD20P06HDL
1
R1, 4, 5, 9, 10, 13, 20,
RESISTOR 4k7 1%
11
R2, 18
RESISTOR 3k3 1%
2
R3, 27, 31
RESISTOR 8K2 1%
3
R11, 12, 29
RESISTOR 1K5 1%
3
R14
VARISTOR S14K300
1
R16
RESISTOR 15K 1%
1
R17
RESISTOR 470R 1%
1
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
7–27
ELECTRONICS
7–28
PV 101+/PV 102+ Service Manual
Reference
Description
Quantity
R19
RESISTOR 18K 1%
1
R21, 67
RESISTOR 56R 1%
2
R22, 34, 72, 73,
RESISTOR 1k 1%
6
R23
RESISTOR 3k9 1%
1
R25
RESISTOR 270k 1%
1
R26, 60, 71, 92, 98
RESISTOR 47k 1%
5
R28, 103
RESISTOR 220k 0,6W
2
R30
RESISTOR 68R 1%
1
R32
RESISTOR 5K6 2W
1
R33
RESISTOR 150R 1%
1
R35, 54
RESISTOR 22k 1%
2
R36
RESISTOR 270R 0,6W
1
R37, 40
RESISTOR 0R68 1W PR01
2
R38, 76, 77
RESISTOR 10R 1%
3
R39
RESISTOR 33R 1%
1
R42, 52, 62, 75, 78, 79
RESISTOR 100k 1%
15
R43
RESISTOR 4K7 2W
1
R48, 57, 64, 65, 87
RESISTOR 10k 1%
7
R50, 51, 70
RESISTOR 6K8 1%
3
R55, 59, 84
RESISTOR 2k2 1%
3
R58
RESISTOR 10k NTC B57621C103
1
R6
RESISTOR B57236-S200-M NTC
1
R61, 94
RESISTOR 470K 1%
2
R63, 91
RESISTOR 47R 1%
2
R66, 105
RESISTOR 68k 1%
2
R69
RESISTOR 12K 1%
1
R7, 15
RESISTOR 680R 1%
2
R8
RESISTOR 1M 0,6W
1
R82, 83
RESISTOR 0R1 0,6W
2
R85, 86
RESISTOR 47R O,6W
2
R88, 89
RESISTOR 100R 1%
2
R90
RESISTOR RXE185 / 3.5 A fuse (wire type)
1
S3
FUSE HOLDER Schurter 5x20m
1
TR1
TRANSFORMER 12uH PLA10
1
TR2
TRANSFORMER ETD34-3F3
1
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
7.8.3
ELECTRONICS
Reference
Description
Quantity
TR3
TRANSFORMER ETD29
1
TR4
TRANSFORMER 7uH PLA10A7720R7D02
1
U1
IC LM324Y
1
U2, 6, 8
IC CNY17F-3 DIL6
3
U3
IC UC 3845D
1
U4
IC LM393D
1
U7
IC UC3825DW
1
U9
IC LM78L05ACM
1
Component list – MDA board
Reference
Description
Quantity
C200
CAPACITOR 0,47u/25V
1
C201, 202
CAPACITOR 2,2nF Y1 DE1210E222M
2
C209, 232-234
CAPACITOR 4,7uF/16V
4
C210
CAPACITOR 10uF/6V
1
C213-217, 249-251
CAPACITOR 1nF NPO
11
C220
CAPACITOR 220u/50V
1
C221
CAPACITOR 0,47F FSOH
1
C222, 224, 225
CAPACITOR 10nF X7R
3
C223
CAPACITOR 10nF X7R
1
C226
CAPACITOR 0,1uF/100V MKT
1
C228
CAPACITOR 3,3nF X7R
1
C230
CAPACITOR 1uF/50V
1
C231, 243
CAPACITOR 1uF/35V
2
C236
CAPACITOR 330uF/25V CV-GX
1
C237
CAPACITOR 33uF/16V
1
C238, 245
CAPACITOR 0,1uF/50V X7R
2
C240-242
CAPACITOR 330pF NPO
3
C244
CAPACITOR 100uF/25V CV-GX
1
C246
CAPACITOR 4,7uF/35V ELYT D55
1
C273
CAPACITOR 10uF/50V MVKF55
1
C282-284, 203, 204
CAPACITOR 100nF X7R
35
D200, 202, 204, 209
DIODE BAS32
15
D203, 212, 213, 219
DIODE BAS85
5
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
7–29
ELECTRONICS
7–30
