Uploaded by Kittiwat Wongsuwan

KTMED MVSA Angelus Vital-Signs-Monitor Service-Manual-v1.10

advertisement
KTMED’s Vital Sign Monitor
ANGELUS
Service Manual
Ver 1.10
KTMED, Inc.
#610, Hansin IT Tower, 235,
Guro-3Dong,
Seoul, Korea
822.2108.7890
Fax: 822.2108.7894
www.ktmed.com
CONTENTS
Section 1 - General Information
Instruction-------------------------------------------------------------------------------------3
Revision History------------------------------------------------------------------------------3
About KTMED Monitor--------------------------------------------------------------------4
General Cautions----------------------------------------------------------------------------------5
Warning----------------------------------------------------------------------------------------5
Warranty----------------------------------------------------------------------------------------6
Service----------------------------------------------------------------------------------------6
Composition-----------------------------------------------------------------------------------7
Monitor Block Diagram-------------------------------------------------------------------------8
Module descriptions------------------------------------------------------------------------------9
Section 2 - Functional Tests and Trouble Shooting
Check List 1: Power-------------------------------------------------------------------------12
Check List 2: Power Cord------------------------------------------------------------------12
Check List 3: Fuse----------------------------------------------------------------------------12
Check List 4: Charge-------------------------------------------------------------------------12
Check List 5: Discharge----------------------------------------------------------------------12
Troubling Shooting----------------------------------------------------------------------13
Error message lists ---------------------------------------------------------------16
Section 3 - Disassembly and Assembly
Prior to Disassembly ---------------------------------------------------------------19
Disassembly Battery --------------------------------------------------------------------19
Disassembly Front and Rear Case---------------------------------------20
Disassembly Controller Board-----------------------------------------------21
Disassembly NIBP& SpO2 Board----------------------------------------22
Disassemb ly Power Board----- - --------- --- --------- --- --------- --- -----24
Replacing Fuse-----------------------------------------------------------------25
Section 4 - Service
SpO2 Calibration -------------------------------------------------------------26
NIBP Calibration------------------------------------------------------------------29
Section 5 - Repair Part Lists
Repair Part Lists-------------------------------------------------------------------44
Section 6 - Contact to Manufacturer
Contact Information ------------------------------------------------------------45
Service Request Form ------------------------------------------------------------46
Section 1 – General Information
Meaning of Symbols used in this manual
Symbols are used to specially emphasize the agreed details as follows. Users need to surely
follow all the caution and notes listed in this Manual.
In the event that the product is damaged due to misuse or negligence by a user, the
manufacturer or authorized dealers shall not be responsible for any damage or loss to the
product.
※ Warning
The title “Warning” is used to inform the users of possible causes that could inflict injury,
death, or properly damage to the patients.
※ Caution
The title “Caution” is used to inform the users of possible causes that could inflict injury to
the patients although it might not severe enough to cause deaths.
※ Note
The title “Note” is used to inform the users of items that are of importance in terms of
installation, operation, or maintenance of the Product although the failure does not inflict
the bodily harm to the patients.
Dangerous warning should never be included under the “Note” signal.
SYMBOLS
The International Electro-technical Commission (IEC) has established a set of
symbols for medical electrical unit regarding connections or warnings of potential
hazards.
The classification and symbols are shown below;
Symbol
Description
Type BF applied parts
-2-
Section 1 – General Information
INSTRUCTION
This manual contains information needed to service the ANGELUS manufactured by
KTMED INC.
This manual purposes to assist its easy maintenance and repair of ANGELUS.
To help you to familiarize yourself quickly with this manual, a series of check list diagram
and explanations have been provided.
For specific information, please refer to the Angelus Operating Manual.
Revision History
Revision No.
1.1
Description
Calibration procedures updated
-3-
Date
2009.3.27
Remark
Section 1 – General Information
About KTMED Monitor
ANGELUS displays numerical values for functional arterial hemoglobin oxygen saturation
(SpO2) in percentage, pulse rate in beats per minute and NIBP on light emitting diode
(LED) digital displays
You can expect outstanding monitoring performance with Angelus during a surgery or
intensive monitoring care
The monitor is powered by AC Power or by a rechargeable battery pack.
