Version: Page: Issue Date: 02 1 of 1 3 June 1990 pH REFERENCE ELECTRODE JUNCTION POTENTIAL CHECK pH Electrode Description or Serial No.: DATE pH 6.88 BUFFER 1:10 pH 6.88 BUFFER TEMP C ANALYST CHECK Measure of Reference Electrode Junction Potential The effect of junction potential can be checked by measuring the pH, of preferably a pH 6.88 phosphate buffer solution and comparing it with the pH of the same buffer solution diluted with 9 parts of freshly boiled distilled water. Pipette 5ml pH 6.88 buffer and dilute to 50mL in a volumetric flask with freshly distilled water. The pH of the diluted buffer should be 0.20.05units higher than the original buffer. If the value obtained is outside the range, the reference electrode may require attention. Inform the Shift Chemist. pH range = 7.03 –7.13 To make pH 6.88 buffer, add 3.387g KH2PO4 and 3.533g Na2HPO4 to 1L volumetric flask and make to volume with freshly boiled (carbon dioxide free) distilled water and mix well. 05-022