PV 101+/PV 102+ Service Manual
Reference
Description
Quantity
D205-208, 214
DIODE SM6T15CA
5
D215, 225
DIODE BYS10-45
2
D216-218, 222-224
DIODE BZX84C10
6
D226-231
DIODE RS3K SMC
6
DR201-205, 208
INDUCTOR DSS306-91FZ103N100
9
DR206, 209
INDUCTOR 330uH HM50-331K
2
DR207
INDUCTOR BL01 RN1-A62
1
DZ201
ZENER DIODE BZX84C6V8
1
P201, 204
POTENTIOMETER 20K TYP3224J
2
P202
POTENTIOMETER 1K TYP3224J
1
P203
POTENTIOMETER 5K TYP3224J
1
P205
POTENTIOMETER 50K TYP3224J
1
PZ200
BUZZER OBO-1206C
1
Q200
TRANSISTOR BC857B
1
Q201, 202, 211-213
TRANSISTOR BC847B
7
Q203
TRANSISTOR LM2575S-5
1
Q205-210, 214
TRANSISTOR RFD16N06LESM
7
Q217
TRANSISTOR MTD20N06HDL
1
R200, 233, 234
RESISTOR 3k3 1%
3
R201, 256
RESISTOR 1k 1%
2
R202, 214, 288-290
RESISTOR 100R 1%
5
R203, 216, 225, 226
RESISTOR 270R 1%
4
R204
RESISTOR 39K 1%
1
R205, 222, 227, 229
RESISTOR 1k 1%
11
R206, 215, 235-238
RESISTOR 10k 1%
15
R210
RESISTOR 1k6 1%
1
R211, 212
RESISTOR 220R 1%
2
R213
RESISTOR 680R 1%
1
R218
RESISTOR 560R 1%
1
R220, 221
RESISTOR 24K9
2
R228, 231, 260, 271, 310
RESISTOR 2k2 1%
5
R239
RESISTOR 100R 1%
1
R240
RESISTOR 0,22R 3W AC03
1
R241-244
RESISTOR 4k7 1%
4
R246, 247, 249, 250
RESISTOR 10k 1%
6
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
ELECTRONICS
Reference
Description
Quantity
R255, 268, 281, 285
RESISTOR 3k3 1%
12
R257
RESISTOR 2k7 1/2W 5%
1
R263, 209
RESISTOR 18K 1%
2
R264, 270, 292
RESISTOR 10R 1%
3
R266, 283, 298
RESISTOR 100k 1%
3
R267
RESISTOR 8K2 1%
1
R269
RESISTOR 1k82 1%
1
R276
RESISTOR 5k6 1%
1
R277
RESISTOR 82k 1%
1
R279, 219, 251, 313
RESISTOR 4k7 1%
8
R280
RESISTOR 820k 1%
1
R293, 272
RESISTOR 1K5 1%
2
R294
RESISTOR 22R 1%
1
R295, 297, 316, 317
RESISTOR 47k 1%
4
R296
RESISTOR 220k 1%
1
R305
RESISTOR 2k 1%
1
R312
RESISTOR 5k1 1%
1
R314
RESISTOR 12K 1%
1
R315
RESISTOR 15K 1%
1
R318
RESISTOR 470K 1%
1
REL200
RELAY RT114024
1
S200
POLYFUSE RXE020
1
U200, 206, 207
IC LM393D
3
U201, 214
IC TLC274CDR
2
U202, 203
IC LM317LM
2
U205
IC MC33035DW
1
U208
IC NE555
1
U209, 212
IC LM358D
2
U210
IC LM2597M-12
1
U213
IC LM317
1
VR202
SENSOR KTY81-220
1
X201
CONNECTOR ANGLED 8P
1
X202
CONNECTOR 1x14POLE
1
X203, 204
CONNECTOR 2P
2
X205
CONNECTOR 8P
1
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
7–31
ELECTRONICS
7–32
PV 101+/PV 102+ Service Manual
Reference
Description
Quantity
X207
CONNECTOR ANGLED 2P
1
BREAS MEDICAL
Doc. No. 003259 En
Issue: X-1
PV 101+/PV 102+ Service Manual
8
FAULT TRACING
FAULT TRACING
This chapter contains a fault-tracing table that may be used when trouble-shooting the
device.