A battery was added to enable the monitor to be used if there is a power outage or to be
unplugged momentarily when moved with the patient from the surgery room to the
recovery room. The ANGELUS monitor should be plugged into AC as much as possible to
give you the maximum battery backup time when there is a power outage. It will maintain
unit operation
-4-
Section 1 – General Information
General Cautions
„
„
„
„
„
„
„
„
„
„
„
„
„
„
Use a grounded wrist strap before starting the assembly and disassembly of the
monitor.
The monitor could be damaged by water and liquid. Do not repair the unit where
water or liquid falls on.
Do not remove the battery during charging
The battery is consumable parts and when purchasing, compared
The use hour is gradually decreases.
The recharging time depends on the voltage level and ambient temperature of the
dry cell.
Do not immerse the monitor in any fluid
Do not dispose of the battery in the regular waste collection.
Do not operate the monitor unless it has been calibrated properly
The monitor does not operate effectively if a patient is having seizure activity
In shock conditions, the low amplitude of the blood pressure waveform may make
it difficult for this monitor to accurately determine the systolic and diastolic
pressures.
NIBP: Accuracy of any blood pressure measurement may be affected by the
position of patient.
All hose and cuffs used to the patient should be approved by KTMED.
Incorrect size or incorrect fitting may be result in the incorrect readings.
WARNING
Service personnel might be electrically shocked or parts might be damaged when repairing
the ANGELUS.
Before repairing, service personnel should follow to;
„
„
„
„
„
„
„
Disconnect the AC power.
Remove internal battery.
Antistatic himself.
Read this manual carefully.
To ensure safety, use only accessories & parts recommended or supplied by
KTMED
Always follow the manufacture’s instruction.
Unauthorized repairs will void the warranty.
-5-
Section 1 – General Information
WARRANTY
KTMED, Inc. warrants from the date of original purchase from KTMED or its authorized
distributors as followings:
Monitor: Two (2) Year
Battery: One (1) Year
Printer: One (1) Year
SpO2 Probes: One (1) Year
Temperature Probes: One (1) Year
Cuffs (all): Out-of-box failure only.
Service
Qualified electronic personnel should perform all repairs under the warranty. Unauthorized
repairs will void the warranty.
We do not support component level repair of the unit, there is in need of high level skill or
a special tool (SW).
As for service parts, refer to the part list in the manual which we provided.
„
„
„
„
„
If you have a problem that you cannot solve, contact KTMED
When returning a unit for service, you must send the form to us at first.
When shipping, please send the unit including accessories that might be
associated with the problem.
KTMED does not provide service without SRF (Service Request Form).
KTMED does not accept returned units without SRF.
SRF includes the following information:
„
„
„
„
„
„
Product name and serial number
Revision level of your software
The specific steps which reproduce your problem
Any error codes (photo) displayed on screen
A daytime phone number, fax number, and/or email address (if available)
Your name / company
Note: Save the original shipping container and it’s inside packing material should the
monitor need to be returned for service.
Refer to the section How To Contact Us, found in the Rear of this manual, for important
phone numbers, fax numbers and email addresses.
-6-
Section 1 – General Information
COMPOSITION
1. Main Body
1-1. Front
1-1-1. Front Case
1-1-2. Controller Board
1-1-3. Window
1-1-4. Key Pad
1-1-5. SpO2 Input Connector
1-1-6. NIBP Cuff Fitting
1-2. Rear
1-2-1. Rear Case
1-2-3. Power Board
1-2-4. NIBP Board
1-2-5. SpO2 Board
1-2-8. AC inlet
1-3. Battery
1-3-1. Rechargeable Battery
1-3-2. Battery Cover
2. Accessory
2-1. SpO2 Probe
2-1-1. Reusable Sensor
2-1-1-1. Adult Finger Sensor
2-1-1-2. Pediatric Finger Sensor
2-1-1-3. Y Probe
2-1-2. Disposable Sensor
2-2-1. Adult Finger Sensor
2-1-2-2. Pediatric Finger Sensor
2-1-2-3. Neonate Sensor
2-1-2-4. Multi-site Sensor
2-2. SpO2 Sensor Extension Cable
2-3 NIBP Cuff
2-3-1Adult Cuff
2-3-2. Child Cuff
2-4. AC Power Cord
-7-
Section 1 – General Information
Monitor Block Diagram
-8-
Section 1 – General Information
Module Descriptions
Controller board
Modules control & Display implementation
Power board
The power supply for each module
-9-
Section 1 – General Information
NIBP board
NIBP measurement module
SpO2 & Temperature board
SpO2 & Temperature measurement module
- 10 -
Section 1 – General Information
- 11 -
Section 2 – Functional Tests and Trouble Shooting
Functional Tests
Check 1: Power
After connect to AC power cord, and than AC power switch on, check the AC LED in
keypad of front panel. If AC LED light on, check 1 is OK. If AC LED light off, check 1 is
fail.
