eDAQ Z100™ Electrochemical Impedance analyzer

advertisement
Modular Impedance Analyzer system
Front Panel
Rear Panel
eDAQ
Z100™ Electrochemical Impedance analyzer
Description
Electrochemical Impedance Spectroscopy (EIS) has emerged as a powerful technique for
defining reaction mechanisms and physical structures involved in a wide variety of applications
including
•
•
•
•
Corrosion processes
Battery and fuel cell
Biological systems
Surface structures
The EDAQ Z100™ EIS System provides the means to turn an existing potentiostat/galvanostat
or electronic load into a fully functioning EIS System with full analysis capability.
The single channel Z100 is based on the experience gained from the development of a 6 channel
EIS system and includes many useful enhancements.
The EDAQ Z100™ EIS system covers the frequency range of mHz to 100kHz and the ZMAN
software provided with the system generates all required analysis outputs.
The complete system is software controlled and all functions such as ranging, calibration, and
measurements can be automated.
System Features
•
•
•
•
•
•
•
•
•
Uses external Potentiostat/Galvanostat or electronic loads to provide a versatile AC
impedance measuring system.
Excitation signal output (Sig out) and 2 Signal Inputs channels (Ein and Iin) to compute
Impedance.
Flexible frequency generator/analyser
Designed for electrochemistry applications
Simulation and fitting with ZMAN
High Current applications with external load or potentiostat/galvanostat.
GUI and software controlled functions
Dual real time graphs (Bode, Nyquist etc) during measurement
Power Spectrum and Waveform Generation
Modes of Operation
With EDAQ EA161 potentiostat or 3rd Party potentiostat/galvanostat.
•
•
•
•
•
•
Potentiostatic EIS (PEIS)
Galvanostatic EIS (GEIS)
Galvanostatic HFR (GHFR)
Cyclic voltametry(CV)
Linear Sweep voltametry (LSV)
Square wave voltametry (SWV)
With External Electronic Load
•
•
Frequency response Analyser (FRA)
High Frequency resistometry
Software features
Z Navigator is a software application included with the system to allow the system to be setup
for measurement, data recording and display. It contains the following functions and Menus
•
•
•
•
•
•
Device Selection: Selects the device to be used in the experiment eg EA161 Potentiostat
Experiment Selection: This menu selects the experiment to be performed eg
Potentiostatic EIS
Measurement Panel: this selects the frequencies that will be applied during the
experiment.
Monitoring (real time) : This window displays the applied signals and the outputs and
generates a table of numerical results.
Display Modes
a. Power Spectrum
b. 3D Impedance Plots (Real time)
c. Nyquist Plot (Real time)
d. Bode Plot (Real time)
e. Lissajous plots
File Management: provisions to save, delete and recall
ZMAN is a software analysis and modeling software supplied with each Z100 system. It
provides full facilities for manipulating , editing , analyzing, modeling and display features – see
ZMAN data sheet.
Specifications
Analog Out:
BNC connector:
Configuration:
Max output:
Resolution:
DC zero offset:
functions as signal generator
Single –Ended
-11V to +11V (DC +AC peak)
16 bit
<0.5mV, software corrected.
DC Bias Range (Volts):
-10 to +10
-0 to +10
-5 to +5
0 to +5
-2.5 to +2.5
-2.5 to +7.5
AC Waveform(Volts)
Types:
DC, Sine, Cosine, Ramp, Sawtooth, Triangle, Square, Pulse, Multi-tone
Frequency :
1µHz to 100KHz
Freq Resolution: 5000 steps /decade
Freq Accuracy:
typ 75ppm, Max +/-200ppm
Freq Stability:
<2ppm @ 1kHz
<20 ppm @ 10kHz
<200ppm @ 100kHz
<2000ppm(0.2%) @ 1MHz
Amplitude range: 1mVpp to 5 Vpp
Gain/Attenuation: -44dB to +40dB (x100) in 6dB steps –Automatic Gain selection
Reconstruction filter: 10kHz to 150kHz 8th Order low pass – 10kHz steps or Bypass
Gain error: <0.5%
Analog In:
System Ports:
Ein BNCfor Voltage signal input
Iin BNC for Current signal input
Configuration: Differential
Max Common Mode voltage: +/-100 VDC
Offset: <0.5mV
Bandwidth: 550 kHz
Input impedance: 110 kOhms
Pre-attenuation: -20dB (x0.1)
Gain/Attenuation: -44dB to +40dB (x100) in 6dB steps –Automatic Gain selection
Anti-Aliasing filter: 10kHz to 150kHz 8th Order low pass – 10kHz step or Bypass
CMRR
>80dB @1kHz
> 60dB @10kHz
>40dB @ 100kHz
I2C port for connection to EDAQ amplifiers
USB 2.0 high Speed communications port
Physical Configuration:
Physical (w x h x d):
Weight:
Operating voltage:
Power Comsumption:
Operating conditions:
300 x 60 x 300 mm (12” x 2.5” x 12”)
3 kg
100-240VAC @250mA 50-60Hz
80VA
0 to 35°C, 0-90% humidity non-condensing