8.1
Fault-tracing table
If the PV 101+/PV 102+ does not work properly try to identify the problem in the table
below. Check the possible causes and carry out the suggested remedial actions.
Symptom
Possible cause
Remedial action
See ref.
The device alarms at
start-up.
• The power cord is not properly connected.
• Connect the power cord.
2.4.2
• Mains fuse has blown.
• Replace the fuse.
7.1
• Leakage from patient circuit/
mask.
• Check the tubes and mask
for leakage.
2.6.6
• Internal leakage.
• Check the tubes inside the
machine.
2.5.3
• Dirty filters.
• Replace or wash the filters.
2.6.5
The ventilator does
not give adequate
pressure/volume.
1 Adjust voltage to transformer.
2 Replace MDA board or
pressure sensor.
4.3
• Trigger settings not correct.
• Adjust settings.
See
PV 101+/
PV 102+
Operating
manual
• Tube to flow sensor has
worked loose.
• Check tube and connectors.
7.4.3
• Pressure sensor faulty.
The PV 101+/
PV 102+ does not
trigger.
• Flow sensor faulty.
• MDA board faulty.
Alarms for mask-off
(PV 102+ only).
• Leakage from patient circuit/
mask.
Doc. No. 003259 En Issue: X-1
1 Adjust voltage to pressure
sensor.
2 Replace MDA board or flow
sensor.
• Check the tubes and mask
for leakage.
BREAS MEDICAL
3.7
4.5
3.7
2.6.6
8–1
FAULT TRACING
8–2
PV 101+PV 102+ Service Manual
BREAS MEDICAL
Doc. No. 003259 En Issue: X-1
PV 101+/PV 102+ Service Manual
9
APPENDICES
APPENDICES
This chapter contains the engineering change history of the PV 101+ and the PV 102+.
Included in the chapter is also a service record and a service report. These pages can
be copied and used for completing the reports.
9.1
Engineering change history – PV 101+
The table below lists the engineering changes made to the PV 101+ by serial number.
From serial no. Changes
9.2
R060334
Software rev 3. Hardware MDA rev. 2, PSU rev. 3, CPU rev. 3
Production start.
R090177
Software upgrade rev 4.
Ref. internal improvement no. 337: The PV 101+ incorrectly turned
itself off 15 seconds after ramp activation when there was no patient
circuit leakage.
R250001
(mandatory field
upgrade from
R020296 to
R230389)
Replacement of the R90 component on the power board
(ref. TBD33).
A 3.5 A fuse (wire link type) has replaced the former PTC fuse, in
order to avoid risk of disturbances due to ambient temperature.
Engineering change history – PV 102+
The table below lists the engineering changes made to the PV 102+ by serial number.
From serial no. Changes
R060335
Software rev 3. Hardware MDA rev. 2, PSU rev. 3, CPU rev. 3
Production start.
R090279
Software upgrade rev 4.
Ref. internal improvement no. 337: The PV 102+ incorrectly turned
itself off 15 seconds after ramp activation when there was no patient
circuit leakage.
R250001
(mandatory field
upgrade from
R020296 to
R230389)
Replacement of the R90 component on the power board (ref. TBD33).
A 3.5 A fuse (wire link type) has replaced the former PTC fuse, in
order to avoid risk of disturbances due to ambient temperature.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
9–1
APPENDICES
9–2
PV 101+/PV 102+ Service Manual
BREAS MEDICAL
Doc. No. 003259 En Issue: X-1
PV 101+/PV 102+ Service Manual
APPENDICES
Service Record for BREAS PV 101+/PV 102+
Use a photocopy of this service record for the maintenance inspection described
in chapter 2 in this service manual. Use the reverse side for comments and notes.