Check 2: Power Cord
If there are open wiring on AC power cord. Change the power cord.
Check 3: Fuse
After identify there aren’t any problems on check 2, if AC LED light off, check the fuses
in AC inlet. If fuse is broken, replace the fuses.
We can find it’s faulty by see with eye but if you can’t, you may use resister. If the
measurement of it’s register below some Ω, there aren’t any faulty but over some KΩ,
replace fuses to new one.
Check 4: Charge
Check the BATT LED in keypad of front panel. If BATT LED light on, check 4 is OK. If
BATT LED light off, check 4 is fail. In case of, you have to replace Power Board.
Check 5: Discharge
After battery charge enough, it will maintain about 1 hour normal operation, the battery is
no problems. However within 40 minutes, if you can find this case, you are needed to
replace the battery.
- 12 -
Section 2 – Functional Tests and Trouble Shooting
TROUBLE SHOOTING
Start
Check 1
Power
Fail
OK
Check 2
Power Cord
Fail
Change
Power Cord
OK
OK
Check 4
Charge
Check 3
Fuse
Fail
Fail
Change Fuse
Change Power
Board
OK
Check 5
Discharge
Fail
Change Battery
OK
Check 6
LED
Fail
Change LED
Fail
Change
Control Board
OK
Check 7
Control
Board
OK
A
- 13 -
Section 2 – Functional Tests and Trouble Shooting
A
Check 8
Module
Fail
Fail SpO2 Board
Change SpO2 Board
Fail NIBP Board
Change NIBP Board
OK
SpO2 Performance Test
NIBP Performance Test
The set has no
problem
SpO2 Performance Test
Change Sensor
or Extension Cable
Check 12
Sensor
Check 13
Waveform
& SpO2
Value
Change SpO2 Board
- 14 -
Section 2 – Functional Tests and Trouble Shooting
OK
The SpO2
Board has no
problem
NIBP Performance Test
Check 14
Cuff Leak
Fail
Change Cuff or
Extension Hose
OK
Check 15
Measuring
Test
Fail
Change NIBP Board
OK
The NIBP
Board has no
problem
- 15 -
Section 2 – Functional Tests and Trouble Shooting
Error message lists:
Message/Problem
Units does not run
Possible Cause
Solution
No power to unit
Check the connection.
Faulty electrical connection
Replace keypad.
Faulty electrical button
Replace power board.
Defective power board
Battery run time is low Defective battery
Replace the battery.
LED of Battery is
Defective battery
Replace the battery.
blinking
Defective power board
Replace power board.
Faulty connection between
Check the connection.
LED and the main board
Replace the LED.
Missing LED display
Replace the main board.
Displayed SpO2 not
accurate
Defective SpO2 sensor
Check the connection.
connection
Replace SpO2 sensor cable.
Defective SpO2 sensor cable
Replace SpO2 board.
Defective SpO2 board
If problem persists, return
Require calibration
for service.
Check cuff gilt and
NIBP error code11
Cuff fitted too loosely and or
position.
not connected
If problem persists, return
for service.
Readjust cuff.
NIBP error code12
Inflation Error
If problem persists, return
for service.
Replace NIBP board.
NIBP error code13
Pulse detection Error
If problem persists, return
for service.
Check that the cuff is in the
NIBP error code14
Out of measurement range
properly tightened.
Check the hose connection.
NIBP error code15
Deflation Error
- 16 -
Check that the cuff hose has
Section 2 – Functional Tests and Trouble Shooting
no bend.
