UK/2972 - Gov.uk

advertisement
UK 2972
V(0)a
EC Type-Approval
pproval Certificate
UK 2972
Issued by the:
National Measurement Office
Notified Body Number 0126
In accordance with the requirements of the Non
Non-Automatic
utomatic Weighing Instruments
Regulations 2000 (SI 2000/3236) which implement, in the United Kingdom, Council
Directive 2009/23/EC, this EC Type-Approval Certificate
ertificate has been issued to:
Excell Precision Co., Ltd
6F, No 127, Lane 235
Pao-Chiao Road
Hsin-Tien District
New Taipei City
Taiwan
in respect of a Non-Automatic
utomatic Weighing Instrument
nstrument designated the FJ3-XXX, GJWXXXX, HJW-XXX or HJP--XXX Hanging Scales where (XXXX
XXXX denotes alternative
approved models) and having the following characteristics:
n ≤ 3 750 with single range and multi range Class III
Max ≤ 15 000 kg (Depending on instrument configuration)
The necessary data (principal characteristics, alterations, securing, functioning etc)
for identification purposes and conditions (when applicable) are set out in the
descriptive annex to this certificate.
Issue Date:
Valid Until:
Reference No:
30 May 2014
29 May 2024
TS1201/0062
Signatory: P R Dixon
for Chief Executive
National Measurement Office | Stanton Avenue | Teddington | TW11 0JZ | United Kingdom
Tel +44 (0)20 8943 7272 | Fax +44 (0)20 8943 7270 | Web www.gov.uk/nmo
NMO is an Executive Agency of the Department for Business Innovation & Skills
0135
NAWI Rev. 2 01 April 2011
Descriptive Annex
1
INTRODUCTION
These weighing instruments are designated the FJ3, GJW, HJW or HJP Hanging
Scales. There are designed to be used as a Class III Non-Automatic Weighing
Instrument. The Hanging Scale are self-indicating, mains or battery-powered, and
can operate as a single or multi range (HJP and HJW) weighing instrument.
The indicator consists of an analogue to digital conversion circuitry, microprocessor
control circuitry, power supply, function keypad, non-volatile memory for storage of
calibration and setup data, and a weight display contained within a single enclosure.
2
FUNCTIONAL DESCRIPTION
2.1
Construction (Figure 1)
2.1.1
Mechanical
− Aluminium Alloy construction
− Six digit segments LED digital screen (FJ3)
− Six digit segments LED or LCD digital screen available in various
colours (HJP, HJW and GJW models)
− Operator function keypad
− Plastic enclosure
− LED indicators (battery status: charging or charged)
− Dual display (HJP and HJW)
2.1.2
Devices
Initial zero setting (≤ 20% Max)
Semi-automatic zero setting (≤ 4% Max)
Zero tracking (≤ 4% Max)
Semi-automatic subtractive tare weighing
Determination of stability of equilibrium
Preset tare
Accumulation
High-resolution mode (5 sec maximum)
Unit switch (kg, t)
Gross, Net, Motion, Accumulation, Preset tare, Weighing range
indicators
− Remote operation
− Hold function
−
−
−
−
−
−
−
−
−
−
2/8
2.2
Load cells
2.2.1
Any load cells may be used provided the following conditions are met:
−
there is a respective OIML Certificate of Conformity (R60) or a EC Test
Certificate (EN 45501) issued for the load cell by a Notified Body
responsible for type examination under Directive 2009/23/EEC;
−
the certificate contains the load cell types and the necessary load cell
data required for the manufacturer’s declaration of compatibility of
modules (WELMEC 2, Issue 5, 2009, No. 11), and any particular
installation requirements. A load cell marked NH is allowed only if
humidity testing to EN 45501 has been conducted on this load cell;
−
the compatibility of load cell and indicator is established by the
manufacturer by means of the compatibility of modules form, contained
in the above WELMEC 2 document, at the time of EC verification or
declaration of EC conformity of type; and
−
the load transmission must conform to one of the examples shown in
the WELMEC 2.4 “Guide for load cells”.