Service Record No:__________
Registration
Model:...................................
Serial No................................... Inventory No.
..................
Accessories: ...............................................................................................................................
Delivery date:......................................
No. of operating hours: ..............................................
Service started:...................................
Signature: ..................................................................
Service completed: .............................
Signature: ..................................................................
Delivered, date: ..................................
Signature: ..................................................................
See instruction ref.
Check OK
Preparing inspection
Verifying the components and the installed software
Initial recording
Markings
Information from the patient/user
Validity of the documentation
2.3.1
2.3.2
2.3.3
2.3.4
2.3.5
.........
.........
.........
.........
.........
2.4.1
2.4.2
2.4.3
.........
.........
.........
Cleaning the inside of the machine
Inspecting the internal wiring
Securing the components
Checking the power supply
Checking Electrical safety measurements
2.5.2
2.5.3
2.5.4
2.5.5
7.5
.........
.........
.........
.........
.........
Blower Assembly (15,000 hours)
Checking if blower unit should be replaced
3.5
.........
Calibration (12 months/15,000 hours)
Calibrating the pressure sensor
Calibrating the flow sensor
Calibrating the high pressure limit
4.3
4.5
4.6
.........
.........
.........
External inspection
Visual inspection for external damage and wear
Checking the power supply connection
Short function check
Internal inspection
Continued on next page
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
9–3
PV 101+/PV 102+
See instruction ref.
Check OK
Final inspection before delivery
Function check
Checking the low pressure leakage alarm (PV 102+)
Checking the trigger function
Checking the pressure and rate
Replacing the filters
Checking the accessories
Adjusting the settings for the patient
2.6.1
2.6.2
2.6.3
2.6.4
2.6.5
2.6.6
2.6.7
.........
.........
.........
.........
.........
.........
.........
Notes
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
BREAS MEDICAL
Doc. No. 003259 En Issue: X-1
PV 101+/PV 102+ Service Manual
9.3
APPENDICES
Returning products to BREAS
You may need to return the device or any components or accessories to BREAS, for
example, for service, warranty, upgrade or repair. If so, follow the instructions below to
ensure that the correct action is taken and to avoid unnecessary delays.
1
Pack the product in its original packaging. If this is not available pack the product
in packaging suitable for transporting the product to BREAS.
2
Take a photocopy of the delivery report on the next page.
3
Fill out the Customer information part of the service report and pack it together
with the product to be returned. The service report will be completed by BREAS
and returned with the product.
Product damage caused by poor packaging or during transport is not covered by the factory warranty.
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
9–5
APPENDICES
9–6
PV 101+/PV 102+ Service Manual
BREAS MEDICAL
Doc. No. 003259 En Issue: X-1
PV 101+/PV 102+
Service Report for BREAS PV 101+/PV 102+
BREAS Ref. No:__________
Customer information
Customer name:...................................................................................................................
Customer address:...............................................................................................................
.............................................................................................................................................
Phone:..................................................................................................................................
Reference person:................................................................................................................
Customer reference number: ...............................................................................................
Product information
Model: .......................
Serial No:..........................
Operating hours: ........................h
Error /Complaint / Accessories
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
.............................................................................................................................................
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
PV 101+/PV 102+
Information from BREAS
Date received by BREAS:......................
Repair
Warranty
Signature ...................................................
Update
Charge
Other
Action taken:
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
Parts used:
Pcs:
Price:
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
Operating hours from BREAS:_____________h
Date returned to customer: ________________ Signature:____________________
Notes:
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
BREAS MEDICAL
Doc. No. 003259 En Issue: X-1
PV 101+/PV 102+ Service Manual
APPENDICES
Notes
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
9–9
APPENDICES
PV 101+/PV 102+ Service Manual
Notes
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
9–10
BREAS MEDICAL
Doc. No. 003259 En Issue: X-1
10
PV 101+ OPERATING MANUAL
BREAS MEDICAL
BREAS MEDICAL
11
PV 102+ OPERATING MANUAL
BREAS MEDICAL
BREAS MEDICAL
Download