Check that the silicon hose
of NIBP board is correct.
Check the Solenoid.
If problem persists, return
for service.
NIBP error code16
Excessive pressure (Out of
IEC)
Check the condition of cuff.
If problem persists, return
for service
Replace NIBP board.
NIBP error code17
Pulse signal is too high
If problem persists, return
for service.
Check that the patient is
NIBP error code18
Two much movement during moving too much.
measurement
If problem persists, return
for service.
Replace NIBP board.
NIBP error code19
Abrupt Leakage
If problem persists, return
for service.
It needs calibration.
NIBP error code20
Systolic Detection Error
If problem persists, return
for service.
It needs calibration.
NIBP error code21
Diastolic Detection Error
If problem persists, return
for service.
Replace NIBP board.
NIBP error code22
Out of pressure range
If problem persists, return
for service
NIBP error code 23
NIBP error code30
Leakage detection during
leakage test
Replace NIBP board.
If problem persists, return
for service.
PUMP open Error
- 17 -
Replace Air pump
Section 2 – Functional Tests and Trouble Shooting
If problem persists, return
for service
Replace Solenoid 1.
NIBP error code31
Solenoid 1 Open Error
If problem persists, return
for service
Replace Solenoid 2.
NIBP error code32
Solenoid 2 Open Error
If problem persists, return
for service
Replace NIBP board.
NIBP error code99
System Error
If problem persists, return
for service.
- 18 -
Section 3 – Disassembly and Assembly
Disassembly and Assembly
Prior to Disassembly
1. Turn the unit power off
2. Disconnect the unit from AC power
3. Remove the battery from the monitor
Disassembly Battery
1. Remove 2 screws of the battery cover as the below figure
2. Remove the battery cover. Disconnect the battery from the unit after disconnecting
the connector. Remove 2 screws in the battery case
- 19 -
Section 3 – Disassembly and Assembly
Disassembly Front and Rear Case
1. Remove 5 screws of Rear case
2. Separate Front and Rear case slowly by using the both hands.
Caution: Take care when pulling the Front case from Rear case.
If the header pins of the control board are got bend, the unit could not work properly.
- 20 -
Section 3 – Disassembly and Assembly
Disassembly Controller Board
1. Disconnect the cable as shown in figure.
1.
Remove 10 screws & disconnect the speaker connector to separate the control board
from the Front case.
- 21 -
Section 3 – Disassembly and Assembly
2.
Separate the Front case and the control board.
Disassembly NIBP & SpO2 Board
1. Remove 3 screws of the NIBP board
2. Disconnect the cable to the NIBP board.
- 22 -
Section 3 – Disassembly and Assembly
3. Remove the 2 hoses.
4. Disconnect the cable from the unit.
- 23 -
Section 3 – Disassembly and Assembly
Disassembly Power board
1. Remove the 1 screw, 2 supporters, battery cable from the Power board as shown in
the figure.
2. Disconnect the cable from the unit
- 24 -
Section 3 – Disassembly and Assembly
Replacing Fuse
1. Turn the monitor “OFF” and disconnect the power cord.
2. Open the fuse box, use of minus screw driver
3. Carefully remove the fuse from the rear panel of the monitor. Refer to the
following Figure.
4. Insert the new fuse
- 25 -
Section 4 – Service
SpO2 Calibration
Equipment Required
SpO2 simulator
DC Power supply
SPo2 Cable & Extension cable
Test Jig
Service Mode
1. Power the monitor on.
2. Press Confirm key.
3. Leave the monitor in Service mode for SpO2 calibration
Calibration Mode
1. Press Setup/Menu key.
2. Press Start/Stop key.
- 26 -
Section 4 – Service
3.
4.
5.
6.
7.
8.
Press ESC/Silence key.
Press Up key.
Press Up key.
Press Down key.
Press Confirm key.
Press Down and Up keys repeatedly till the below mode appears.
9. Press Confirm key.
10. Connect the unit to SpO2 simulator using the sensor cables
- 27 -
Section 4 – Service
11. Press Start/Stop key.
12. Press Confirm key when showing more than 5 bars on the display of 10-Bars.
13. You can see that the displayed value on the DIA/MAP is changed from 100 to 98
14. Set SpO2 simulator to 98.
- 28 -
Section 4 – Service
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
31.