2.3
Operation
2.3.1
Switch-on
At switch-on, the instrument performs a self-test during which all the relevant
segments of the display are lit. The instrument attempts to set zero provided that it
is within a range not exceeding 20 % of its capacity.
2.3.2
Semi-automatic zero setting
The zero button operates provided that the instrument is within a range not
exceeding 4 % of its capacity.
2.3.3
Over- and under-range
If the load is less than zero, then the weight indications show an error “E2”. The load
may be shown up to a maximum of nine divisions above Max, above which the
weight indications are inhibited and replaced by “E1”.
2.3.4
Tare
Subtractive tare balancing can be performed.
3/8
3
TECHNICAL DATA
3.1
Indictor specification
Power supply
6 V DC 7 or 10 AH
Maximum number of scale intervals
Maximum Tare
Maximum Preset Tare
Load cell excitation voltage
Minimum load cell impedance
Maximum load cell impedance
Minimum input voltage per scale
interval
Measuring range minimum voltage
Measuring range maximum voltage
Fraction of maximum permissible error
Operating temperature range
Load cell connection
3 750
6 kg or 15 000 kg (depending on configuration)
6 kg to 15 000 kg (depending on configuration)
5 V DC
350 Ω
1 200 Ω
1 µV
3.2
-1 mV/V
3 mV/V
Pind = 0.5
-10 °C / +40 °C
4 or 6 wire connection maximum length = 2 m/mm2
Power supply
The Hanging Scales are mains or battery-powered. A mains adaptor supplies 6 V
DC to the Hanging Scales, any CE-marked adaptor may be used. The battery may
be a 6 V DC 7 or 10 AH rechargeable battery. If the battery or supply voltage is too
low, the display will show a battery symbol.
4
PERIPHERAL DEVICES AND INTERFACES
There are no interface ports, and therefore no peripheral devices can be connected.
4.3
Remote operation
The instrument may be operated remotely from an internally wired remote control
unit (HJW model) or via a non-wired remote control unit (GJW, FJ3, HJP model).
5
APPROVAL CONDITIONS
The certificate is issued subject to the following conditions:
5.1
Legends and inscriptions
5.1.1
The instrument shall bear the following legends near the display of the
weighing result:
Max
Min
e=
4/8
5.1.2
The instrument shall bear the following legends
CE mark
Green M
Accuracy class
Serial number
Manufacturer’s mark or name
Certificate number
6
LOCATION OF SEALS AND VERIFICATION MARKS
6.1
The rating plate (Figure 5) is located at the back of the instrument. The
CE mark shall be impossible to remove without damaging it. The rating plate shall be
impossible to remove without it being destroyed.
The markings and inscriptions shall fulfil the requirements of Paragraph 1 of
Annex IV of the Directive 2009/23/EC.
−
−
7
a mark of the manufacturer and/or manufacturer’s representative, or
an official mark of a verification officer.
ALTERNATIVES
There are no authorised alternatives.
8
ILLUSTRATIONS
Figure 1
Figure 2
Figure 3
Figure 4
Figure 5
FJ3 – Weighing Hanging Scale (300 kg ~ 15 000 kg)
GJW – Weighing Hanging Scale (200 kg ~ 1 000 kg)
HJW – Weighing Hanging Scale (6 kg ~ 60 kg)
HJP – Pricing Hanging Scale (6 kg ~ 100 kg)
Rating plate
9
CERTIFICATE HISTORY
Issue №.
Date
Description
UK 2972
11 June 2014
Type approval first issued.
---
---
---
5/8
Figure 1 FJ3 – Weighing Hanging Scale (300 kg ~ 15 000 kg)
6/8
Figure 2 GJW – Weighing Hanging Scale (200 kg ~ 1 000 kg)
Figure 3 HJW – Weighing Hanging Scale (6 kg ~ 60 kg)
7/8
Figure 4 HJP – Pricing Hanging Scale (6 kg ~ 100 kg)
Serial No. XXXxxxxx (X=A~Z, x=0~9)
Figure 5 Rating plate
Crown Copyright 2014
National Measurement Office
Department of Innovation, Universities & Skills
8/8