Press Confirm key when showing more than 5 bars on the display of 10-Bars.
You can see that the displayed value on the Dia/MAP is changed from 98 to 95.
Set SpO2 simulator to 95.
Press Confirm key when showing more than 5 bars on the display of 10-Bars.
You can see that the displayed value on the Dia/MAP is changed from 95 to 90.
Set SpO2 simulator to 90.
Press Confirm key when showing more than 5 bars on the display of 10-Bars.
You can see that the displayed value on the Dia/MAP is changed from 90 to 85.
Set SpO2 simulator to 85.
Press Confirm key when showing more than 5 bars on the display of 10-Bars.
You can see that the displayed value on the Dia/MAP is changed from 85 to 65.
Set SpO2 simulator to 65.
Press Confirm key when showing more than 5 bars on the display of 10-Bars.
You can see that the displayed value on the Dia/MAP is changed from 65 to 30.
Set SpO2 simulator to 30.
Press Confirm key
Press Confirm key to save this process.
NIBP Calibration
NIBP device requires that the accuracy shall be +/-3mmHg. It’s important to verify the
calibration. Return the unit to our distributor or KTMED if you need the calibration.
Equipment Required
NIBP simulator (Digital manometer)
DC Power supply
Voltage meter
- 29 -
Section 4 – Service
Service Mode
1. Power the monitor on.
2. Press Confirm key.
3. Leave the monitor in Service mode for NIBP calibration
Calibration Mode
1.
2.
3.
4.
5.
6.
7.
Press Setup/Menu key.
Press Start/Stop key.
Press ESC/Silence key.
Press Up key.
Press Up key.
Press Down key.
Press Confirm key.
- 30 -
Section 4 – Service
8. Press Down and Up keys repeatedly till the below mode appears.
9. Press Confirm key.
- 31 -
Section 4 – Service
10. Press Start/Stop key
11.
12. Verify that the displayed values on the SYS and DIA are within 0.5+/- 0.03.
The current pressure is 0 mmHg.
(Note: If those values are beyond the limit, it needs to adjust hardware)
13. Press Confirm key
14. Attach the hose (the target pressure: 300mmHg) to the monitor when 3 LED bars
are on. (Note: Please neglect “Err” displayed on Temp)
- 32 -
Section 4 – Service
Verify that the displayed values on the SYS and DIA are within 3.2+/- 0.1.
(Note: If those values are beyond the limit, it needs to adjust hardware)
15. Press Confirm key to save
16. Detach the pressure hose from the monitor and Press Confirm key
when 4 LED bars are on.
- 33 -
Section 4 – Service
17. Press ESC/Silence key
18. Press ESC/Silence key to exit the calibration mode.
NIBP Adjustment
This section describes the calibration procedure in case that the measurement values of
NIBP are beyond the spec.
Set up
1. Remove the NIBP module (MST modules) form the main unit as below.
- 34 -
Section 4 – Service
2. Set the voltage of DC power supply to 5V+/- 0.1V.
3. Connect the DC input cable to the power socket as below.
1 pin
- 35 -
Section 4 – Service
Power
input
-
++
4. Switch Digital nanometer or equivalent to 330mmHg.
Adjusting the measurement under 0 mmHg
1. Remove the hoses on Sens 1 and Sens 2.
- 36 -
Section 4 – Service
Sens
pin
2. Connect the (-) ground & (+) probe of DC Voltage meter to GND pad and TP 1
(Test point) on the board as bellows.
GND
pin
Remove TP
VR
- 37 -
Section 4 – Service
3. Set the displayed voltage on DC voltage meter to within 0.5V+/- 0.01V by
switching VR2 on board.
Target value
4. Reconnect the (-) ground & (+) probe of DC Voltage meter to GND pad and TP 3
(Test point) on board as belows.
- 38 -
Section 4 – Service
GND
pin
Remove
TP
5. Set the displayed voltage on DC voltage meter to within 0.5V+/- 0.01V by
switching VR4 on the board.
- 39 -
Section 4 – Service
Target value
Adjusting the measurement under 330 mmHg
1. Remove the hose of Sens2.
2. Connect the hose (the target pressure: 330mmHg) on only Sens 1.
Sens
1 pin
1
- 40 -
Section 4 – Service
3. Connect the (-) ground & (+) probe of DC Voltage meter to GND pad and TP 1
(Test point) on board as below.
GND
pin
TP
4. Set the displayed voltage on DC voltage meter to within 3.5V+/- 0.1V by
switching VR1 on board.
- 41 -
Section 4 – Service
Target value
5. Remove the hose on Sens 1. Connect the hose (the target pressure: 330mmHg) on
only Sens 2.
Sens
1 pin
2
6. Connect the (-) ground & (+) probe of DC Voltage meter to GND pad and TP 3
(Test point) on board as below.
- 42 -
Section 4 – Service
GND
pin
TP
7. Set the displayed voltage on DC voltage meter to within 3.5V+/- 0.1V by
switching VR3 on board.
Target value
7. Reassembly
- 43 -
Section 5 – Replace Part Lists
Replace Part Lists
Part Name
ANGELUS-ABS
ANGELUS-WINDOW
ANGELUS-RUBBER
FOOT
ANGELUS-KEY PAD
ANGELUSCONTROL PCB
ANGELUS-POWER
PCB
ANGELUS-MODULE
PCB
ANGELUS-MODULE
PCB
ANGELUS-SPO2
INNER CABLE
ANGELUS-NIBP
INPUT FITTING
ANGELUS-LAN
INNER CABLE
ANGELUSSPEAKER CABLE
ANGELUS-DC
INNER CABLE
ANGELUS-AC
INLET
ANGELUSBATTERY SET
AC Power Cord
NIBP Adult Big Size
Cuff
NIBP Adult Cuff
NIBP Child Cuff
Cuff Hose
ADP-02
C Clamp
ANGELUSCALIBRATION
Part No
ANKP-001-001
ANKP-001-003
ANKP-001-004
Description
Plastic case
LED Screen Window
Epoxy
ANKP-001-005
ANKP-004-018
Membrane Key Pad
Main Controller Board
ANKP-004-013
Power Board
ANKP-004-016
SpO2
ANKP-004-015
NIBP Board
ANKP-005-0001
SpO2 Inner Harness Cable
ANKP-005-0002
ANKP-005-0003
NIBP
Input
Fitting(within
Silicon Hose)
LAN Port Inner Harness Cable
ANKP-005-0004
Speaker Inner Harness Cable
ANKP-005-0005
DC Jack Inner Harness Cable
ANKP-005-0006
AC Inlet(within Fuse)
ANKP-002-023
Battery Pack
AC Power Cord(1.8m)
KTPM-2002, Angelus
KTPM-2002, Angelus
KTPM-2002, Angelus
KTPM-2002, Angelus
KTPM-2002, EPOX, Angelus
POX, EPOX, Angelus IV Clamp
Calibration
- 44 -
Section 6 – Contact to Manufacture
Contact Information
If you need additional information or if you have any technical, methodical or
clinical questions please contact the manufacturer by e-mail:
E-Mail :
master@ktmed.com
regu@ktmed.com
Otherwise you can contact us as follows:
Phone :
Facsimile :
+82-2-2108-7890
+82-2-2108-7894
Mail : #610, Hanshin IT Tower, 235
Guro-3Dong, Guro-Gu
Seoul, Korea
In the case of a request we would ask you to inform us about the serial number of
your device which you can find on the type plate on the back or bottom side of the
device.
- 45 -
Section 6 – Contact to Manufacture
Service Request Form
Please fill one form out for each individual unit, which needs service.
Return completed form to KTMED: +822 21087894 (fax) or email: regu@ktmed.com
Distributor/Hospital : ___________________________Date: ___________
Contact Person: ____________________________________
Phone: ________________________
Fax: ______________________
Email: ________________________
Address: _________________________________Country:___________________
Unit Information/Problem
Name of Product: _________________ Unit Serial #:
Accessories & Serial #: ______________________
___________________
A. User’s Description of Problem (step by step) :
NOTE: Please send the faulty unit including all accessories (cables, adapter, etc) associated with
the problem.
Service Description
A. Date Received:
B. Service Technician’s Diagnosis
C. Service Technician’s Action Description:
- 46 -
Download